GE LightSpeed VCT / Light Speed VCT standard configuration… and 4 more products

May 15, 2007

Date posted by the FDA
May 15, 2007
Status
Terminated
Filed by
GE Healthcare
Reason
Oil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.
Distribution pattern
Worldwide Distribution: USA in the following states and locations: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, Guam and Puerto Rico. Foreign countries include ; Albania, Algeria, Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Canada, Chile, China, Columbia, Czechoslovakia, Denmark, Finland, France, Germany, Great Britain, Greece, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Martinique, Mexico, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Reunion Island, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Venezuela, and Vietnam

5 products in this recall

GE LightSpeed VCT / Light Speed VCT standard configuration…

Product
GE LightSpeed VCT / Light Speed VCT standard configuration, Model 5124069, CT Scanner System.
Quantity
1,172
Codes and lots
All serial numbers
Recall started
Dec. 21, 2006
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Other
What the company told customers
A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.
Recall number
Z-0802-2007

GE LightSpeed VCT Select / LightSpeed Pro32…

Product
GE LightSpeed VCT Select / LightSpeed Pro32, Model 5126093, CT Scanner system.
Quantity
158
Codes and lots
All serial numbers
Recall started
Dec. 21, 2006
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Other
What the company told customers
A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.
Recall number
Z-0803-2007

GE LightSpeed Pro 16 (100kW/80kW) Models 2357739…

Product
GE LightSpeed Pro 16 (100kW/80kW) Models 2357739, 2357739-3, 2357739-4, 2357739-6, 2357739-7 and 2357739-8, CT Scanner systems
Quantity
462
Codes and lots
All serial numbers
Recall started
Dec. 21, 2006
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Other
What the company told customers
A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.
Recall number
Z-0804-2007

GE LightSpeed Xtra / GE LightSpeed RT Pro 16…

Product
GE LightSpeed Xtra / GE LightSpeed RT Pro 16, Model 2374681-6, CT Scanner systems.
Quantity
21
Codes and lots
All serial numbers
Recall started
Dec. 21, 2006
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Other
What the company told customers
A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.
Recall number
Z-0805-2007

GE Discovery VCT…

Product
GE Discovery VCT, Volume PET-CT scanner, Model 5124069-3.
Quantity
4
Codes and lots
All serial numbers
Recall started
Dec. 21, 2006
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Other
What the company told customers
A Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.
Recall number
Z-0806-2007