FDA · Medical device recall
HiSpeed LXI (2200997) and 23 more products
- Date posted by the FDA
- Jan. 4, 2011
- Status
- Terminated
- Filed by
- GE Healthcare, LLC
- Reason
- A defect with the hardware and software displaying the incorrect dose data.
- Distribution pattern
- Nationwide Distribution
24 products in this recall
HiSpeed LXI (2200997)
- Product
- HiSpeed LXI (2200997)
- Quantity
- 102 total in US
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0744-2011
HiSpeed LX/l (2200997-2)
- Product
- HiSpeed LX/l (2200997-2)
- Quantity
- 102 total in US
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0745-2011
HiSpeed DX/I (2201000)
- Product
- HiSpeed DX/I (2201000)
- Quantity
- 102 total in US
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0746-2011
HiSpeed DX/I (2201000-2)
- Product
- HiSpeed DX/I (2201000-2)
- Quantity
- 102
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0747-2011
HiSpeed DX/I (2249696)
- Product
- HiSpeed DX/I (2249696)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0768-2011
Hi Speed DX/I (2249696-2)
- Product
- Hi Speed DX/I (2249696-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0770-2011
HiSpeed FX/I (2200997 with Gantry mo…
- Product
- HiSpeed FX/I (2200997 with Gantry mo. #2200997)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0771-2011
HiSpeed FX/I (2200997 with Gantry mo…
- Product
- HiSpeed FX/I (2200997 with Gantry mo. #2201000)
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0772-2011
HiSpeed FX/I (2200997 with Gantry mo…
- Product
- HiSpeed FX/I (2200997 with Gantry mo. #2200997-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0773-2011
HiSpeed FX/I (2200997 with Gantry mo…
- Product
- HiSpeed FX/I (2200997 with Gantry mo. #2201000-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0774-2011
HiSpeed FX/I (2200997 with Gantry mo…
- Product
- HiSpeed FX/I (2200997 with Gantry mo. #2249696)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0775-2011
HiSpeed FX/I (2200997 with Gantry mo…
- Product
- HiSpeed FX/I (2200997 with Gantry mo. #2249696-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0776-2011
HiSpeed FX/I (2200997-2 with Gantry mo…
- Product
- HiSpeed FX/I (2200997-2 with Gantry mo. #2200997)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0777-2011
HiSpeed FX/I (2200997-2 with Gantry mo…
- Product
- HiSpeed FX/I (2200997-2 with Gantry mo. #2200997-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0778-2011
HiSpeed FX/I (2200997-2 with Gantry mo…
- Product
- HiSpeed FX/I (2200997-2 with Gantry mo. #2201000)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0779-2011
HiSpeed FX/I (2200997-2 with Gantry mo…
- Product
- HiSpeed FX/I (2200997-2 with Gantry mo. #2201000-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0780-2011
HiSpeed FX/I (2200997-2 with Gantry mo…
- Product
- HiSpeed FX/I (2200997-2 with Gantry mo. #2249696)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0781-2011
HiSpeed FX/I (2200997-2 with Gantry mo…
- Product
- HiSpeed FX/I (2200997-2 with Gantry mo. #2249696-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0782-2011
HiSpeed FX/I (2201000-2 with Gantry mo…
- Product
- HiSpeed FX/I (2201000-2 with Gantry mo. #2200997)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0783-2011
HiSpeed FX/I (2201000-2 with Gantry mo…
- Product
- HiSpeed FX/I (2201000-2 with Gantry mo. #2200997-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0784-2011
HiSpeed FX/I (2201000-2 with Gantry mo…
- Product
- HiSpeed FX/I (2201000-2 with Gantry mo. #2201000)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0785-2011
HiSpeed FX/I (2201000-2 with Gantry mo…
- Product
- HiSpeed FX/I (2201000-2 with Gantry mo. #2201000-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0786-2011
HiSpeed FX/I (2201000-2 with Gantry mo…
- Product
- HiSpeed FX/I (2201000-2 with Gantry mo. #2249696)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0787-2011
HiSpeed FX/I (2201000-2 with Gantry mo…
- Product
- HiSpeed FX/I (2201000-2 with Gantry mo. #2249696-2)
- Quantity
- 102 total
- Codes and lots
- With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
- Recall started
- Feb. 1, 2010
- Device type
- System, X-Ray, Tomography, Computed
- Root cause (FDA)
- Radiation Control for Health and Safety Act
- What the company told customers
- Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
- Recall number
- Z-0788-2011
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-0744-2011, Z-0745-2011, Z-0746-2011, Z-0747-2011, Z-0768-2011, Z-0770-2011, Z-0771-2011, Z-0772-2011 and others. The agency record governs; check it and the company before acting.