FDA · Medical device recall

BD Q-Syte… and 18 more products

Feb. 2, 2010

Date posted by the FDA
Feb. 2, 2010
Status
Terminated
Filed by
Becton Dickinson Medical Systems
Reason
Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
Distribution pattern
Nationwide distribution.

19 products in this recall

BD Q-Syte…

Product
BD Q-Syte, Luer Access Split-Septum, 0.10 ml, REF 385100, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
950,000 units
Codes and lots
Lot Numbers 8268863, 8269020, 8275798, 8308321, 8308323, 8309553, 8331937, 8331940, 8354558, 9007921, 9009643, 9009646, 9028838, 9035029, 99035032.
Recall started
Oct. 19, 2009
Device type
Set, Administration, Intravascular
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0750-2010

BD Q-Syte Extension Set…

Product
BD Q-Syte Extension Set, 15cm Std Bore, P.V. 1.14 ml Luer-Lok, REF 385101, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
42,600 units
Codes and lots
Lot numbers: 8305510, 8331944.
Recall started
Oct. 19, 2009
Device type
Set, Administration, Intravascular
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0751-2010

BD Q-Syte Extension Set…

Product
BD Q-Syte Extension Set, 15 cm, P.V. 0.21 ml, REF 385102, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
144,900 units
Codes and lots
Lot numbers: 8305511, 8308330, 8354561, 9012072
Recall started
Oct. 19, 2009
Device type
Set, Administration, Intravascular
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0752-2010

BD Q-Syte Vial Access Adapter…

Product
BD Q-Syte Vial Access Adapter, 0.16 ml, REF 385108, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
37,900 units
Codes and lots
Lot numbers: 9007918, 9035042, 9035044.
Recall started
Oct. 19, 2009
Device type
Set, Administration, Intravascular
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0753-2010

MPS Acacia Extension Set…

Product
MPS Acacia Extension Set, 15cm RightBore-18, Spin Nut w/BBD Q-Syte, REF 385150,
Quantity
73,050 units
Codes and lots
Lot numbers: A1967, A1979, A2056.
Recall started
Oct. 19, 2009
Device type
Set, Administration, Intravascular
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0754-2010

MPS Acacia Extension Set…

Product
MPS Acacia Extension Set, 15cm Small Bore, Spin Nut w/BD Q-Syte, REF 385151
Quantity
45,350 units
Codes and lots
Lot numbers: A1950, A2399
Recall started
Oct. 19, 2009
Device type
Set, Administration, Intravascular
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0755-2010

MPS Acacia Tri-Extension Set…

Product
MPS Acacia Tri-Extension Set, 15cm Small Bore, Spin Nut w/BD Q-Syte, REF 385164
Quantity
2,300 units
Codes and lots
Lot number: A1951
Recall started
Oct. 19, 2009
Device type
Set, Administration, Intravascular
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0756-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383530, 24GA 0.56IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
28,000 units
Codes and lots
Lot numbers: 8238450, 8319936, 8345957, 9013868.
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0829-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383531, 24GA 0.75IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
190,000 units
Codes and lots
Lot numbers: 8284049, 8284050, 8305438, 8308349, 8345951, 9013867, 9026722.
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0830-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383532, 22GA 1.00IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
827,280 units
Codes and lots
Lot numbers: 8248349, 8259950, 8295761, 8296792, 8296793, 8309299, 8309662, 8311150, 8319961, 8322609, 8331771, 8338737, 8338738, 8358628, 9020033.
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0831-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383533, 20GA 1.00IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
47,200 units
Codes and lots
Lot numbers: 8291980, 8330279.
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0832-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383534, 20GA 1.25IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
8960 units
Codes and lots
Lot number: 8331763.
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0833-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383536, 20GA 1.00IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
1,021,600 units
Codes and lots
Lot number: 8248355, 8249034, 8249035, 8269338, 8270576, 8274565, 8275207, 8275210, 8283964, 8311167, 8312211, 8319016, 8347056, 9007360, 9015539, 9022039.
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0834-2010

BD Nexiva HF Closed IV Catheter System…

Product
BD Nexiva HF Closed IV Catheter System, REF 383537, 20GA 1.25IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
180,160 units
Codes and lots
Lot numbers: 8263294, 8270579, 8347058, 9013864.
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0835-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383538, 20GA 1.75IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
5200 units
Codes and lots
Lot number: 8312220
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0836-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383539, 18GA 1.25IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
303,280 units
Codes and lots
Lot number: 8263293, 8274555, 8275429, 8290311, 8340308, 9016819.
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0837-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383540, 18GA 1.75IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
4960 units
Codes and lots
Lot number: 8340302
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0838-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383647, 20GA 1.25IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
21,680 units
Codes and lots
Lot Number: 9034319
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0839-2010

BD Nexiva Closed IV Catheter System…

Product
BD Nexiva Closed IV Catheter System, REF 383649, 18GA 1.25IN, Becton Dickinson Infusion Therapy Systems Inc.
Quantity
13,360 units
Codes and lots
Lot Number: 9041552
Recall started
Oct. 19, 2009
Device type
Catheter, Intravascular, Therapeutic, Short-Term Less Than 30 Days
Root cause (FDA)
Process control
What the company told customers
Consignees were notified by B-D Urgent Medical Device Recall letter on 10/28/2009 and asked to return all unused affected lots. For information, call the company at 800-453-4538, option 2, extension 2341. Another letter was sent to all affected customers on 02/08/2010 notifying them of the extension of the recall to include Nexiva products. A response card was included to be returned whether or not product was on hand. B-D issued a press release on Febriuary 8, 2010. It can be found on www.bd.com at http://www.bd.com/contentmanager/b_article.asp?Item_ID=24486&ContentType_ID=1&BusinessCode=20001&d=BD+Worldwide&s=&dTitle=&dc=&dcTitle=
Recall number
Z-0840-2010

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0750-2010, Z-0751-2010, Z-0752-2010, Z-0753-2010, Z-0754-2010, Z-0755-2010, Z-0756-2010, Z-0829-2010 and others. The agency record governs; check it and the company before acting.