FDA · Medical device recall
RunWay 6F Cardiovascular Guide Catheter
- Date posted by the FDA
- Jan. 28, 2005
- Status
- Terminated
- Filed by
- Boston Scientific Scimed
- Product
- RunWay 6F Cardiovascular Guide Catheter.
- Reason
- The Runway guide catheter do not meet internal product specifications. Specifically, the affected catheters have exposed or protruding braid wires in the inner lumen of the catheter tip. This condition could have the potential to damage a POBA or Stent Delivery System balloon
- Distribution pattern
- Throughout United States and into 15 other countries.
- Quantity
- 76,984
- Codes and lots
- all models and lot numbers of the Boston Scientific RunWay 6F Guide Catheter.
- Recall started
- Oct. 29, 2004
- Device type
- Catheter, Percutaneous
- Root cause (FDA)
- Other
- What the company told customers
- On 10/29/04 the Boston Scientific Sales force was notified via voice mail of the recall. Sales personnel were told to contact their accounts/customers by the end of the business day on 11/02/04. Accounts/customers were to receive written notice of the recall via Fed Ex on Monday 11/01/04. Sales Personnel are to ensure that accounts/customers got notice of the recall, understand instructions and are taking appropriate steps. Catheters are to be returned to Boston Scientific Corporation, Quincy, MA.
- Recall number
- Z-0454-05