Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Hook… and 8 more products

Jan. 27, 2009

Date posted by the FDA
Jan. 27, 2009
Status
Terminated
Filed by
Ethicon Endo-Surgery Inc
Reason
Nickel exposure: The stainless steel tips on the affected product were degrading, causing a small amount of nickel to be released, which may lead to health risks in some patients, particularly those with a known or suspected nickel allergy or sensitivity. Ethicon Endo-Surgery determined the presence of free nickel in the affected product could elicit a reaction in those patients with nickel sens
Distribution pattern
Worldwide distribution: USA, PUERTO RICO, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, DOMINICAN REPUBLIC, EGYPT, ENGLAND, EQUADOR, GERMANY, GREECE, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, KOREA, MALAYSIA, MEXICO, PAKISTAN, PANAMA, POLAND, REPUBLIC OF YEMEN, RUSSIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD, TURKEY, UNITED ARAB EMIRATES, URUGUAY, and VENEZUELA.

9 products in this recall

Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Hook Probe (Product Code EPS01)

Product
Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Hook Probe (Product Code EPS01)
Quantity
40 lots totalling 27,480 pieces
Codes and lots
Product Code# EPS01 - Lot # E4K594, Exp Date 2012-12; Lot # E4K67R, Exp Date 2012-12; Lot # E4K93J, Exp Date 2012-12; Lot # E4KC5J, Exp Date 2013-01; Lot # E4KF21, Exp Date 2013-01; Lot # E4KK1M, Exp Date 2013-02; Lot # E4KK7Z, Exp Date 2013-02; Lot # E4KL9T, Exp Date 2013-02; Lot # E4KM4A, Exp Date 2013-02; Lot # E4KP38, Exp Date 2013-02; Lot # E4KT81, Exp Date 2013-03; Lot # E4KU2X, Exp Date 2013-03; Lot # E4KX2D, Exp Date 2013-03; Lot # E4L01C, Exp Date 2013-04; Lot # E4L02M, Exp Date 2013-04; Lot # E4L21T, Exp Date 2013-04; Lot # E4L66N, Exp Date 2013-04; Lot # E4L738, Exp Date 2013-05; Lot # E4L82C, Exp Date 2013-05; Lot # E4LA6P, Exp Date 2013-05; Lot # E4LC88, Exp Date 2013-05; Lot # E4LE62, Exp Date 2013-05; Lot # E4LG4C, Exp Date 2013-06; Lot # E4LJ8W, Exp Date 2013-06; Lot # E4LK99, Exp Date 2013-06; Lot # E4LM57, Exp Date 2013-06; Lot # E4LP5G, Exp Date 2013-07; Lot # E4LU30, Exp Date 2013-07; Lot # E4LV48, Exp Date 2013-07; Lot # E4LX91, Exp Date 2013-07; Lot # E4M068, Exp Date 2013-08; Lot # E4M12K, Exp Date 2013-08; Lot # E4M23T, Exp Date 2013-08; Lot # E4M33W, Exp Date 2013-08; Lot # E4M387, Exp Date 2013-08; Lot # E4M43R, Exp Date 2013-08; Lot # E4M54H, Exp Date 2013-08; Lot # E4M75K, Exp Date 2013-09; Lot # E4M973, Exp Date 2013-09; and Lot # E4MA8R, Exp Date 2013-09.
Recall started
Dec. 9, 2008
Device type
Laparoscope, General & Plastic Surgery
Root cause (FDA)
Process design
What the company told customers
On 12/9/2008 Ethicon Endo-Surgery sent recall notification letters to their customers instructing them to stop using the product, examine all inventory to determine if they have the affected product on hand and remove the affected products from inventory. They requested that each customer complete the Business Reply Form and fax it back to Stericycle and also return any affected product to Stericycle.
Recall number
Z-0655-2009

Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Spatula Probe (Product Code EPS02)

