FDA · Medical device recall

Hoana LifeBed Patient Vigilance System…

Jan. 24, 2008

Date posted by the FDA
Jan. 24, 2008
Status
Terminated
Filed by
Hoana Medical
Product
Hoana LifeBed Patient Vigilance System, (Intelligent Medical Vigilance System),Model Number: 010135-01, Product distributed by Hoana Medical, Inc., 828 Fort St. Mall, Suite 620, Honolulu, HI
Reason
Bed Exit Failure: Failure to recognize "Bed Exit" when the feature is active due to a software anomaly.
Distribution pattern
Nationwide including states of HI, CA, TX and MA
Quantity
216 units
Codes and lots
All Serial Numbers
Recall started
Nov. 7, 2007
Device type
Monitor, Bed Patient
Root cause (FDA)
Software design
What the company told customers
On 11/7/07, all consignees were notified via an untitled letter and on-site visit by the firm's sales representative, informing them of the affected product and providing instructions on the recall. The recalling firm will arrange for installation of a software fix at each consignee.
Recall number
Z-0592-2008