FDA · Medical device recall
Stradis Healthcare… and 77 more products
- Date
- Dec. 21, 2022
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- Stradis Medical, LLC dba Stradis Healthcare
- Reason
- Medical and surgical procedure trays and kits underwent certain manufacturing conditions that may result in the outer bag being incompletely sealed, which may result in a breach in the sterility of the kit.
- Distribution pattern
- US nationwide distribution, and Canada.
78 products in this recall
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, WECKCELL EYE SPEARS, Item Numbers: a) 9430, b) 9431
- Quantity
- 80 kits
- Codes and lots
- a) 9430: UDI/DI (case) M75294301, UDI/DI (kit)M75294301, Serial/Lot Numbers: 22270492023 b) 9431: UDI/DI (case) M75294301, UDI/DI (kit)M75294301, Serial/Lot Numbers: 22255491774, 22255491484
- Recall started
- Nov. 2, 2022
- Device type
- Sponge, Ophthalmic
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0481-2023
ACE Surgical Supply…
- Product
- ACE Surgical Supply, 17539 CUSTOM KIT STERILE Item No.003-3880, oral surgery tray
- Quantity
- 10 kits
- Codes and lots
- UDI/DI (case) M75200338800, UDI/DI (kit)M75200338800, Serial/Lot Numbers: 22242490999
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0482-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Perio Surgery, Item No. 20177RPSS
- Quantity
- 25 units
- Codes and lots
- UDI/DI (case) M75220177RPSS1, UDI/DI (kit)M75220177RPSS0, Serial/Lot Numbers: 22263491083
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0483-2023
Basic Implant Pack…
- Product
- BASIC IMPLANT PACK, Item No. 40310SBI, oral surgery tray
- Quantity
- 80 units
- Codes and lots
- UDI/DI (case) M75240310SBI1, UDI/DI (kit)M75240310SBI0, Serial/Lot Numbers: 22241489853
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0484-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Surg Pack, Item No.40450SMS, dental
- Quantity
- 12 units
- Codes and lots
- UDI/DI (case) M75240450SMS1, UDI/DI (kit)M75240450SMS0, Serial/Lot Numbers: 22262490676
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0485-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Surgical Pk, Item No.40510TWS, oral surgery tray
- Quantity
- 18 units
- Codes and lots
- UDI/DI (case) M75240510TWS1, UDI/DI (kit)M75240510TWS0, Serial/Lot Numbers: 22236490637
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0486-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Surg, Item No.40551TMS. oral surgery tray
- Quantity
- 6 units
- Codes and lots
- UDI/DI (case) M75240551TMS1, UDI/DI (kit)M75240551TMS0, Serial/Lot Numbers: 22230489112
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0487-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Basic Implant Pack, Item No.40561BIP, oral surgery tray
- Quantity
- 20 units
- Codes and lots
- UDI/DI (case) M75240561BIP1, UDI/DI (kit)M75240561BIP0, Serial/Lot Numbers: 22228490663, 22242491516, 22242491516
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0488-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Surgical Pk, Item No.40649ECS, dental
- Quantity
- 6 units
- Codes and lots
- UDI/DI (case) M75240649ECS1, UDI/DI (kit)M75240649ECS0, Serial/Lot Numbers: 22270491425
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0489-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Surg Pk, Item No.40714IES, oral surgery tray
- Quantity
- 18 units
- Codes and lots
- UDI/DI (case) M75240714IES1, UDI/DI (kit)M75240714IES0, Serial/Lot Numbers: 22229490633, 22229490633, 22229490633
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0490-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, pack Item No.40788SMS, oral surgery tray
- Quantity
- 6 units
- Codes and lots
- UDI/DI (case) M75240788SMS1, UDI/DI (kit)M75240788SMS0, Serial/Lot Numbers: 22234489435
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0491-2023
Surgical room…
- Product
- Surgical room, SURGICAL PACK, Item No.40996WLS, oral surgery tray
- Quantity
- 180 units
- Codes and lots
- UDI/DI (case) M75240996WLS1, UDI/DI (kit)M75240996WLS0, Serial/Lot Numbers: 22230489371
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0492-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Implants, Item No.41015NSI, dental tray
- Quantity
- 30 units
- Codes and lots
