Menu
FDA · Medical device recall

QMS Tacrolimus Calibrators…

Jan. 17, 2024 Class II

Date
Jan. 17, 2024
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Ortho-Clinical Diagnostics, Inc.
Product
QMS Tacrolimus Calibrators, Product code 10015573
Reason
Ortho-Clinical Diagnostics, Inc. is recalling QMS Tacrolimus Calibrators, Lot 74859619 expiry date: May 11, 2024, due to improper storage that is expected to degrade the quality of the calibrator and render it unsuitable for use. Use of the affected lot may results in erroneous calibration results, causing delay in testing.
Distribution pattern
Domestic distribution to IL, IN, FL, NJ, NM, TX. International distribution to Chile.
Quantity
10 sales units US; 1 sales unit OUS
Codes and lots
No UDI Lot 74859619, Expiry 11-May-2024
Recall started
Nov. 21, 2023
Device type
Calibrator, Secondary
Root cause (FDA)
Storage
What the company told customers
On 21 November 2023, all consignees who were shipped the VITROS QMS Tacrolimus Calibrators Lot 74859619 expiry date: 11-May-2024 from QuidelOrtho distribution center were notified of the issue and instructed to: -Discontinue using, render unusable, and discard your inventory of QMS Tacrolimus Calibrators, Lot 74859619, which was shipped to the laboratory from QuidelOrtho. -Post the notification by the laboratory s inventory storage area until a replacement lot of QMS Tacrolimus Calibrator has been received. -Forward the notification if the affected product was distributed outside their facility. If you have questions about this recall, please contact QuidelOrtho's Global Services Organization (formerly Ortho Care) at 1-800-421-3311.
Recall number
Z-0712-2024

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0712-2024. The agency record governs; check it and the company before acting.