Product
Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Spatula Probe (Product Code EPS02)
Quantity
32 lots, 18,528 pieces
Codes and lots
Product Code EPS02 - Lot # E4K181, Exp. Date 2012-12; Lot # E4K58C, Exp. Date 2012-12; Lot # E4KC5X, Exp. Date 2013-01; Lot # E4KD1P, Exp. Date 2013-01; Lot # E4KF2N, Exp. Date 2013-01; Lot # E4KG70, Exp. Date 2013-01; Lot # E4KK22, Exp. Date 2013-02; Lot # E4KP2F, Exp. Date 2013-02; Lot # E4KV17, Exp. Date 2013-03; Lot # E4KY1A, Exp. Date 2013-03; Lot # E4KY70, Exp. Date 2013-03; Lot # E4L03H, Exp. Date 2013-04; Lot # E4L519, Exp. Date 2013-04; Lot # E4L73Z, Exp. Date 2013-05; Lot # E4L916, Exp. Date 2013-05; Lot # E4LC3K, Exp. Date 2013-05; Lot # E4LD8L, Exp. Date 2013-05; Lot # E4LF60, Exp. Date 2013-05; Lot # E4LG4D, Exp. Date 2013-06; Lot # E4LL7P, Exp. Date 2013-06; Lot # E4LT6E, Exp. Date 2013-07; Lot # E4LX4H, Exp. Date 2013-07; Lot # E4M045, Exp. Date 2013-08; Lot # E4M12L, Exp. Date 2013-08; Lot # E4M27N, Exp. Date 2013-08; Lot # E4M466, Exp. Date 2013-08; Lot # E4M535, Exp. Date 2013-08; Lot # E4M54F, Exp. Date 2013-08; Lot # E4M69H, Exp. Date 2013-09; Lot # E4M838, Exp. Date 2013-09; Lot # E4MA4M, Exp. Date 2013-09; and Lot # E4MC3A, Exp. Date 2013-09
Recall started
Dec. 9, 2008
Device type
Laparoscope, General & Plastic Surgery
Root cause (FDA)
Process design
What the company told customers
On 12/9/2008 Ethicon Endo-Surgery sent recall notification letters to their customers instructing them to stop using the product, examine all inventory to determine if they have the affected product on hand and remove the affected products from inventory. They requested that each customer complete the Business Reply Form and fax it back to Stericycle and also return any affected product to Stericycle.
Recall number
Z-0656-2009

Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Right Angle Probe (Product Code EPS03)

