Menu
FDA · Medical device recall

Platelet Aggregation Kit… and 1 more product

Dec. 4, 2019 Class III

Date
Dec. 4, 2019
Risk class
Class III (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences)
Status
Ongoing
Filed by
Helena Laboratories, Corp.
Distribution pattern
US:TX, WV, OH, TN, GA, TN, NY, ME, MA,LA, AZ OUS: Israel, Korea, Australia, Indonesia, Canada, Thailand, Colombia, Brasil , China , Taiwan

2 products in this recall

Platelet Aggregation Kit…

Product
Platelet Aggregation Kit, Cat. No.5369, For use in platelet aggregation studies, Contains: 2X1 mL ADP Reagent containing adenosine diphosphate, 2X 1 mL Collagen Reagent containing collagen (equine tendon), 2 x 1 mL Epinephrine Reagent containing L-epinephrine, FOR IN-VITRO DIAGNOSTIC USE, Store at 2 to 8 C, 0011724/14(3)
Reason
Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet Aggregation Kit may contain particulate matter or microbial growth.
Quantity
383 Collagen Kits
Codes and lots
Component Lot # 5-17-5368 (Collagen), Exp. 5/31/19 Kit Lot # 3-17-5369 (Aggregation kit containing Collagen), Exp. 11/30/18 Kit Lot # 5-17-5368 (Aggregation Kit containing Collagen), Exp. 05/31/2019
Recall started
May 23, 2018
Device type
Reagent, Platelet Aggregation
Root cause (FDA)
Unknown/Undetermined by firm
What the company told customers
On 05/23/2018, the firm's technical services contacted consignees via phone call with a email follow-up of their recall notification letter. Consignees were advised of the following: 1. Prevent any shipment or schedule demonstrations with these lots 2. Remove the lots from available inventory and discard COLLAGEN vials. 3. Verify amount REMOVED FROM STOCK 4. COMPLETE THE FORM AND RETURN TO HELENA LABORATORIES, USA VIA: fax: 409-842-1874 OR E-MAIL TO JJONES@HELENA.COM. For questions, contact Jessica W. Jones, MD, MPH - Mgr., Technical Services at: 1 (409) 842-3714 ext. 1177
Recall number
Z-0536-2020

Collagen Reagent…

Product
Collagen Reagent, Cat. No. 5368, 2 X 1 mL, For use in platelet aggregation studies, Contains: Collagen (equine tendon) For In-Vitro Diagnostic Use,
Reason
Through an investigation, it has been determined that vials of Collagen packaged within the Collagen Reagent and Platelet aggregation Kit, may contain particulate matter or microbial growth.
Quantity
64 Aggregation Kits
Codes and lots
Component Lot # 5-17-5368 (Collagen), Exp. 5/31/19 Kit Lot # 3-17-5369
Recall started
May 23, 2018
Device type
Reagent, Platelet Aggregation
Root cause (FDA)
Unknown/Undetermined by firm
What the company told customers
On 05/23/2018, the firm's technical services contacted consignees via phone call with a email follow-up of their recall notification letter. Consignees were advised of the following: 1. Prevent any shipment or schedule demonstrations with these lots 2. Remove the lots from available inventory and discard COLLAGEN vials. 3. Verify amount REMOVED FROM STOCK 4. COMPLETE THE FORM AND RETURN TO HELENA LABORATORIES, USA VIA: fax: 409-842-1874 OR E-MAIL TO JJONES@HELENA.COM. For questions, contact Jessica W. Jones, MD, MPH - Mgr., Technical Services at: 1 (409) 842-3714 ext. 1177
Recall number
Z-0537-2020

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0536-2020, Z-0537-2020. The agency record governs; check it and the company before acting.

Report a mistake on this page