FDA · Medical device recall

Stradis Healthcare… and 77 more products

Dec. 21, 2022 Class II

Date
Dec. 21, 2022
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Stradis Medical, LLC dba Stradis Healthcare
Reason
Medical and surgical procedure trays and kits underwent certain manufacturing conditions that may result in the outer bag being incompletely sealed, which may result in a breach in the sterility of the kit.
Distribution pattern
US nationwide distribution, and Canada.

78 products in this recall

Stradis Healthcare…

Product
STRADIS HEALTHCARE, WECKCELL EYE SPEARS, Item Numbers: a) 9430, b) 9431
Quantity
80 kits
Codes and lots
a) 9430: UDI/DI (case) M75294301, UDI/DI (kit)M75294301, Serial/Lot Numbers: 22270492023 b) 9431: UDI/DI (case) M75294301, UDI/DI (kit)M75294301, Serial/Lot Numbers: 22255491774, 22255491484
Recall started
Nov. 2, 2022
Device type
Sponge, Ophthalmic
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0481-2023

ACE Surgical Supply…

Product
ACE Surgical Supply, 17539 CUSTOM KIT STERILE Item No.003-3880, oral surgery tray
Quantity
10 kits
Codes and lots
UDI/DI (case) M75200338800, UDI/DI (kit)M75200338800, Serial/Lot Numbers: 22242490999
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0482-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Perio Surgery, Item No. 20177RPSS
Quantity
25 units
Codes and lots
UDI/DI (case) M75220177RPSS1, UDI/DI (kit)M75220177RPSS0, Serial/Lot Numbers: 22263491083
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0483-2023

Basic Implant Pack…

Product
BASIC IMPLANT PACK, Item No. 40310SBI, oral surgery tray
Quantity
80 units
Codes and lots
UDI/DI (case) M75240310SBI1, UDI/DI (kit)M75240310SBI0, Serial/Lot Numbers: 22241489853
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0484-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Surg Pack, Item No.40450SMS, dental
Quantity
12 units
Codes and lots
UDI/DI (case) M75240450SMS1, UDI/DI (kit)M75240450SMS0, Serial/Lot Numbers: 22262490676
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0485-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Surgical Pk, Item No.40510TWS, oral surgery tray
Quantity
18 units
Codes and lots
UDI/DI (case) M75240510TWS1, UDI/DI (kit)M75240510TWS0, Serial/Lot Numbers: 22236490637
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0486-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Surg, Item No.40551TMS. oral surgery tray
Quantity
6 units
Codes and lots
UDI/DI (case) M75240551TMS1, UDI/DI (kit)M75240551TMS0, Serial/Lot Numbers: 22230489112
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0487-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Basic Implant Pack, Item No.40561BIP, oral surgery tray
Quantity
20 units
Codes and lots
UDI/DI (case) M75240561BIP1, UDI/DI (kit)M75240561BIP0, Serial/Lot Numbers: 22228490663, 22242491516, 22242491516
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0488-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Surgical Pk, Item No.40649ECS, dental
Quantity
6 units
Codes and lots
UDI/DI (case) M75240649ECS1, UDI/DI (kit)M75240649ECS0, Serial/Lot Numbers: 22270491425
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0489-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Surg Pk, Item No.40714IES, oral surgery tray
Quantity
18 units
Codes and lots
UDI/DI (case) M75240714IES1, UDI/DI (kit)M75240714IES0, Serial/Lot Numbers: 22229490633, 22229490633, 22229490633
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0490-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, pack Item No.40788SMS, oral surgery tray
Quantity
6 units
Codes and lots
UDI/DI (case) M75240788SMS1, UDI/DI (kit)M75240788SMS0, Serial/Lot Numbers: 22234489435
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0491-2023

Surgical room…

Product
Surgical room, SURGICAL PACK, Item No.40996WLS, oral surgery tray
Quantity
180 units
Codes and lots
UDI/DI (case) M75240996WLS1, UDI/DI (kit)M75240996WLS0, Serial/Lot Numbers: 22230489371
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0492-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Implants, Item No.41015NSI, dental tray
Quantity
30 units
Codes and lots
UDI/DI (case) M75241015NSI1, UDI/DI (kit)M75241015NSI0, Serial/Lot Numbers: 22230491043
Recall started
Nov. 2, 2022
Device type
General Purpose Dental Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0493-2023