Product
Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Right Angle Probe (Product Code EPS03)
Quantity
48 (labeled as LLC) lots + 4 (labeled as Inc.) lots, totaling 33,654 pieces
Codes and lots
Product Code EPS03 - Lot # E4K182, Exp. Date 2012-12; Lot # E4K59R, Exp. Date 2012-12; Lot # E4K67T, Exp. Date 2012-12; Lot # E4K81D, Exp. Date 2012-12; Lot # E4K938, Exp. Date 2012-12; Lot # D4JL20, Exp. Date 2012-09; Lot # D4JL21, Exp. Date 2012-09; Lot # D4JL22, Exp. Date 2012-09; Lot # E4KC1M, Exp. Date 2013-01; Lot # E4KF20, Exp. Date 2013-01; Lot # E4KK23, Exp. Date 2013-02; Lot # E4KK6V, Exp. Date 2013-02; Lot # E4KL0N, Exp. Date 2013-02; Lot # E4KR6Y, Exp. Date 2013-03; Lot # E4KV1X, Exp. Date 2013-03; Lot # E4KX2Z, Exp. Date 2013-03; Lot # E4KY5H, Exp. Date 2013-03; Lot # E4L18A, Exp. Date 2013-04; Lot # E4L404, Exp. Date 2013-04; Lot # E4L66M, Exp. Date 2013-04; Lot # E4L69A, Exp. Date 2013-05; Lot # E4L871, Exp. Date 2013-05; Lot # E4LA7Z, Exp. Date 2013-05; Lot # E4LD0T, Exp. Date 2013-05; Lot # E4LF5F, Exp. Date 2013-05; Lot # E4LJ46, Exp. Date 2013-06; Lot # E4LJ4F, Exp. Date 2013-06; Lot # E4LK9C, Exp. Date 2013-06; Lot # E4LN16, Exp. Date 2013-06; Lot # E4LP57, Exp. Date 2013-07; Lot # E4LR66, Exp. Date 2013-07; Lot # E4LT3Y, Exp. Date 2013-07; Lot # E4LU2Z, Exp. Date 2013-07; Lot # E4LV56, Exp. Date 2013-07; Lot # E4M00W, Exp. Date 2013-08; Lot # E4M01X, Exp. Date 2013-08; Lot # E4M120, Exp. Date 2013-08; Lot # E4M20L, Exp. Date 2013-08; Lot # E4M23R, Exp. Date 2013-08; Lot # E4M31A, Exp. Date 2013-08; Lot # E4M376, Exp. Date 2013-08; Lot # E4M39C, Exp. Date 2013-08; Lot # E4M572, Exp. Date 2013-08; Lot # E4M57U, Exp. Date 2013-08; Lot # E4M62M, Exp. Date 2013-08; Lot # E4M914, Exp. Date 2013-09; Lot # E4MA2R, Exp. Date 2013-09; and Lot # E4MD45, Exp. Date 2013-09
Recall started
Dec. 9, 2008
Device type
Laparoscope, General & Plastic Surgery
Root cause (FDA)
Process design
What the company told customers
On 12/9/2008 Ethicon Endo-Surgery sent recall notification letters to their customers instructing them to stop using the product, examine all inventory to determine if they have the affected product on hand and remove the affected products from inventory. They requested that each customer complete the Business Reply Form and fax it back to Stericycle and also return any affected product to Stericycle.
Recall number
Z-0657-2009

Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Curved Dissector Probe (Product Code EPS04)

Product
Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 34 cm Curved Dissector Probe (Product Code EPS04)
Quantity
13 lots, 2,304 pieces
Codes and lots
Product Code EPS04 - Lot # E4KC2J, Exp. Date 2013-01; Lot # E4KG8M, Exp. Date 2013-01; Lot # E4KP8H, Exp. Date 2013-02; Lot # E4KV0T, Exp. Date 2013-03; Lot # E4L030, Exp. Date 2013-04; Lot # E4L892, Exp. Date 2013-05; Lot # E4LD0R, Exp. Date 2013-05; Lot # E4LK7V, Exp. Date 2013-06; Lot # E4LX1D, Exp. Date 2013-07; Lot # E4M072, Exp. Date 2013-08; Lot # E4M28F, Exp. Date 2013-08; Lot # E4M91D, Exp. Date 2013-09; and Lot # E4MA2P, Exp. Date 2013-09
Recall started
Dec. 9, 2008
Device type
Laparoscope, General & Plastic Surgery
Root cause (FDA)
Process design
What the company told customers
On 12/9/2008 Ethicon Endo-Surgery sent recall notification letters to their customers instructing them to stop using the product, examine all inventory to determine if they have the affected product on hand and remove the affected products from inventory. They requested that each customer complete the Business Reply Form and fax it back to Stericycle and also return any affected product to Stericycle.
Recall number
Z-0658-2009

Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Hook Probe (Product Code EPS05)

