Menu
FDA · Medical device recall

GE HEALTHCARE CARESCAPE R860 Inspiratory Safety Guard…

Dec. 12, 2018 Class I

Date
Dec. 12, 2018
Risk class
Class I (FDA's definition, 21 CFR 7.3(m): a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
GE HEALTHCARE CARESCAPE R860 Inspiratory Safety Guard, Part Numbers: a) 2066713-001 (single pack) b) 2083208-001 (box of ten singles) The CARESCAPE R860 ventilator is designed to provide mechanical ventilation or support to neonatal, pediatric, and adult patients weighing 0.25 kg and above. The CARESCAPE R860 ventilator is a microprocessor based, electronically controlled, pneumatically driven ventilator that includes integrated monitoring of FiO2, airway pressure, flow, and volume.
Reason
The ventilator Inspiratory Safety Guard (ISG) may disconnect from the breathing circuit pathway. As a result, this could create a loss of ventilation which may lead to inadequate oxygenation for patients, increasing the possibility of hypoxia.
Distribution pattern
Worldwide Distribution: US (Nationwide) including Puerto Rico, and to countries of: Albania, Argentina, Australia, Austria, Bangladesh, Bolivia, Bosnia and Herzegovina, Brazil, Cambodia, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Greece, Honduras, Hungary, India, Indonesia, Iraq, Italy, Japan, Kazakhstan, Korea (Republic of), Kuwait, Latvia, Lebanon, Libya, Malaysia, Maldives, Mexico, Myanmar, Nepal, Netherlands, Nicaragua, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, Uruguay, and Vietnam.
Quantity
1140 units (307 units in US)
Codes and lots
a) 2066713-001 Lot Numbers: 17/00951, 17/01174, 17/01937, 17/01994, 17/02372, 17/02393, 18/00126, 18/00127, 18/00128, 18/00129, 18/00130 b) 2083208-001 Lot Numbers: 17/00951, 17/01174, 17/01937, 17/01994, 17/02372, 17/02393, 18/00126, 18/00127, 18/00128, 18/00129, 18/00130
Recall started
Oct. 5, 2018
Device type
Ventilator, Continuous, Facility Use
Root cause (FDA)
Process control
What the company told customers
The firm, GE Healthcare, initiated an "URGENT MEDICAL DEVICE CORRECTION" on 10/05/2018 by letter to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to do the following: Follow the safety instructions and If you have any incorrect ISGs, return to GE Healthcare or destroy on site, and complete and return the attached Customer Response form checking box #2 to indicate that you do have affected ISGs. E-mail the completed form to Recall34095.InspiratorySG@ge.com. Please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative. The firm initiated a new recall notice identifying a new issue on 02/01/2019
Recall number
Z-0463-2019

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0463-2019. The agency record governs; check it and the company before acting.

Report a mistake on this page