CNSystems TASK FORCE MONITOR 3040i…

Jan. 13, 2016 Class II

Date
Jan. 13, 2016
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Partners In Medicine Llc
Product
CNSystems TASK FORCE MONITOR 3040i, plethysmograph, impedance. Noninvasively measure and display patient's hemodynamic parameters time synchronized using Impedance Cardiography (ICG), Electrocardiography (ECG), oscillometric Blood Pressure (oscBP) and continuous Blood Pressure (contBP). Furthermore Sympathic Tone and Parasympathic Tone are calculated by power spectral analysis of ECG (heart rate variability) and continuous blood pressure (blood pressure variability) as well as the Baroreceptor Reflex Sensitivity (BRRS). For use by medically trained personnel. It has been designed for diagnoses aiding and must not be used for vital sign monitoring of critically ill patients under absence of additional, suitable self-monitoring devices or medical trained personnel. The device measures continuously the subject's hemodynamic parameters without reporting any diagnosis. It is the physician's responsibility to make proper judgment based on these parameters.
Reason
Device non-conformity which may lead to an incorrect oscillometric measurement of upper arm blood pressure values in some subjects - mainly in tall and slim persons (overestimation of systolic blood pressure, underestimation of diastolic blood pressure).
Distribution pattern
Nationwide Distribution including MI and CA.
Quantity
28 worldwide 2 monitors distributed in the U.S.
Codes and lots
Serial numbers: 003040i-2014-161-014-GG-0000, 003040i-2015-101-003-GG-0000, 003040i-2014-161-001-GG-0000, 003040i-2014-161-002-GG-0000, 003040i-2014-161-003-GG-0000, 003040i-2014-161-004-GG-0000, 003040i-2014-161-005-GG-0000, 003040i-2014-161-006-GG-0000, 003040i-2014-161-007-GG-0000, 003040i-2014-161-008-GG-0000, 003040i-2014-161-009-GG-0000, 003040i-2014-161-010-GG-0000, 003040i-2014-161-011-GG-0000, 003040i-2014-161-012-GG-0000, 003040i-2014-161-013-GG-0000, 003040i-2014-161-015-GG-0000, 003040i-2014-161-016-GG-0000, 003040i-2014-161-017-GG-0000, 003040i-2014-161-018-GG-0000, 003040i-2014-161-019-GG-0000, 003040i-2014-161-020-GG-0000, 003040i-2014-161-021-GG-0000, 003040i-2014-161-022-GG-0000, 003040i-2014-161-023-GG-0000, 003040i-2014-161-024-GG-0000, 003040i-2014-161-025-GG-0000, 003040i-2015-101-001-GG-0000, and 003040i-2015-101-002-GG-0000
Recall started
Dec. 1, 2015
Device type
Plethysmograph, Impedance
Root cause (FDA)
Mixed-up of materials/components
What the company told customers
CNSystems sent an Urgent Medical Device Recall Safety Notice dated December 1, 2015, to all affected customers. The notice identified the product, the problem, and the action to be taken by the customer. CNSystems decided to perform a voluntary recall of the affected devices and to return the devices after repair (free of charge). For questions regarding this recall call 314-324-2470.
Recall number
Z-0572-2016

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0572-2016. The agency record governs; check it and the company before acting.