Medline Fluid Delivery Set with Drip Chamber: 1)… and 1 more product

Recall started · FDA enforcement report Jan. 22, 2025 Class I

Reason
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Status
Ongoing
Distribution
US Nationwide Distribution

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Jan. 22, 2025
Risk class
Class I (FDA's definition, 21 CFR 7.3(m): a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
MEDLINE INDUSTRIES, LP - Northfield
Reason
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Distribution pattern
US Nationwide Distribution

What this record is about

A generic icon of the kind of thing the record is about; the record's own words are below. It is not a picture of the product.

A generic icon of the kind of thing the record is aboutA simplified icon of the kind of thing the record is about and the phrases of the record's own text beside it.1 The record: Medline Fluid Delivery Set with Drip…2 In the record's words:3 Only the kind of thing is drawn4 The record's own words are below“Fluid Delivery Sets were incorrectly assembled with a white macrodrip chamber instead of the required grey micro drip chamber.”MEDLINE FLUID DELIVERY…

“Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.”Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223, FDA enforcement report, 2025-01-22. The caption is the record's own words.

The picture shows the kind of thing the record is about (a part, a product, a package), not the product itself; the ring marks it. Nothing in it is a claim beyond the record's own words below.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Set, Administration, Intravascular

2 products in this recall

Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG…

Product
Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223
Quantity
13,600 units
Codes and lots
1)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000128741; 2)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000134505; 3)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000133893; 4)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000135769; 5)99000069UDI-D:10193489069587(each)30193489069581(case), Lot Number:0000139596; 6)949000222UDI-D:10197344018550(each)30197344018554(case), Lot Number:0000133018; 7)949000224UDI-D:10197344018642(each)30197344018646(case), Lot Number:0000136573; 8)949000221UDI-D:10197344018567(each)30197344018561(case), Lot Number:0000133036; 9)949000223UDI-D:10197344018543(each)30197344018547(case), Lot Number:0000133031;
Recall started
Dec. 16, 2024
Voluntary or mandated
Voluntary: Firm initiated
Device type
Set, Administration, Intravascular
Root cause (FDA)
Under Investigation by firm
What the company told customers
An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 12/16/24 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-24-270-FG Recall Code: 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. Update: A follow-up letter titled IMMEDIATE ACTION REQUIRED dated 2/11/25 was sent to customers. Medline would like to update our action to instruct you to please destroy all impacted convenience kits and Medline will provide replacement kit(s) with the correct Fluid Delivery Set. REQUIRED ACTION: 1. Immediately check your stock for the affected kit number(s) and lot number(s) listed on the attachment. Please destroy all affected kits. 2. Please complete and return the enclosed destruction form listing the quantity of affected kits destroyed. Even if you do not have any affected product, please complete and return the form. 3. Upon receipt of your completed destruction form, Medline will issue replacem
Recall number
Z-0764-2025

Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES…

Product
Medline Fluid Delivery Set with Drip Chamber: 1) MTO,135,YELLOW,MICRO,SPK-MAIMONIDES, Item Number: 60130527; 2) MTO,ANGIO,DOUBLE,SPIKE,KIT,KENNEDY HLTH, Item Number: 65111341; 3) MTO,ANGIO,SINGLE,SPIKE,KIT, Item Number: 65232486; 4) MTO,ANGIO,SINGLE,SPIKE,KIT,GOOD,SAM, Item Number: 65071193; 5) MTO,NEURO,FLUSH,SPIKE,KIT,BLUE,FLORIDA, Item Number: 600604422; 6) MTO,NEURO,KIT,-,JERSEY,SHORE PG, Item Number: 60100045; 7) MTO,RADIOLOGY,KIT,HAMOT MED CEN, Item Number: 650801715; 8) MTS,135,CDS,IV,TUBE-JEFFERSON UNIV., Item Number: 60232481; 9) MTS,135,YELLOW,MICRO,SPIKE,LUTHERAN MED, Item Number: 60120583; 10) MTS,ANGIO,NEURO,KIT,NORTH SHORE UNIV, Item Number: 601403920; 11) MTS,CUST. FILTER-NEWYORK UNIVERSITY, Item Number: 62140602; 12) MTS,INTERV,NEURO,KIT-WESTCHESTER,MC, Item Number: 602301614; 13) MTS,NEU,NORTH SHORE HOSPITAL, Item Number: 6010771601; 14) MTS,NEURO,KIT, Item Number: 600602914; 15) MTS,NEURO,KIT,2,NEWENGLAND MED CTR, Item Number: 601400111; 16) MTS,NEURO,KIT,YLW&RED,FLORIDA HOSP, Item Number: 600604423; 17) MTS,NEURO,KIT-JOHN,F,KENNEDY MED CENTER, Item Number: 60110552; 18) MTS,NEURO,KIT-OVERLOOK HOSPITAL PG, Item Number: 60150035; 19) MTS,NEURO,SPIKE,KIT,CAPITAL HLTHSYS-FULD, Item Number: 60182723; 20) MTS,NEURO-RADIOLOGY,KIT-YALE-NEWHAVEN, Item Number: 602500214; 21) MTS,STROKE/EMBO,KIT,NORTH SHORE UNIV, Item Number: 601403921; 22) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601305210; 23) MTS,YELLOW,/,RED,MULTILINE,KIT,MAIMONIDE, Item Number: 601400110;
Quantity
3930 units
Codes and lots
1) 60130527 UDI-D:10193489046465(each), 20193489046462(case), Lot Number:0000133032; 2) 65111341 UDI-D:10193489061314(each), 20193489061311(case), Lot Number:0000129541; 3) 65111341 UDI-D:10193489061314(each), 20193489061311(case), Lot Number:0000134701; 4) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000129544; 5) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000128715; 6) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000133060; 7) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000131657; 8) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000135723; 9) 65232486 UDI-D:10193489063950(each), 20193489063957(case), Lot Number:0000135294; 10) 65071193 UDI-D:10193489060904(each), 20193489060901(case), Lot Number:0000130171; 11) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000135606; 12) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000134729; 13) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000137115; 14) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000130833; 15) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000131669; 16) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000129513; 17) 600604422 UDI-D:10193489070224(each), 20193489070221(case), Lot Number:0000128721; 18) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000129523; 19) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000132393; 20) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000134940; 21) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000144158; 22) 60100045 UDI-D:10193489044928(each), 30193489044922(case), Lot Number:0000139082; 23) 650801715 UDI-D:10193489075458(each), 20193489075455(case), Lot Number:0000130138; 24) 650801715 UDI-D:10193489075458(each), 20193489075455(case), Lot Number:0000132350; 25) 650801715 UDI-D:10193489075458(each), 20193489075455(case), Lot Number:0000135736; 26) 60232481 UDI-D:10193489053326(each), 30193489053320(case), Lot Number:0000129534; 27) 60120583 UDI-D:10193489045864(each), 30193489045868(case), Lot Number:0000133660; 28) 60120583 UDI-D:10193489045864(each), 30193489045868(case), Lot Number:0000136501; 29) 60120583 UDI-D:10193489045864(each), 30193489045868(case), Lot Number:0000131600; 30) 601403920 UDI-D:10193489071498(each), 30193489071492(case), Lot Number:0000131376; 31) 601403920 UDI-D:10193489071498(each), 30193489071492(case), Lot Number:0000130940; 32) 62140602 UDI-D:10193489057805(each), 30193489057809(case), Lot Number:0000132222; 33) 62140602 UDI-D:10193489057805(each), 30193489057809(case), Lot Number:0000141215; 34) 602301614 UDI-D:10193489072754(each), 30193489072758(case), Lot Number:0000133659; 35) 6010771601 UDI-D:10193489099843(each), 30193489099847(case), Lot Number:0000127859; 36) 600602914 UDI-D:10193489070170(each), 30193489070174(case), Lot Number:0000135717; 37) 600602914 UDI-D:10193489070170(each), 30193489070174(case), Lot Number:0000128447; 38) 601400111 UDI-D:10193489071344(each), 30193489071348(case), Lot Number:0000133121; 39) 601400111 UDI-D:10193489071344(each), 30193489071348(case), Lot Number:0000130573; 40) 600604423 UDI-D:10193489070231(each), 30193489070235(case), Lot Number:0000128607; 41) 600604423 UDI-D:10193489070231(each), 30193489070235(case), Lot Number:0000137751; 42) 60110552 UDI-D:10193489045611(each), 30193489045615(case), Lot Number:0000131878; 43) 60110552 UDI-D:10193489045611(each), 30193489045615(case), Lot Number:0000130577; 44) 60110552 UDI-D:10193489045611(each), 30193489045615(case), Lot Number:0000138084; 45) 60150035 UDI-D:10193489048292(each), 30193489048296(case), Lot Number:0000130187; 46) 60150035 UDI-D:10193489048292(each), 30193489048296(case), Lot Number:0000131883; 47) 60150035 UDI-D:10193489048292(each), 30193489048296(case), Lot Number:0000143095; 48) 60182723 UDI-D:10193489049817(each), 30193489049811(case), Lot Number:0000130988; 49) 602500214 UDI-D:10193489072815(each), 30193489072819(case), Lot Number:0000127874; 50) 601403921 UDI-D:30193489071508(each), 30193489071508(case), Lot Number:0000131619; 51) 601403921 UDI-D:30193489071508(each), 30193489071508(case), Lot Number:0000133025; 52) 601403921 UDI-D:30193489071508(each), 30193489071508(case), Lot Number:0000136559; 53) 601305210 UDI-D:10193489070989(each), 30193489070983(case), Lot Number:0000130984; 54) 601400110 UDI-D:10193489071337(each), 30193489071331(case), Lot Number:0000128757;
Recall started
Dec. 16, 2024
Voluntary or mandated
Voluntary: Firm initiated
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Under Investigation by firm
What the company told customers
An IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification letter dated 12/16/24 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. Quarantine all affected product immediately. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-24-270-FG Recall Code: 4. If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. Update: A follow-up letter titled IMMEDIATE ACTION REQUIRED dated 2/11/25 was sent to customers. Medline would like to update our action to instruct you to please destroy all impacted convenience kits and Medline will provide replacement kit(s) with the correct Fluid Delivery Set. REQUIRED ACTION: 1. Immediately check your stock for the affected kit number(s) and lot number(s) listed on the attachment. Please destroy all affected kits. 2. Please complete and return the enclosed destruction form listing the quantity of affected kits destroyed. Even if you do not have any affected product, please complete and return the form. 3. Upon receipt of your completed destruction form, Medline will issue replacem
Recall number
Z-0765-2025

This recall among the others

  • FDA's enforcement report came 37 days after the firm started the recall. For half of the 972 FDA medical device recall events reported in 2025 with a start date, it took 44 days or fewer (by year).
  • MEDLINE INDUSTRIES: 91 other recalls or notices on this site in the three years before this one (from Jan. 23, 2022).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement reports, recall numbers Z-0764-2025, Z-0765-2025. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-0764-2025, Jan. 22, 2025. "Medline Fluid Delivery Set with Drip Chamber: 1)… and 1 more product." PlainRecords, https://plainrecords.com/recall/device/z-0764-2025/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance