FDA · Medical device recall
PROCARE B20 Patient Monitor… and 4 more products
- Date
- Dec. 19, 2018
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- GE Healthcare, LLC
- Reason
- Patient monitors may restart due to network overload caused by network configuration.
- Distribution pattern
- Worldwide Distribution: US (nationwide) including Puerto Rico; and countries of: Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bangladesh, Belgium, Bolivia, Bosnia-Herzegovina, Brazil, Brunei, Bulgaria, Burma (Myanmar), Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Ethiopia, Finland, France, Gabon, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Jamaica, Jordan, Kampuchea, Kazakstan, Kenya, Korea, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Mexico, Moldova, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, and Zambia.
5 products in this recall
PROCARE B20 Patient Monitor…
- Product
- PROCARE B20 Patient Monitor, V1; and B20 Patient Monitor, V2. The PROCARE Monitor B20 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intrahospital transport.
- Quantity
- 28,004 patient monitors
- Codes and lots
- All monitors
- Recall started
- Sept. 11, 2018
- Device type
- Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Root cause (FDA)
- Software design
- What the company told customers
- The recalling firm, GE Healthcare, issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 9/11/2018 via trackable mail on 9/11/2018 to its customers. The letters described the product, problem and actions to be taken. The letters informed the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Bx0 and/or B1x5 patient monitor on the network does not resume normal functionality in approximately 90 seconds. The customers were instructed to follow the safety instructions. GE Healthcare will provide revised network installation and user instructions documentation at no cost to you in a future mailing. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-0583-2019
PROCARE B40 Patient Monitor…
- Product
- PROCARE B40 Patient Monitor, V1; and B40 Patient Monitor, V2, V2.1, and V3. The PROCARE Monitor B40 is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
- Quantity
- 33,065 patient monitors
- Codes and lots
- All monitors
- Recall started
- Sept. 11, 2018
- Device type
- Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Root cause (FDA)
- Software design
- What the company told customers
- The recalling firm, GE Healthcare, issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 9/11/2018 via trackable mail on 9/11/2018 to its customers. The letters described the product, problem and actions to be taken. The letters informed the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Bx0 and/or B1x5 patient monitor on the network does not resume normal functionality in approximately 90 seconds. The customers were instructed to follow the safety instructions. GE Healthcare will provide revised network installation and user instructions documentation at no cost to you in a future mailing. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-0584-2019
B20i Patient Monitor and B20i V2 Patient Monitor…
- Product
- B20i Patient Monitor and B20i V2 Patient Monitor. The device is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
- Quantity
- 6,458 patient monitors
- Codes and lots
- All monitors
- Recall started
- Sept. 11, 2018
- Device type
- Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Root cause (FDA)
- Software design
- What the company told customers
- The recalling firm, GE Healthcare, issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 9/11/2018 via trackable mail on 9/11/2018 to its customers. The letters described the product, problem and actions to be taken. The letters informed the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Bx0 and/or B1x5 patient monitor on the network does not resume normal functionality in approximately 90 seconds. The customers were instructed to follow the safety instructions. GE Healthcare will provide revised network installation and user instructions documentation at no cost to you in a future mailing. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-0585-2019
B40i Patient Monitor and B40i Patient Monitor V3 The device is a portable multiparameter unit to be used for…
- Product
- B40i Patient Monitor and B40i Patient Monitor V3 The device is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
- Quantity
- 6,864 patient monitors
- Codes and lots
- All monitors
- Recall started
- Sept. 11, 2018
- Device type
- Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Root cause (FDA)
- Software design
- What the company told customers
- The recalling firm, GE Healthcare, issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 9/11/2018 via trackable mail on 9/11/2018 to its customers. The letters described the product, problem and actions to be taken. The letters informed the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Bx0 and/or B1x5 patient monitor on the network does not resume normal functionality in approximately 90 seconds. The customers were instructed to follow the safety instructions. GE Healthcare will provide revised network installation and user instructions documentation at no cost to you in a future mailing. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-0586-2019
B105/125 Patient Monitor…
- Product
- B105/125 Patient Monitor. The device is a portable multiparameter unit to be used for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adult, pediatric, and neonatal patients in a hospital environment and during intra-hospital transport.
- Quantity
- 5,516 patient monitors
- Codes and lots
- All monitors
- Recall started
- Sept. 11, 2018
- Device type
- Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Root cause (FDA)
- Software design
- What the company told customers
- The recalling firm, GE Healthcare, issued "URGENT MEDICAL DEVICE CORRECTION" letters dated 9/11/2018 via trackable mail on 9/11/2018 to its customers. The letters described the product, problem and actions to be taken. The letters informed the customer of the issue and providing safety instructions for mitigation of the issue if a prolonged network overload occurs and the Bx0 and/or B1x5 patient monitor on the network does not resume normal functionality in approximately 90 seconds. The customers were instructed to follow the safety instructions. GE Healthcare will provide revised network installation and user instructions documentation at no cost to you in a future mailing. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-0587-2019
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-0583-2019, Z-0584-2019, Z-0585-2019, Z-0586-2019, Z-0587-2019. The agency record governs; check it and the company before acting.