- UDI/DI (case) M75241015NSI1, UDI/DI (kit)M75241015NSI0, Serial/Lot Numbers: 22230491043
- Recall started
- Nov. 2, 2022
- Device type
- General Purpose Dental Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0493-2023
Surgical room…
- Product
- Surgical room, Implant Pack, Item No.41111UTI, oral surgery tray
- Quantity
- 162 units
- Codes and lots
- UDI/DI (case) M7524111UTI0, UDI/DI (kit)M7524111UTI0, Serial/Lot Numbers: 22229491074
- Recall started
- Nov. 2, 2022
- Device type
- Endodontic Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0494-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, DENTAL CUSTOM PACK, Item No.41137TDS,
- Quantity
- 20 units
- Codes and lots
- UDI/DI (case) M75241137TDS1, UDI/DI (kit)M75241137TDS0, Serial/Lot Numbers: 22230489097
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0495-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Custom Pack, Item No.41213ATCP, dental
- Quantity
- 12 units
- Codes and lots
- UDI/DI (case) M75241213ATCP1, UDI/DI (kit)M75241213ATCP0, Serial/Lot Numbers: 22262491452
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0496-2023
Surgical room…
- Product
- Surgical room, DENTAL SURGICAL PACK, Item No.41236SDSP
- Quantity
- 20 units
- Codes and lots
- UDI/DI (case) M75241236SDSP1, UDI/DI (kit)M75241236SDSP0, Serial/Lot Numbers: 22235489632
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0497-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, COTTON APPLICATOR 3" Item No.505-010,
- Quantity
- 100 units
- Codes and lots
- UDI/DI (case) M7525050101, UDI/DI (kit)M7525050100, Serial/Lot Numbers: 22269491064,
- Recall started
- Nov. 2, 2022
- Device type
- Applicator, Absorbent Tipped, Sterile
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0498-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, SAFETY PINS, SMALL 1" Item No.515-001
- Quantity
- 100 units
- Codes and lots
- UDI/DI (case) M7525150011, UDI/DI (kit)M7525150011, Serial/Lot Numbers: 22258491483
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0499-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Introducer Kit Micro, Item No.525-038, cardiovascular
- Quantity
- 180 units
- Codes and lots
- UDI/DI (case) M7525250381, UDI/DI (kit)M7525250380, Serial/Lot Numbers: 22251491752,
- Recall started
- Nov. 2, 2022
- Device type
- Percutaneous Sheath Introducer Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0500-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, PROBE COVER W/GEL PACK, Item No.525-071, general surgery tray
- Quantity
- 150 units
- Codes and lots
- UDI/DI (case) M7525250711, UDI/DI (kit)M7525250710, Serial/Lot Numbers: 22252491061
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0501-2023
Henry Schein…
- Product
- HENRY SCHEIN, Breast Augmentation tray, Item No.570-0611
- Quantity
- 10 units
- Codes and lots
- UDI/DI (case) H65857006111, UDI/DI (kit)M75257006110, Serial/Lot Numbers: 22257490683
- Recall started
- Nov. 2, 2022
- Device type
- Plastic Surgery And Accessories Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0502-2023
Henry Schein…
- Product
- HENRY SCHEIN, Vein Clinics 1, Item No.570-1426
- Quantity
- 800 units
- Codes and lots
- UDI/DI (case) H65857014261, UDI/DI (kit)M75257014260, Serial/Lot Numbers: 22257490682
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0503-2023
Henry Schein…
- Product
- HENRY SCHEIN, Novasure Pack, Item No.570-1462, general surgery tray
- Quantity
- 15 units
- Codes and lots
- UDI/DI (case) H65857014621, UDI/DI (kit)M75257014620, Serial/Lot Numbers: 22230491082
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0504-2023
Henry Schein…
- Product
- HENRY SCHEIN, SURGICAL PAIN TRAY, Item No.570-1482
- Quantity
- 20 units
- Codes and lots
- UDI/DI (case) M75257014821, UDI/DI (kit)M75257014820, Serial/Lot Numbers: 22235489736
- Recall started
- Nov. 2, 2022
- Device type
- Laparoscopic Cholecystectomy Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0505-2023
Henry Schein…
- Product
- HENRY SCHEIN, PROCEDURE KIT, Item No.570-1508, general surgery tray
- Quantity
- 35 units
- Codes and lots
- UDI/DI (case) H65857015081, UDI/DI (kit)M75257015080, Serial/Lot Numbers: 22252490204
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0506-2023
Henry Schein…
- Product
- HENRY SCHEIN, Basin Pack, Item No.570-1692
- Quantity
- 30 units
- Codes and lots
- UDI/DI (case) H65857016921, UDI/DI (kit)M75257016920, Serial/Lot Numbers: 22257490755
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0507-2023
Henry Schein…
- Product
- HENRY SCHEIN, Basin Set, Item No.570-1803
- Quantity
- 96 units
- Codes and lots
- UDI/DI (case) H65857018031, UDI/DI (kit)M75257018030, Serial/Lot Numbers: 22258490734
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0508-2023
Henry Schein…
- Product
- HENRY SCHEIN, BREAST PACK, Item No.570-2397
- Quantity
- 8 units
- Codes and lots
- UDI/DI (case) H65857023971, UDI/DI (kit)M75257023970, Serial/Lot Numbers: 22235489451
- Recall started
- Nov. 2, 2022
- Device type
- Plastic Surgery And Accessories Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0509-2023
Henry Schein…
- Product
- HENRY SCHEIN, GYN PACK, Item No.570-2625
- Quantity
- 20 units
- Codes and lots
- UDI/DI (case) H65857026251, UDI/DI (kit)M75257026250, Serial/Lot Numbers: 22230491067
- Recall started
- Nov. 2, 2022
- Device type
- Vaginal Examination Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0510-2023
Henry Schein…
- Product
- HENRY SCHEIN, ENT PACK, Item No.570-2718
- Quantity
- 104 units
- Codes and lots
- UDI/DI (case) H65857027181, UDI/DI (kit)M75257027180, Serial/Lot Numbers: 22242489582, 22256491864
- Recall started
- Nov. 2, 2022
- Device type
- Oral Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0511-2023
Henry Schein…
- Product
- HENRY SCHEIN, BASIC PACK, Item No.570-2719, UDI/DI (case) H65857027191, UDI/DI (kit)M75257027190, Serial/Lot Numbers: 22242489505
- Quantity
- 24 units
- Codes and lots
- UDI/DI (case) H65857027181, UDI/DI (kit)M75257027180, Serial/Lot Numbers: 22242489582
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0512-2023
Henry Schein…
- Product
- HENRY SCHEIN, VEIN TRAY, Item No.570-2738,
- Quantity
- 18 units
- Codes and lots
- UDI/DI (case) H65857027381, UDI/DI (kit)M75257027380, Serial/Lot Numbers: 22257490591
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0513-2023
Henry Schein…
- Product
- HENRY SCHEIN, ENT SET UP PACK, Item No.570-2742
- Quantity
- 12 units
- Codes and lots
- UDI/DI (case) H65857027421, UDI/DI (kit)M75257027420, Serial/Lot Numbers: 22252489778
- Recall started
- Nov. 2, 2022
- Device type
- Ear, Nose, And Throat Surgical Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0514-2023
Henry Schein…
- Product
- HENRY SCHEIN, OMFS, Item No.570-2779,
- Quantity
- 90 units
- Codes and lots
- UDI/DI (case) H65857027791, UDI/DI (kit)M75257027790, Serial/Lot Numbers: 22262489585, 22270490757
- Recall started
- Nov. 2, 2022
- Device type
- General Purpose Dental Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0515-2023
Henry Schein…
- Product
- HENRY SCHEIN, MAJOR EXTREMITY PACK, Item No.570-2783,
- Quantity
- 5 units
- Codes and lots
- UDI/DI (case) H65857027831, UDI/DI (kit)M75257027830, Serial/Lot Numbers: 22234489305
- Recall started
- Nov. 2, 2022
- Device type
- Orthopedic Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0516-2023
Henry Schein…
- Product
- HENRY SCHEIN, UNIVERSAL TRAY, Item No.570-2787
- Quantity
- 675 units
- Codes and lots
- UDI/DI (case) H65857027871, UDI/DI (kit)M75257027870, Serial/Lot Numbers: 22262491981
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0517-2023
Henry Schein…
- Product
- HENRY SCHEIN, CATH PACK Item No.570-2789
- Quantity
- 100 units
- Codes and lots
- UDI/DI (case) H65857027891, UDI/DI (kit)M75257027890, Serial/Lot Numbers: 22252491561
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0518-2023
Henry Schein…
- Product
- HENRY SCHEIN, PACEMAKER PACK IItem No.570-2793
- Quantity
- 18 units
- Codes and lots
- UDI/DI (case) H65857027931, UDI/DI (kit)M75257027930, Serial/Lot Numbers: 22241489575
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0519-2023
Henry Schein…
- Product
- HENRY SCHEIN, TEMP STIMULATOR SUPPLIES, Item No.570-2883
- Quantity
- 24 units
- Codes and lots
- UDI/DI (case) H65857028831, UDI/DI (kit)M75257028831, Serial/Lot Numbers: 22229489323
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0520-2023
Henry Schein…
- Product
- HENRY SCHEIN, BASIC EXTENDED PACK, Item No.570-2890
- Quantity
- 25 units
- Codes and lots
- UDI/DI (case) H65857028901, UDI/DI (kit)M75257028900, Serial/Lot Numbers: 22262492001