Surgical room…

Product
Surgical room, Implant Pack, Item No.41111UTI, oral surgery tray
Quantity
162 units
Codes and lots
UDI/DI (case) M7524111UTI0, UDI/DI (kit)M7524111UTI0, Serial/Lot Numbers: 22229491074
Recall started
Nov. 2, 2022
Device type
Endodontic Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0494-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, DENTAL CUSTOM PACK, Item No.41137TDS,
Quantity
20 units
Codes and lots
UDI/DI (case) M75241137TDS1, UDI/DI (kit)M75241137TDS0, Serial/Lot Numbers: 22230489097
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0495-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Custom Pack, Item No.41213ATCP, dental
Quantity
12 units
Codes and lots
UDI/DI (case) M75241213ATCP1, UDI/DI (kit)M75241213ATCP0, Serial/Lot Numbers: 22262491452
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0496-2023

Surgical room…

Product
Surgical room, DENTAL SURGICAL PACK, Item No.41236SDSP
Quantity
20 units
Codes and lots
UDI/DI (case) M75241236SDSP1, UDI/DI (kit)M75241236SDSP0, Serial/Lot Numbers: 22235489632
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0497-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, COTTON APPLICATOR 3" Item No.505-010,
Quantity
100 units
Codes and lots
UDI/DI (case) M7525050101, UDI/DI (kit)M7525050100, Serial/Lot Numbers: 22269491064,
Recall started
Nov. 2, 2022
Device type
Applicator, Absorbent Tipped, Sterile
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0498-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, SAFETY PINS, SMALL 1" Item No.515-001
Quantity
100 units
Codes and lots
UDI/DI (case) M7525150011, UDI/DI (kit)M7525150011, Serial/Lot Numbers: 22258491483
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0499-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Introducer Kit Micro, Item No.525-038, cardiovascular
Quantity
180 units
Codes and lots
UDI/DI (case) M7525250381, UDI/DI (kit)M7525250380, Serial/Lot Numbers: 22251491752,
Recall started
Nov. 2, 2022
Device type
Percutaneous Sheath Introducer Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0500-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, PROBE COVER W/GEL PACK, Item No.525-071, general surgery tray
Quantity
150 units
Codes and lots
UDI/DI (case) M7525250711, UDI/DI (kit)M7525250710, Serial/Lot Numbers: 22252491061
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0501-2023

Henry Schein…

Product
HENRY SCHEIN, Breast Augmentation tray, Item No.570-0611
Quantity
10 units
Codes and lots
UDI/DI (case) H65857006111, UDI/DI (kit)M75257006110, Serial/Lot Numbers: 22257490683
Recall started
Nov. 2, 2022
Device type
Plastic Surgery And Accessories Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0502-2023

Henry Schein…

Product
HENRY SCHEIN, Vein Clinics 1, Item No.570-1426
Quantity
800 units
Codes and lots
UDI/DI (case) H65857014261, UDI/DI (kit)M75257014260, Serial/Lot Numbers: 22257490682
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0503-2023

Henry Schein…

Product
HENRY SCHEIN, Novasure Pack, Item No.570-1462, general surgery tray
Quantity
15 units
Codes and lots
UDI/DI (case) H65857014621, UDI/DI (kit)M75257014620, Serial/Lot Numbers: 22230491082
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0504-2023

Henry Schein…

Product
HENRY SCHEIN, SURGICAL PAIN TRAY, Item No.570-1482
Quantity
20 units
Codes and lots
UDI/DI (case) M75257014821, UDI/DI (kit)M75257014820, Serial/Lot Numbers: 22235489736
Recall started
Nov. 2, 2022
Device type
Laparoscopic Cholecystectomy Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0505-2023

