FDA · Medical device recall
VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products…
- Date
- Nov. 25, 2020
- Risk class
- Class III (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences)
- Status
- Ongoing
- Filed by
- Ortho Clinical Diagnostics Inc
- Product
- VITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated System - Product Code: 8231474
- Reason
- Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures
- Distribution pattern
- Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
- Quantity
- 10788 total: US 4680; OUS 6108
- Codes and lots
- Lot Q7692 , EXP: 01-NOV-2021 UDI: 10758750004577
- Recall started
- Oct. 16, 2020
- Device type
- Multi-Analyte Controls, All Kinds (Assayed)
- Root cause (FDA)
- Under Investigation by firm
- What the company told customers
- Ortho Clinical Diagnostics issued letter IMPORTANT PRODUCT CORRECTION NOTIFICATION (CL2020-243) on 16 October 2020 via FedEx/US Priority Mail or ORTHO Plus e-Communications for delivery on October 19, 2020, stating reason for recall, health risk and action to take: If their laboratory used VITROS Performance Verifier II Lot Q7692 to verify performance of VITROS ALKP Slides and stores an inventory of unopened vials in the refrigerator: Discard any remaining refrigerated inventory of unopened VITROS Performance Verifier II Lot Q7692. If their laboratory does not store inventory of VITROS Performance Verifier II Lot Q7692 in the refrigerator or does not use Lot Q7692 to verify performance of VITROS ALKP Slides: No Action Is Required. To verify performance of VITROS ALKP Slides, use frozen (never refrigerated) inventory of VITROS Performance Verifier II Lot Q7692 or switch to another lot of VITROS Performance Verifier II. Ortho will replace or credit your account as indicated on your Confirmation of Receipt form Retain this notification as verification of the revised instructions. OUS notified by email on 09/14/20 to global affiliates for distribution to distributors and customers.
- Recall number
- Z-0471-2021
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0471-2021. The agency record governs; check it and the company before acting.