Product
Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Hook Probe (Product Code EPS05)
Quantity
21 lots, 5,952 pieces
Codes and lots
Product Code EPS05 - Lot # E4K72Z, Exp. Date 2012-12; Lot # E4K92E, Exp. Date 2012-12; Lot # E4KE29, Exp. Date 2013-01; Lot # E4KH2L, Exp. Date 2013-01; Lot # E4KP2E, Exp. Date 2013-02; Lot # E4KU5K, Exp. Date 2013-03; Lot # E4KX1Y, Exp. Date 2013-03; Lot # E4L03G, Exp. Date 2013-04; Lot # E4L69Y, Exp. Date 2013-05; Lot # E4LA5X, Exp. Date 2013-05; Lot # E4LE64, Exp. Date 2013-05; Lot # E4LF6H, Exp. Date 2013-05; Lot # E4LN1K, Exp. Date 2013-06; Lot # E4LR3W, Exp. Date 2013-07; Lot # E4LU50, Exp. Date 2013-07; Lot # E4M00X, Exp. Date 2013-08; Lot # E4M24N, Exp. Date 2013-08; Lot # E4M506, Exp. Date 2013-08; Lot # E4M62L, Exp. Date 2013-08; Lot # E4M79W, Exp. Date 2013-09; and Lot # E4MA4P, Exp. Date 2013-09
Recall started
Dec. 9, 2008
Device type
Laparoscope, General & Plastic Surgery
Root cause (FDA)
Process design
What the company told customers
On 12/9/2008 Ethicon Endo-Surgery sent recall notification letters to their customers instructing them to stop using the product, examine all inventory to determine if they have the affected product on hand and remove the affected products from inventory. They requested that each customer complete the Business Reply Form and fax it back to Stericycle and also return any affected product to Stericycle.
Recall number
Z-0659-2009

Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Spatula Probe (Product Code EPS06)

Product
Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Spatula Probe (Product Code EPS06)
Quantity
30 lots, 11,352 pieces
Codes and lots
Product Code EPS06 - Lot # E4K730, Exp. Date 2012-12; Lot # E4K937, Exp. Date 2012-12; Lot # E4KE28, Exp. Date 2013-01; Lot # E4KF4E, Exp. Date 2013-01; Lot # E4KK3D, Exp. Date 2013-02; Lot # E4KK9Z, Exp. Date 2013-02; Lot # E4KT18, Exp. Date 2013-03; Lot # E4KU5A, Exp. Date 2013-03; Lot # E4KW7W, Exp. Date 2013-03; Lot # E4KY71, Exp. Date 2013-03; Lot # E4KZ5L, Exp. Date 2013-03; Lot # E4KZ9G, Exp. Date 2013-04; Lot # E4L189, Exp. Date 2013-04; Lot # E4L41U, Exp. Date 2013-04; Lot # E4L53G, Exp. Date 2013-04; Lot # E4L54M, Exp. Date 2013-04; Lot # E4L891, Exp. Date 2013-05; Lot # E4LA6F, Exp. Date 2013-05; Lot # E4LD1N, Exp. Date 2013-05; Lot # E4LF86, Exp. Date 2013-05; Lot # E4LJ4C, Exp. Date 2013-06; Lot # E4LN1J, Exp. Date 2013-06; Lot # E4LT6C, Exp. Date 2013-07; Lot # E4LU3P, Exp. Date 2013-07; Lot # E4M20H, Exp. Date 2013-08; Lot # E4M338, Exp. Date 2013-08; Lot # E4M457, Exp. Date 2013-08; Lot # E4M54G, Exp. Date 2013-08; Lot # E4M634, Exp. Date 2013-08; and Lot # E4MA4N, Exp. Date 2013-09
Recall started
Dec. 9, 2008
Device type
Laparoscope, General & Plastic Surgery
Root cause (FDA)
Process design
What the company told customers
On 12/9/2008 Ethicon Endo-Surgery sent recall notification letters to their customers instructing them to stop using the product, examine all inventory to determine if they have the affected product on hand and remove the affected products from inventory. They requested that each customer complete the Business Reply Form and fax it back to Stericycle and also return any affected product to Stericycle.
Recall number
Z-0660-2009

Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Right Angle Probe (Product Code EPS07)