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0521-2023
Henry Schein…
- Product
- HENRY SCHEIN, LOOP RECORDER PACK, Item No.570-2891
- Quantity
- 144 units
- Codes and lots
- UDI/DI (case) H65857028911, UDI/DI (kit)M75257028910, Serial/Lot Numbers: 22252491813
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0522-2023
Henry Schein…
- Product
- HENRY SCHEIN, PROCEDURE TRAY, Item No.570-2900,
- Quantity
- 10 units
- Codes and lots
- UDI/DI (case) H65857029001, UDI/DI (kit)M75257029000, Serial/Lot Numbers: 22252489760
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0523-2023
Henry Schein…
- Product
- HENRY SCHEIN, Knee Pack, Item No.570-2908,
- Quantity
- 48 units
- Codes and lots
- UDI/DI (case) H65857029081, UDI/DI (kit)M75257029080, Serial/Lot Numbers: 22241489773
- Recall started
- Nov. 2, 2022
- Device type
- Orthopedic Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0524-2023
Henry Schein…
- Product
- HENRY SCHEIN, CATH ANGIO PACK, Item No.570-2930
- Quantity
- 16 units
- Codes and lots
- UDI/DI (case) H65857029301, UDI/DI (kit)M75257029300, Serial/Lot Numbers: 22235490920
- Recall started
- Nov. 2, 2022
- Device type
- Angiography/Angioplasty Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0525-2023
Henry Schein…
- Product
- HENRY SCHEIN, EXTREMITY PACK, Item No.570-2946
- Quantity
- 144 units
- Codes and lots
- UDI/DI (case) H65857029461, UDI/DI (kit)M75257029460, Serial/Lot Numbers: 22242490745
- Recall started
- Nov. 2, 2022
- Device type
- Orthopedic Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0526-2023
Henry Schein…
- Product
- HENRY SCHEIN, Amnio Tray, Item No.570-3059,
- Quantity
- 510 units
- Codes and lots
- UDI/DI (case) H65857030591, UDI/DI (kit)M75257030590, Serial/Lot Numbers: 22257487536
- Recall started
- Nov. 2, 2022
- Device type
- Sampler, Amniotic Fluid (Amniocentesis Tray)
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0527-2023
Henry Schein…
- Product
- HENRY SCHEIN, RHINO PACK, Item No.570-3070
- Quantity
- 328 units
- Codes and lots
- UDI/DI (case) H65857030701, UDI/DI (kit)M75257030700, Serial/Lot Numbers: 22235491255, 22241491256
- Recall started
- Nov. 2, 2022
- Device type
- Plastic Surgery And Accessories Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0528-2023
Henry Schein…
- Product
- HENRY SCHEIN, CUSTOM PK 2, Item No.570-3072
- Quantity
- 27 units
- Codes and lots
- UDI/DI (case) H65857030721, UDI/DI (kit)M75257030720, Serial/Lot Numbers: 22241490953
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0529-2023
Henry Schein…
- Product
- HENRY SCHEIN, SINGLE SHOT EPIDURAL, Item No.570-3089
- Quantity
- 80 units
- Codes and lots
- UDI/DI (case) H65857030891, UDI/DI (kit)M75257030890, Serial/Lot Numbers: 22251491690
- Recall started
- Nov. 2, 2022
- Device type
- Epidural Anesthesia Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0530-2023
Henry Schein…
- Product
- HENRY SCHEIN, Blue Tail Custom Pack, Item No.570-3091,
- Quantity
- 72 units
- Codes and lots
- UDI/DI (case) H65857030911, UDI/DI (kit)M75257030910, Serial/Lot Numbers: 22258490762
- Recall started
- Nov. 2, 2022
- Device type
- Bone Marrow Biopsy Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0531-2023
Henry Schein…
- Product
- HENRY SCHEIN, BONE MARROW KIT, Item No.570-3092
- Quantity
- 581 units
- Codes and lots
- UDI/DI (case) H65857030921, UDI/DI (kit)M75257030920, Serial/Lot Numbers: 22231491193
- Recall started
- Nov. 2, 2022
- Device type
- Bone Marrow Biopsy Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0532-2023
Henry Schein…
- Product
- HENRY SCHEIN, Labiaplasty PackItem No.570-3118,
- Quantity
- 42 units
- Codes and lots
- UDI/DI (case) H65857031181, UDI/DI (kit)M75257031180, Serial/Lot Numbers: 22234491292
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0533-2023
Henry Schein…
- Product
- HENRY SCHEIN, ILR Insertion Tray, Item No.570-3131
- Quantity
- 140 units
- Codes and lots
- UDI/DI (case) H65857031311, UDI/DI (kit)M75257031310, Serial/Lot Numbers: 22257490596
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0534-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, CURVED TIP SYRINGE, Item No.60060GCTS,
- Quantity
- 300 units
- Codes and lots
- UDI/DI (case) M75260060GCTS1, UDI/DI (kit)M75260060GCTS0, Serial/Lot Numbers: 22257490438
- Recall started
- Nov. 2, 2022
- Device type
