The LightSpeed RT16 CT Scanner System is indicated for head and whole… and 1 more product

Feb. 15, 2007

Date posted by the FDA
Feb. 15, 2007
Status
Terminated
Filed by
GE Healthcare
Reason
When the manual cradle release feature on the table is used to move a patient out of the gantry, a discrepancy of up to 6mm between the numerical display on the gantry and the actual position of the table could be introduced.
Distribution pattern
Worldwide, including USA, Puerto Rico, Albania, Argentina, Austria, Australia, Belarus, Belgium, Bulgaria, Brazil, Canada, Chile, China, Columbia, Denmark, Finland, France, Germany, Great Britain, Greece, Guam, Hong Kong, Honduras, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Malaysia , Martinique, Mexico, Morocco, Netherlands, New Zealand, Norway, Panama, Philippines, Poland, Portugal, Reunion Island, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain , Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Venezuela, Vietnam, and Yugoslavia.

2 products in this recall

The LightSpeed RT16 CT Scanner System is indicated for head and whole body X-ray Computed Tomography…

Product
The LightSpeed RT16 CT Scanner System is indicated for head and whole body X-ray Computed Tomography applications. The system is capable of generating images for the guidance of minimally invasive procedures such as biopsy and ablation of tumors and pathology. The system allows imaging of obese patients, up to and including the obese population (BMI > 40). When used in the LightSpeed RT 16 configuration, the system acquires CT anatomical images that are clinically useful in the simulation and planning of radiation therapy for the treatment of cancer.
Quantity
22 units
Codes and lots
All serial numbers
Recall started
Dec. 8, 2006
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Other
What the company told customers
A Product Safety Advisory Notice, dated December 8, 2006, was issued to all customers informing them of the safety issue and production correction. Customers are advised to operate the CT cradle as intended with the system''s electronic controls. If a critical need arises to release the cradle during a radiation therapy simulation scan, they are advised to independently verify patient position prior to marking the patient for therapy to avoid the potential inaccuracy. A software update will be installed to correct this problem by a GE representative.
Recall number
Z-0506-2007

The LightSpeed VCT System is intended for head and whole body X-ray Computed Tomography applications

Product
The LightSpeed VCT System is intended for head and whole body X-ray Computed Tomography applications.
Quantity
1,256 units
Codes and lots
All serial numbers
Recall started
Dec. 8, 2006
Device type
System, X-Ray, Tomography, Computed
Root cause (FDA)
Other
What the company told customers
A Product Safety Advisory Notice, dated December 8, 2006, was issued to all customers informing them of the safety issue and production correction. Customers are advised to operate the CT cradle as intended with the system''s electronic controls. If a critical need arises to release the cradle during a radiation therapy simulation scan, they are advised to independently verify patient position prior to marking the patient for therapy to avoid the potential inaccuracy. A software update will be installed to correct this problem by a GE representative.
Recall number
Z-0507-2007

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0506-2007, Z-0507-2007. The agency record governs; check it and the company before acting.