Product
Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Right Angle Probe (Product Code EPS07)
Quantity
19 lots, 7,104 pieces
Codes and lots
Product Code EPS07 - Lot # E4KC1V, Exp Date 2013-01; Lot # E4KJ1J, Exp Date 2013-02; Lot # E4KM09, Exp Date 2013-02; Lot # E4KV1N, Exp Date 2013-03; Lot # E4KZ5M, Exp Date 2013-03; Lot # E4L464, Exp Date 2013-04; Lot # E4L664, Exp Date 2013-04; Lot # E4LA0W, Exp Date 2013-05; Lot # E4LC3M, Exp Date 2013-05; Lot # E4LE63, Exp Date 2013-05; Lot # E4LJ89, Exp Date 2013-06; Lot # E4LK9K, Exp Date 2013-06; Lot # E4LR3Y, Exp Date 2013-07; Lot # E4LX68, Exp Date 2013-07; Lot # E4LX88, Exp Date 2013-07; Lot # E4M071, Exp Date 2013-08; Lot # E4M43P, Exp Date 2013-08; Lot # E4M83K, Exp Date 2013-09; and Lot # E4M98Y, Exp Date 2013-09.
Recall started
Dec. 9, 2008
Device type
Laparoscope, General & Plastic Surgery
Root cause (FDA)
Process design
What the company told customers
On 12/9/2008 Ethicon Endo-Surgery sent recall notification letters to their customers instructing them to stop using the product, examine all inventory to determine if they have the affected product on hand and remove the affected products from inventory. They requested that each customer complete the Business Reply Form and fax it back to Stericycle and also return any affected product to Stericycle.
Recall number
Z-0661-2009

Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Curved Dissector Probe (Product Code EPS08)

Product
Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Curved Dissector Probe (Product Code EPS08)
Quantity
4 lots, 552 pieces
Codes and lots
Product Code EPS08 - Lot # E4KF22, Exp. Date 2013-01; Lot # E4L088, Exp. Date 2013-04; Lot # E4LR3X, Exp. Date 2013-07; and Lot # E4MD46, Exp. Date 2013-09
Recall started
Dec. 9, 2008
Device type
Laparoscope, General & Plastic Surgery
Root cause (FDA)
Process design
What the company told customers
On 12/9/2008 Ethicon Endo-Surgery sent recall notification letters to their customers instructing them to stop using the product, examine all inventory to determine if they have the affected product on hand and remove the affected products from inventory. They requested that each customer complete the Business Reply Form and fax it back to Stericycle and also return any affected product to Stericycle.
Recall number
Z-0662-2009

Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Needle Probe (Product Code EPS09)

Product
Ethicon Endo-Surgery ENDOPATH Electrosurgery Probe Plus II 29 cm Needle Probe (Product Code EPS09)
Quantity
13 lots, 2,448 pieces
Codes and lots
Product code EPS09 - Lot # E4KC2H, Exp. Date 2013-01; Lot # E4KJ1H, Exp. Date 2013-02; Lot # E4KR73, Exp. Date 2013-03; Lot # E4KW8R, Exp. Date 2013-03; Lot # E4L465, Exp. Date 2013-04; Lot # E4LA5T, Exp. Date 2013-05; Lot # E4LD9H, Exp. Date 2013-05; Lot # E4LF6M, Exp. Date 2013-05; Lot # E4LK7U, Exp. Date 2013-06; Lot # E4LU3N, Exp. Date 2013-07; Lot # E4M33D, Exp. Date 2013-08; Lot # E4M69J, Exp. Date 2013-09; and Lot # E4M83N, Exp. Date 2013-09.
Recall started
Dec. 9, 2008
Device type
Laparoscope, General & Plastic Surgery
Root cause (FDA)
Process design
What the company told customers
On 12/9/2008 Ethicon Endo-Surgery sent recall notification letters to their customers instructing them to stop using the product, examine all inventory to determine if they have the affected product on hand and remove the affected products from inventory. They requested that each customer complete the Business Reply Form and fax it back to Stericycle and also return any affected product to Stericycle.
Recall number
Z-0663-2009

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0655-2009, Z-0656-2009, Z-0657-2009, Z-0658-2009, Z-0659-2009, Z-0660-2009, Z-0661-2009, Z-0662-2009 and others. The agency record governs; check it and the company before acting.