Henry Schein…

Product
HENRY SCHEIN, PROCEDURE KIT, Item No.570-1508, general surgery tray
Quantity
35 units
Codes and lots
UDI/DI (case) H65857015081, UDI/DI (kit)M75257015080, Serial/Lot Numbers: 22252490204
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0506-2023

Henry Schein…

Product
HENRY SCHEIN, Basin Pack, Item No.570-1692
Quantity
30 units
Codes and lots
UDI/DI (case) H65857016921, UDI/DI (kit)M75257016920, Serial/Lot Numbers: 22257490755
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0507-2023

Henry Schein…

Product
HENRY SCHEIN, Basin Set, Item No.570-1803
Quantity
96 units
Codes and lots
UDI/DI (case) H65857018031, UDI/DI (kit)M75257018030, Serial/Lot Numbers: 22258490734
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0508-2023

Henry Schein…

Product
HENRY SCHEIN, BREAST PACK, Item No.570-2397
Quantity
8 units
Codes and lots
UDI/DI (case) H65857023971, UDI/DI (kit)M75257023970, Serial/Lot Numbers: 22235489451
Recall started
Nov. 2, 2022
Device type
Plastic Surgery And Accessories Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0509-2023

Henry Schein…

Product
HENRY SCHEIN, GYN PACK, Item No.570-2625
Quantity
20 units
Codes and lots
UDI/DI (case) H65857026251, UDI/DI (kit)M75257026250, Serial/Lot Numbers: 22230491067
Recall started
Nov. 2, 2022
Device type
Vaginal Examination Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0510-2023

Henry Schein…

Product
HENRY SCHEIN, ENT PACK, Item No.570-2718
Quantity
104 units
Codes and lots
UDI/DI (case) H65857027181, UDI/DI (kit)M75257027180, Serial/Lot Numbers: 22242489582, 22256491864
Recall started
Nov. 2, 2022
Device type
Oral Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0511-2023

Henry Schein…

Product
HENRY SCHEIN, BASIC PACK, Item No.570-2719, UDI/DI (case) H65857027191, UDI/DI (kit)M75257027190, Serial/Lot Numbers: 22242489505
Quantity
24 units
Codes and lots
UDI/DI (case) H65857027181, UDI/DI (kit)M75257027180, Serial/Lot Numbers: 22242489582
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0512-2023

Henry Schein…

Product
HENRY SCHEIN, VEIN TRAY, Item No.570-2738,
Quantity
18 units
Codes and lots
UDI/DI (case) H65857027381, UDI/DI (kit)M75257027380, Serial/Lot Numbers: 22257490591
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0513-2023

Henry Schein…

Product
HENRY SCHEIN, ENT SET UP PACK, Item No.570-2742
Quantity
12 units
Codes and lots
UDI/DI (case) H65857027421, UDI/DI (kit)M75257027420, Serial/Lot Numbers: 22252489778
Recall started
Nov. 2, 2022
Device type
Ear, Nose, And Throat Surgical Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0514-2023

Henry Schein…

Product
HENRY SCHEIN, OMFS, Item No.570-2779,
Quantity
90 units
Codes and lots
UDI/DI (case) H65857027791, UDI/DI (kit)M75257027790, Serial/Lot Numbers: 22262489585, 22270490757
Recall started
Nov. 2, 2022
Device type
General Purpose Dental Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0515-2023

Henry Schein…

Product
HENRY SCHEIN, MAJOR EXTREMITY PACK, Item No.570-2783,
Quantity
5 units
Codes and lots
UDI/DI (case) H65857027831, UDI/DI (kit)M75257027830, Serial/Lot Numbers: 22234489305
Recall started
Nov. 2, 2022
Device type
Orthopedic Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0516-2023

Henry Schein…

Product
HENRY SCHEIN, UNIVERSAL TRAY, Item No.570-2787
Quantity
675 units
Codes and lots
UDI/DI (case) H65857027871, UDI/DI (kit)M75257027870, Serial/Lot Numbers: 22262491981
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0517-2023

Henry Schein…

Product
HENRY SCHEIN, CATH PACK Item No.570-2789
Quantity
100 units
Codes and lots
UDI/DI (case) H65857027891, UDI/DI (kit)M75257027890, Serial/Lot Numbers: 22252491561
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0518-2023

