Arjohuntleigh Getinge Group…

Feb. 13, 2019 Class II

Date
Feb. 13, 2019
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Arjohuntleigh Magog
Product
ARJOHUNTLEIGH GETINGE GROUP, RotoProne Therapy System, Model Number 209800
Reason
Inability or difficulty in unbuckling/releasing the proning arm buckle might delay patient access in an emergency situation such as cardiac arrest. This, in turn, could contribute to a life-threatening event.
Distribution pattern
US, Saudi Arabia, and Kuwait
Quantity
231 units
Codes and lots
All Serial Numbers
Recall started
Dec. 19, 2018
Device type
Bed, Patient Rotation, Powered
Root cause (FDA)
Component design/selection
What the company told customers
Arjo hand-delivered the notice with each RotoProne Therapy System rental and delivery beginning on 12/19/2018. Arjo will require the customer to confirm the notification has been received and understood via the signed delivery notification.
Recall number
Z-0802-2019

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0802-2019. The agency record governs; check it and the company before acting.