AQURE System…

Feb. 10, 2016 Class II

Date
Feb. 10, 2016
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Radiometer America Inc
Product
AQURE System; Model Number: 933-599. The AQURE System manages blood gas and immunoassay analyzers.
Reason
The AQURE System has a design error regarding sample type in which sample type may be specified for some results and left blank for other results. The fact that the sample type field is left blank in some cases could lead to misinterpretation of results.
Distribution pattern
Worldwide Distribution-US (nationwide) including the states of PA, GA, MI, OH, VA, FL, AL, TX, OR, OK, TN, NY, and MA, and the countries of Austria, Czech Republic, Denmark, Estonia, United Emirates, Kuwait, Qatar, Saudi Arabia, Oman, Jordan, Finland, France, Germany, Hungry, Ireland, United Kingdom, Italy, Netherlands, Norway, Poland, Spain, Sweden, Switzerland, Australia, New Zealand, China, Hong Kong, Taiwan, Japan, Russia, Singapore, South Africa, Republic of Korea, and Malaysia.
Quantity
375
Codes and lots
Model Number: 933-599 Version 1.8.3
Recall started
Dec. 29, 2015
Device type
Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Root cause (FDA)
Device Design
What the company told customers
Customers were sent a notification letter on 12/29/15. The notification states that the AQURE System can be used, but to eliminate the potential confusion, blank fields for sample type are to be interpreted as not specified. Customers are told that a Radiometer representative will be in contact to schedule a visit or a remote session. During the visit or remote session the representative will run a database script on the AQURE system which will serve for two purposes: 1. Update existing patient results where the sample type field is blank to include Not specified. 2. Install a so-called database-trigger, which will include the text Not specified in a blank sample type field for future patient results. An upgraded version of the AQURE software will include modified device drivers to ensure that Not specified is inserted if the device transmits an empty field for sample type. The new software version will be installed by local engineer when available.
Recall number
Z-0748-2016

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0748-2016. The agency record governs; check it and the company before acting.