Henry Schein…

Product
HENRY SCHEIN, PACEMAKER PACK IItem No.570-2793
Quantity
18 units
Codes and lots
UDI/DI (case) H65857027931, UDI/DI (kit)M75257027930, Serial/Lot Numbers: 22241489575
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0519-2023

Henry Schein…

Product
HENRY SCHEIN, TEMP STIMULATOR SUPPLIES, Item No.570-2883
Quantity
24 units
Codes and lots
UDI/DI (case) H65857028831, UDI/DI (kit)M75257028831, Serial/Lot Numbers: 22229489323
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0520-2023

Henry Schein…

Product
HENRY SCHEIN, BASIC EXTENDED PACK, Item No.570-2890
Quantity
25 units
Codes and lots
UDI/DI (case) H65857028901, UDI/DI (kit)M75257028900, Serial/Lot Numbers: 22262492001
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0521-2023

Henry Schein…

Product
HENRY SCHEIN, LOOP RECORDER PACK, Item No.570-2891
Quantity
144 units
Codes and lots
UDI/DI (case) H65857028911, UDI/DI (kit)M75257028910, Serial/Lot Numbers: 22252491813
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0522-2023

Henry Schein…

Product
HENRY SCHEIN, PROCEDURE TRAY, Item No.570-2900,
Quantity
10 units
Codes and lots
UDI/DI (case) H65857029001, UDI/DI (kit)M75257029000, Serial/Lot Numbers: 22252489760
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0523-2023

Henry Schein…

Product
HENRY SCHEIN, Knee Pack, Item No.570-2908,
Quantity
48 units
Codes and lots
UDI/DI (case) H65857029081, UDI/DI (kit)M75257029080, Serial/Lot Numbers: 22241489773
Recall started
Nov. 2, 2022
Device type
Orthopedic Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0524-2023

Henry Schein…

Product
HENRY SCHEIN, CATH ANGIO PACK, Item No.570-2930
Quantity
16 units
Codes and lots
UDI/DI (case) H65857029301, UDI/DI (kit)M75257029300, Serial/Lot Numbers: 22235490920
Recall started
Nov. 2, 2022
Device type
Angiography/Angioplasty Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0525-2023

Henry Schein…

Product
HENRY SCHEIN, EXTREMITY PACK, Item No.570-2946
Quantity
144 units
Codes and lots
UDI/DI (case) H65857029461, UDI/DI (kit)M75257029460, Serial/Lot Numbers: 22242490745
Recall started
Nov. 2, 2022
Device type
Orthopedic Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0526-2023

Henry Schein…

Product
HENRY SCHEIN, Amnio Tray, Item No.570-3059,
Quantity
510 units
Codes and lots
UDI/DI (case) H65857030591, UDI/DI (kit)M75257030590, Serial/Lot Numbers: 22257487536
Recall started
Nov. 2, 2022
Device type
Sampler, Amniotic Fluid (Amniocentesis Tray)
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0527-2023

Henry Schein…

Product
HENRY SCHEIN, RHINO PACK, Item No.570-3070
Quantity
328 units
Codes and lots
UDI/DI (case) H65857030701, UDI/DI (kit)M75257030700, Serial/Lot Numbers: 22235491255, 22241491256
Recall started
Nov. 2, 2022
Device type
Plastic Surgery And Accessories Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0528-2023

Henry Schein…

Product
HENRY SCHEIN, CUSTOM PK 2, Item No.570-3072
Quantity
27 units
Codes and lots
UDI/DI (case) H65857030721, UDI/DI (kit)M75257030720, Serial/Lot Numbers: 22241490953
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0529-2023

Henry Schein…

Product
HENRY SCHEIN, SINGLE SHOT EPIDURAL, Item No.570-3089
Quantity
80 units
Codes and lots
UDI/DI (case) H65857030891, UDI/DI (kit)M75257030890, Serial/Lot Numbers: 22251491690
Recall started
Nov. 2, 2022
Device type
Epidural Anesthesia Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0530-2023

