Mac Vu360…

Dec. 16, 2020 Class II

Date
Dec. 16, 2020
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
MAC VU360, Electrocardiograph
Reason
Incorrect patient identification and/or patient demographic errors.
Distribution pattern
Domestic Distribution: AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI,MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. International Distribution: AUSTRALIA, AUSTRIA, BELGIUM, Canada, CZECH REPUBLIC, DENMARK, FINLAND, FRANCE, Germany, Hong Kong, IRELAND, ISRAEL, ITALY, Korea, KUWAIT, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, QATAR, SINGAPORE, SPAIN, SWEDEN, SWITZERLAND, THAILAND, UNITED ARAB EMIRATES, United Kingdom.
Quantity
5,284 units
Codes and lots
Model: 2030360 001
Recall started
Oct. 27, 2020
Device type
Electrocardiograph
Root cause (FDA)
Use error
What the company told customers
On 10/27/2020 the firm sent a letter to its consignees with the following instructions: You can continue to use your MAC VU360 system: At system setup: -Disable Auto-ECG in System Settings until the software has been upgraded. When performing ECGs: - Ensure the current patient lead wires are disconnected before selecting the Start New Patient workflow in the user interface. - Verify there are live scrolling waveforms on the screen before acquiring an ECG on the patient. - Verify the date and time of the ECG report corresponds to the patient for whom you are acquiring the ECG. - Always enter patient demographic data for each patient before acquiring an ECG GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the MAC VU360 software to be updated. Providing GE Healthcare with an email address in the attached customer response form, allows us to deliver this software update to you electronically and provide you with notifications of future software updates, as they become available. Note: After the MAC VU360 has been updated, discontinue usage of and destroy any media containing MAC VU360 V1.01 SP06 and any prior versions of MAC VU360 software. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
Recall number
Z-0543-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0543-2021. The agency record governs; check it and the company before acting.