- Syringe, Periodontic, Endodontic, Irrigating
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0535-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Lipo, Item No.681-254,
- Quantity
- 6 units
- Codes and lots
- UDI/DI (case) M7526812541, UDI/DI (kit)M7526812540, Serial/Lot Numbers: 22270492010
- Recall started
- Nov. 2, 2022
- Device type
- Plastic Surgery And Accessories Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0536-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Phlebectomy Pack, Item No.682-001,
- Quantity
- 15 units
- Codes and lots
- UDI/DI (case) M7526820011, UDI/DI (kit)M7526820010, Serial/Lot Numbers: 22236489633
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0537-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Thoracic Phleb, Item No.682-1228,
- Quantity
- 5 units
- Codes and lots
- UDI/DI (case) M7268212281, UDI/DI (kit)M7268212280, Serial/Lot Numbers: 22237491041
- Recall started
- Nov. 2, 2022
- Device type
- Plastic Surgery And Accessories Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0538-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Ablation Pack, Item No.682-130,
- Quantity
- 40 units
- Codes and lots
- UDI/DI (case) M7526821301, UDI/DI (kit)M7526821300, Serial/Lot Numbers: 22237491040
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0539-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Phleb, Item No.682-1425,
- Quantity
- 20 units
- Codes and lots
- UDI/DI (case) M75268214251, UDI/DI (kit)M75268214250, Serial/Lot Numbers: 22236490624
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0540-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Vasc Vein, Item No.682-1583,
- Quantity
- 20 units
- Codes and lots
- UDI/DI (case) M75268215831, UDI/DI (kit)M75268215830, Serial/Lot Numbers: 22237491038
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0541-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Surgery PK, Item No.682-1728,
- Quantity
- 40 units
- Codes and lots
- UDI/DI (case) M75268217281, UDI/DI (kit)M75268217280, Serial/Lot Numbers: 22235489419
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0542-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Vein, Item No.682-1732,
- Quantity
- 10 units
- Codes and lots
- UDI/DI (case) M75268217321, UDI/DI (kit)M75268217320, Serial/Lot Numbers: 22230489185
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0543-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Angioplasty, Item No.682-1937,
- Quantity
- 100 units
- Codes and lots
- UDI/DI (case) M75268219371, UDI/DI (kit)M75268219370, Serial/Lot Numbers: 22237490509
- Recall started
- Nov. 2, 2022
- Device type
- Angiography/Angioplasty Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0544-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Vein RF Pack, Item No.682-1942NL1,
- Quantity
- 400 units
- Codes and lots
- UDI/DI (case) M7526821942NL11, UDI/DI (kit)M7526821942NL10, Serial/Lot Numbers: 22251491808
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0545-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, VENOUS ACCESS PACK, Item No.682-1993R1,
- Quantity
- 30 units
- Codes and lots
- UDI/DI (case) M7526821993R11, UDI/DI (kit)M7526821993R10, Serial/Lot Numbers: 22235491208
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0546-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Medical Vein, Item No.682-330,
- Quantity
- 40 units
- Codes and lots
- UDI/DI (case) M7526823301, UDI/DI (kit)M7526823300, Serial/Lot Numbers: 22238490518
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0547-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Vein & Laser, Item No.682-334R1,
- Quantity
- 10 units
- Codes and lots
- UDI/DI (case) M752682334R11, UDI/DI (kit)M752682334R10, Serial/Lot Numbers: 22231489619
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0548-2023
www.StradisHealthcare.com…
- Product
- www.StradisHealthcare.com, Endovenous Pack, Item No.682-494,
- Quantity
- 10 units
- Codes and lots
- UDI/DI (case) M7526824941, UDI/DI (kit)M7526824940, Serial/Lot Numbers: 22234489198
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0549-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Endo Pack, Item No.682-509,
- Quantity
- 25 units
- Codes and lots
- UDI/DI (case) M7526825091, UDI/DI (kit)M7526825090, Serial/Lot Numbers: 22256490827