Henry Schein…

Product
HENRY SCHEIN, Blue Tail Custom Pack, Item No.570-3091,
Quantity
72 units
Codes and lots
UDI/DI (case) H65857030911, UDI/DI (kit)M75257030910, Serial/Lot Numbers: 22258490762
Recall started
Nov. 2, 2022
Device type
Bone Marrow Biopsy Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0531-2023

Henry Schein…

Product
HENRY SCHEIN, BONE MARROW KIT, Item No.570-3092
Quantity
581 units
Codes and lots
UDI/DI (case) H65857030921, UDI/DI (kit)M75257030920, Serial/Lot Numbers: 22231491193
Recall started
Nov. 2, 2022
Device type
Bone Marrow Biopsy Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0532-2023

Henry Schein…

Product
HENRY SCHEIN, Labiaplasty PackItem No.570-3118,
Quantity
42 units
Codes and lots
UDI/DI (case) H65857031181, UDI/DI (kit)M75257031180, Serial/Lot Numbers: 22234491292
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0533-2023

Henry Schein…

Product
HENRY SCHEIN, ILR Insertion Tray, Item No.570-3131
Quantity
140 units
Codes and lots
UDI/DI (case) H65857031311, UDI/DI (kit)M75257031310, Serial/Lot Numbers: 22257490596
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0534-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, CURVED TIP SYRINGE, Item No.60060GCTS,
Quantity
300 units
Codes and lots
UDI/DI (case) M75260060GCTS1, UDI/DI (kit)M75260060GCTS0, Serial/Lot Numbers: 22257490438
Recall started
Nov. 2, 2022
Device type
Syringe, Periodontic, Endodontic, Irrigating
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0535-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Lipo, Item No.681-254,
Quantity
6 units
Codes and lots
UDI/DI (case) M7526812541, UDI/DI (kit)M7526812540, Serial/Lot Numbers: 22270492010
Recall started
Nov. 2, 2022
Device type
Plastic Surgery And Accessories Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0536-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Phlebectomy Pack, Item No.682-001,
Quantity
15 units
Codes and lots
UDI/DI (case) M7526820011, UDI/DI (kit)M7526820010, Serial/Lot Numbers: 22236489633
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0537-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Thoracic Phleb, Item No.682-1228,
Quantity
5 units
Codes and lots
UDI/DI (case) M7268212281, UDI/DI (kit)M7268212280, Serial/Lot Numbers: 22237491041
Recall started
Nov. 2, 2022
Device type
Plastic Surgery And Accessories Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0538-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Ablation Pack, Item No.682-130,
Quantity
40 units
Codes and lots
UDI/DI (case) M7526821301, UDI/DI (kit)M7526821300, Serial/Lot Numbers: 22237491040
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0539-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Phleb, Item No.682-1425,
Quantity
20 units
Codes and lots
UDI/DI (case) M75268214251, UDI/DI (kit)M75268214250, Serial/Lot Numbers: 22236490624
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0540-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Vasc Vein, Item No.682-1583,
Quantity
20 units
Codes and lots
UDI/DI (case) M75268215831, UDI/DI (kit)M75268215830, Serial/Lot Numbers: 22237491038
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0541-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Surgery PK, Item No.682-1728,
Quantity
40 units
Codes and lots
UDI/DI (case) M75268217281, UDI/DI (kit)M75268217280, Serial/Lot Numbers: 22235489419
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0542-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Vein, Item No.682-1732,
Quantity
10 units
Codes and lots
UDI/DI (case) M75268217321, UDI/DI (kit)M75268217320, Serial/Lot Numbers: 22230489185
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0543-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Angioplasty, Item No.682-1937,
Quantity
100 units
Codes and lots
UDI/DI (case) M75268219371, UDI/DI (kit)M75268219370, Serial/Lot Numbers: 22237490509
Recall started
Nov. 2, 2022
Device type
Angiography/Angioplasty Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0544-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Vein RF Pack, Item No.682-1942NL1,
Quantity
400 units
Codes and lots
UDI/DI (case) M7526821942NL11, UDI/DI (kit)M7526821942NL10, Serial/Lot Numbers: 22251491808
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0545-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, VENOUS ACCESS PACK, Item No.682-1993R1,
Quantity
30 units
Codes and lots
UDI/DI (case) M7526821993R11, UDI/DI (kit)M7526821993R10, Serial/Lot Numbers: 22235491208
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0546-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Medical Vein, Item No.682-330,
Quantity
40 units
Codes and lots
UDI/DI (case) M7526823301, UDI/DI (kit)M7526823300, Serial/Lot Numbers: 22238490518
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0547-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Vein & Laser, Item No.682-334R1,
Quantity
10 units
Codes and lots
UDI/DI (case) M752682334R11, UDI/DI (kit)M752682334R10, Serial/Lot Numbers: 22231489619
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0548-2023

