FDA · Medical device recall
VITROS Chemistry Products Calibrator Kit 16…
- Date posted by the FDA
- Jan. 11, 2012
- Status
- Terminated
- Filed by
- Ortho-Clinical Diagnostics
- Product
- VITROS Chemistry Products Calibrator Kit 16, REF 680 1700, 5x1 mL --- Ortho-Clinical Diagnostics, Rochester, NY USA -- Ortho-Clinical Diagnostics High Wycomber, UK. VITROS Chemistry Products Calibrator Kit 16 is used in conjunction with VITROS Chemistry Products FS Calibrator 1 to calibrate the VITROS 5,1 FS Chemistry System, the VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the quantitative measurement of rheumatoid factor (RF).
- Reason
- Ortho Clinical Diagnostics is recalling Lot 1630 of VITROS Calibrator Kit 16 because it will not likely meet its specified 12 months of stability.
- Distribution pattern
- Class II Recall: Worldwide Distribution - USA (Nationwide) and the countries of Brazil, Canada, Chile, China, Colombia, England, France, Mexico, Panama, Singapore, Spain, Venezuela.
- Quantity
- 768 kits
- Codes and lots
- Lot 1630
- Recall started
- Nov. 10, 2011
- Device type
- System, Test, Rheumatoid Factor
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- A letter was sent to customers via overnight delivery on 11/10/2011. The letter informed customers they should discontinue using and discard all remaining VITROS Calibrator Kit 16, Lot 1630; Complete and return the attached Confirmation of Receipt form upon receipt of the notification. Once the fax reply is received, the firm will begin the replacement order process; forward the notification to all departments and areas that use the VITROS Calibrator Kit 16, Lot 1630; and post the notification by each VITROS Chemistry or Integrated System in your facility that uses the VITROS Calibrator Kit 16, Lot 1630 or with VITROS user documentation.
- Recall number
- Z-0714-2012
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0714-2012. The agency record governs; check it and the company before acting.