VITROS Chemistry Products Performance Verifier is used as an assayed control to monitor the performance on…
- Date
- March 7, 2018
- Risk class
- Class III (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences)
- Status
- Terminated
- Filed by
- Ortho Clinical Diagnostics Inc
- Product
- VITROS Chemistry Products Performance Verifier is used as an assayed control to monitor the performance on VITROS 250/350/5, 1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Used as a control for the VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
- Reason
- VITROS Performance Verifier II (PVII) Lot K4852 does not meet long term stability requirements for the AST Analyte (only) when the product is stored refrigerated for up to 6 months.
- Distribution pattern
- Worldwide
- Quantity
- 10,021 units
- Codes and lots
- Lot K4852, Expiration 22 Feb 2018
- Recall started
- July 10, 2017
- Device type
- Multi-Analyte Controls, All Kinds (Assayed)
- Root cause (FDA)
- Unknown/Undetermined by firm
- What the company told customers
- On July 10, 2017 the firm sent notification letters and an e-communications to their customers. Foreign customers were notified by email. Customers were instructed to complete a response form with the amount of product on hand and to discontinue use of the recalled product. For further questions, please call (585) 453-3452.
- Recall number
- Z-0807-2018
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0807-2018. The agency record governs; check it and the company before acting.