www.StradisHealthcare.com…

Product
www.StradisHealthcare.com, Endovenous Pack, Item No.682-494,
Quantity
10 units
Codes and lots
UDI/DI (case) M7526824941, UDI/DI (kit)M7526824940, Serial/Lot Numbers: 22234489198
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0549-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Endo Pack, Item No.682-509,
Quantity
25 units
Codes and lots
UDI/DI (case) M7526825091, UDI/DI (kit)M7526825090, Serial/Lot Numbers: 22256490827
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0550-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Vein, Item No.682-577,
Quantity
15 units
Codes and lots
UDI/DI (case) M7526825771, UDI/DI (kit)M7526585770, Serial/Lot Numbers: 22230490618
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0551-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Varicosity Basic Pack, Item No.682-947,
Quantity
105 units
Codes and lots
UDI/DI (case) M7526829471, UDI/DI (kit)M7526829470, Serial/Lot Numbers: 22236491032
Recall started
Nov. 2, 2022
Device type
Cardiovascular Procedure Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0552-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Femoral, Vasc, Item No.686-142,
Quantity
10
Codes and lots
UDI/DI (case) M7526861421, UDI/DI (kit)M7526861420, Serial/Lot Numbers: 22236491028
Recall started
Nov. 2, 2022
Device type
Angiography/Angioplasty Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0553-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Angio Pack, Item No.686-199,
Quantity
24 units
Codes and lots
UDI/DI (case) M7526861991, UDI/DI (kit)M7526861990, Serial/Lot Numbers: 22236491017
Recall started
Nov. 2, 2022
Device type
Angiography/Angioplasty Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0554-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, Breast, Item No.693-276,
Quantity
16 units
Codes and lots
UDI/DI (case) M7526932761, UDI/DI (kit)M7526932760, Serial/Lot Numbers: 22242490001
Recall started
Nov. 2, 2022
Device type
Plastic Surgery And Accessories Kit
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0555-2023

Henry Schein…

Product
HENRY SCHEIN, Vas Pack, Item No.888-2520,
Quantity
304 units
Codes and lots
UDI/DI (case) H65888825201, UDI/DI (kit)M75288825200, Serial/Lot Numbers: 22256490684
Recall started
Nov. 2, 2022
Device type
General Surgery Tray
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0556-2023

Henry Schein…

Product
HENRY SCHEIN, Vas Reversal Pack, Item No.888-2549,
Quantity
3 units
Codes and lots
UDI/DI (case) H65888825491, UDI/DI (kit)M75288825490, Serial/Lot Numbers: 22230489459
Recall started
Nov. 2, 2022
Device type
Vasovasostomy Set
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0557-2023

Stradis Healthcare…

Product
STRADIS HEALTHCARE, TIP, SURGE-O-VAC .125" GREEN, Item No.SS077225,
Quantity
75 units
Codes and lots
UDI/DI (case) M752SS0772251, UDI/DI (kit)M752SS0772251, Serial/Lot Numbers: 22255490439
Recall started
Nov. 2, 2022
Device type
Evacuator, Oral Cavity
Root cause (FDA)
Packaging process control
What the company told customers
Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
Recall number
Z-0558-2023

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0481-2023, Z-0482-2023, Z-0483-2023, Z-0484-2023, Z-0485-2023, Z-0486-2023, Z-0487-2023, Z-0488-2023 and others. The agency record governs; check it and the company before acting.