HiSpeed LXI (2200997) and 23 more products

Jan. 4, 2011

Date posted by the FDA
Jan. 4, 2011
Status
Terminated
Filed by
GE Healthcare, LLC
Reason
A defect with the hardware and software displaying the incorrect dose data.
Distribution pattern
Nationwide Distribution

24 products in this recall

HiSpeed LXI (2200997)

Product
HiSpeed LXI (2200997)
Quantity
102 total in US
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0744-2011

HiSpeed LX/l (2200997-2)

Product
HiSpeed LX/l (2200997-2)
Quantity
102 total in US
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0745-2011

HiSpeed DX/I (2201000)

Product
HiSpeed DX/I (2201000)
Quantity
102 total in US
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0746-2011

HiSpeed DX/I (2201000-2)

Product
HiSpeed DX/I (2201000-2)
Quantity
102
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0747-2011

HiSpeed DX/I (2249696)

Product
HiSpeed DX/I (2249696)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0768-2011

Hi Speed DX/I (2249696-2)

Product
Hi Speed DX/I (2249696-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0770-2011

HiSpeed FX/I (2200997 with Gantry mo…

Product
HiSpeed FX/I (2200997 with Gantry mo. #2200997)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0771-2011

HiSpeed FX/I (2200997 with Gantry mo…

Product
HiSpeed FX/I (2200997 with Gantry mo. #2201000)
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0772-2011

HiSpeed FX/I (2200997 with Gantry mo…

Product
HiSpeed FX/I (2200997 with Gantry mo. #2200997-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0773-2011

HiSpeed FX/I (2200997 with Gantry mo…

Product
HiSpeed FX/I (2200997 with Gantry mo. #2201000-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0774-2011

HiSpeed FX/I (2200997 with Gantry mo…

Product
HiSpeed FX/I (2200997 with Gantry mo. #2249696)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0775-2011

HiSpeed FX/I (2200997 with Gantry mo…

Product
HiSpeed FX/I (2200997 with Gantry mo. #2249696-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0776-2011

HiSpeed FX/I (2200997-2 with Gantry mo…

Product
HiSpeed FX/I (2200997-2 with Gantry mo. #2200997)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0777-2011

HiSpeed FX/I (2200997-2 with Gantry mo…

Product
HiSpeed FX/I (2200997-2 with Gantry mo. #2200997-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0778-2011

HiSpeed FX/I (2200997-2 with Gantry mo…

Product
HiSpeed FX/I (2200997-2 with Gantry mo. #2201000)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0779-2011

HiSpeed FX/I (2200997-2 with Gantry mo…

Product
HiSpeed FX/I (2200997-2 with Gantry mo. #2201000-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0780-2011

HiSpeed FX/I (2200997-2 with Gantry mo…

Product
HiSpeed FX/I (2200997-2 with Gantry mo. #2249696)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0781-2011

HiSpeed FX/I (2200997-2 with Gantry mo…

Product
HiSpeed FX/I (2200997-2 with Gantry mo. #2249696-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0782-2011

HiSpeed FX/I (2201000-2 with Gantry mo…

Product
HiSpeed FX/I (2201000-2 with Gantry mo. #2200997)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0783-2011

HiSpeed FX/I (2201000-2 with Gantry mo…

Product
HiSpeed FX/I (2201000-2 with Gantry mo. #2200997-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0784-2011

HiSpeed FX/I (2201000-2 with Gantry mo…

Product
HiSpeed FX/I (2201000-2 with Gantry mo. #2201000)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0785-2011

HiSpeed FX/I (2201000-2 with Gantry mo…

Product
HiSpeed FX/I (2201000-2 with Gantry mo. #2201000-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0786-2011

HiSpeed FX/I (2201000-2 with Gantry mo…

Product
HiSpeed FX/I (2201000-2 with Gantry mo. #2249696)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0787-2011

HiSpeed FX/I (2201000-2 with Gantry mo…

Product
HiSpeed FX/I (2201000-2 with Gantry mo. #2249696-2)
Quantity
102 total
Codes and lots
With Gantry model numbers 2200997, 2200997-2, 2201000, 2201000-2, 2249696 and 2249696-2.
Recall started
Feb. 1, 2010
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
Important Electronic Product Radiation Warning letter was sent to customers identifying the affected product, explaining the defect and related hazards, giving instructions, and the product correction. Customers are to decrease the CTDI value by a factor of 30% when using 5, 7, 10 mm slice thickness, and by a factor of 20% when using 3 mm slice thickness for more accurate estimate of the CTDI. GE Healthcare will remedy the defect free of charge and provide an updated Operator Manual and configuration update to correct the CTDI display to all affected systems. Questions should be directed to the contact numbers provided in the letter.
Recall number
Z-0788-2011

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0744-2011, Z-0745-2011, Z-0746-2011, Z-0747-2011, Z-0768-2011, Z-0770-2011, Z-0771-2011, Z-0772-2011 and others. The agency record governs; check it and the company before acting.