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0550-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Vein, Item No.682-577,
- Quantity
- 15 units
- Codes and lots
- UDI/DI (case) M7526825771, UDI/DI (kit)M7526585770, Serial/Lot Numbers: 22230490618
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0551-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Varicosity Basic Pack, Item No.682-947,
- Quantity
- 105 units
- Codes and lots
- UDI/DI (case) M7526829471, UDI/DI (kit)M7526829470, Serial/Lot Numbers: 22236491032
- Recall started
- Nov. 2, 2022
- Device type
- Cardiovascular Procedure Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0552-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Femoral, Vasc, Item No.686-142,
- Quantity
- 10
- Codes and lots
- UDI/DI (case) M7526861421, UDI/DI (kit)M7526861420, Serial/Lot Numbers: 22236491028
- Recall started
- Nov. 2, 2022
- Device type
- Angiography/Angioplasty Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0553-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Angio Pack, Item No.686-199,
- Quantity
- 24 units
- Codes and lots
- UDI/DI (case) M7526861991, UDI/DI (kit)M7526861990, Serial/Lot Numbers: 22236491017
- Recall started
- Nov. 2, 2022
- Device type
- Angiography/Angioplasty Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0554-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, Breast, Item No.693-276,
- Quantity
- 16 units
- Codes and lots
- UDI/DI (case) M7526932761, UDI/DI (kit)M7526932760, Serial/Lot Numbers: 22242490001
- Recall started
- Nov. 2, 2022
- Device type
- Plastic Surgery And Accessories Kit
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0555-2023
Henry Schein…
- Product
- HENRY SCHEIN, Vas Pack, Item No.888-2520,
- Quantity
- 304 units
- Codes and lots
- UDI/DI (case) H65888825201, UDI/DI (kit)M75288825200, Serial/Lot Numbers: 22256490684
- Recall started
- Nov. 2, 2022
- Device type
- General Surgery Tray
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0556-2023
Henry Schein…
- Product
- HENRY SCHEIN, Vas Reversal Pack, Item No.888-2549,
- Quantity
- 3 units
- Codes and lots
- UDI/DI (case) H65888825491, UDI/DI (kit)M75288825490, Serial/Lot Numbers: 22230489459
- Recall started
- Nov. 2, 2022
- Device type
- Vasovasostomy Set
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0557-2023
Stradis Healthcare…
- Product
- STRADIS HEALTHCARE, TIP, SURGE-O-VAC .125" GREEN, Item No.SS077225,
- Quantity
- 75 units
- Codes and lots
- UDI/DI (case) M752SS0772251, UDI/DI (kit)M752SS0772251, Serial/Lot Numbers: 22255490439
- Recall started
- Nov. 2, 2022
- Device type
- Evacuator, Oral Cavity
- Root cause (FDA)
- Packaging process control
- What the company told customers
- Stradis Healthcare issued an URGENT: MEDICAL DEVICE RECALL to its consignees on 11/2/2022 by email. The notice explained the problem, risk, and requested the following action be taken: Please examine your inventory and immediately quarantine any affected product. Please contact Stradis Customer Care to arrange for return shipment of the recalled product along with a full refund, credit or product substitute. Stradis Customer Care can be reached at 1-800-886-7257 or by email at cs@stradishealthcare.com. In the event you further distributed the affected kits, please identify your customers, and notify them within forty-eight (48) hours of receipt of this recall notification to inform them of this issue. Your customers may contact Stradis Customer Care to arrange for any of the affected surgical kits to be returned to Stradis Healthcare and credited accordingly. Alternatively, if they communicate through you, we ask that you immediately inform Stradis Healthcare as you become aware. Notification to your customers may be enhanced by including a copy of this recall letter. Stradis Healthcare is also requesting return of the attached Recall Acknowledgement Form to confirm receipt of this recall communication and that you have taken all necessary actions as described above.
- Recall number
- Z-0558-2023
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-0481-2023, Z-0482-2023, Z-0483-2023, Z-0484-2023, Z-0485-2023, Z-0486-2023, Z-0487-2023, Z-0488-2023 and others. The agency record governs; check it and the company before acting.