FDA · Medical device recall
K-ACT Actalyke Clotting Test Tubes that are used in the: ACTALYKE XL ACTIVATED CLOTTING TIME ANALYZER…
- Date
- Dec. 23, 2020
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- Helena Laboratories, Corp.
- Product
- K-ACT Actalyke Clotting Test Tubes that are used in the: ACTALYKE XL ACTIVATED CLOTTING TIME ANALYZER, and ACTALYKE MINI ACTIVATED CLOTTING TIME TEST SYSTEM.
- Reason
- Due to complaints received associated with cracked/split test tube caps.
- Distribution pattern
- Worldwide distribution - U.S. Nationwide including in the states of NJ, NY, PA and TX. The countries of Austria, Canada, England, France, Italy, Portugal, Russia, Spain.
- Quantity
- Lot 1-19-K-ACT = 314 tubes and Lot 2-19-K-ACT = 297 tubes
- Codes and lots
- Catalog Number: K-ACT Lot Number: 1-19-K-ACT and 2-19-K-ACT
- Recall started
- Nov. 9, 2020
- Device type
- System, Multipurpose For In Vitro Coagulation Studies
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- On 11/09/2020, international distributors were initially contacted and provided with a notification for distributors letter. On 11/16/2020 domestic customers were contacted via telephone and then a formal notification was delivered. The Recalling Firm was informing its customers to the potential of defect associated with the (flip-top) caps for some test tubes where the cap my split resulting in leaking of biohazardous fluids as users are instructed to shake the tube for proper mixing. Customers are instructed to immediately examine their inventory and quarantine affected product. If customers have further distributed the affected product, they are asked to identify the users within their facility to notify them at once of this product recall. Customers are also asked to complete and return the response form as soon as possible. For any questions, customer can email or call the Recalling Firm as jhanka@Helena.com or 1-800-231-5663 extension 1177.
- Recall number
- Z-0635-2021
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0635-2021. The agency record governs; check it and the company before acting.