NovoTHOR Gen 2.0 whole body red light therapy bed… and 3 more products

Dec. 13, 2023 Class II

Date
Dec. 13, 2023
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Thor Photomedicine Ltd
Reason
Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.
Distribution pattern
Worldwide Distribution: US (Nationwide): Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Maryland, Michigan, Minnesota, Missouri, Nevada, New Hampshire, North Carolina, Ohio, Oregon, Texas, Utah, Virginia, West Virginia. OUS (foreign) to countries of: Australia, Canada, Netherlands, New Zealand, Singapore, United Kingdom.

4 products in this recall

NovoTHOR Gen 2.0 whole body red light therapy bed…

Product
NovoTHOR Gen 2.0 whole body red light therapy bed, model S1176
Quantity
19
Codes and lots
Serial numbers: 2003, 2004, 2005, 2006, 2007, 2009, 2010, 2011, 2017, 2018, 2015, 2020, 2019, 2001, 2021, 2028, 2033, 2036, 2030, 2029.
Recall started
Sept. 13, 2022
Device type
Lamp, Infrared, Therapeutic Heating
Root cause (FDA)
Device Design
What the company told customers
The "IMPORTANT MEDICAL DEVICE CORRECTION" notification issued October 05, 2023 by email with follow-up phone call. The notice advises users (1) how to recognize the issue. (2) THOR is replacing all ball studs on all affected devices. Should you continue to have users/patients use the NovoTHOR, then be aware it may potentially cause harm should the failure occur whilst they are lifting or lowering the canopy. " Ensure you warn all users of the potential risk " Ensure someone is present to lift and lower the canopy " Ensure the canopy is lifted and lowered gently and slowly " Ensure someone is present in the room or immediately outside the room during use. A service representative will contact customers within 6 weeks to schedule correction. Customers are requested to respond to the email notice with their acknowledgement. Contact phone number for Thor is (877) 355 3151.
Recall number
Z-0453-2024

G041 Pod…

Product
G041 POD, THOR NOVO GEN 2.0 XL whole body red light therapy bed, model S1177
Quantity
6
Codes and lots
Serial numbers: 3001, 3002, 3004, 3008, 3009, 3006, 3003.
Recall started
Sept. 13, 2022
Device type
Lamp, Infrared, Therapeutic Heating
Root cause (FDA)
Device Design
What the company told customers
The "IMPORTANT MEDICAL DEVICE CORRECTION" notification issued October 05, 2023 by email with follow-up phone call. The notice advises users (1) how to recognize the issue. (2) THOR is replacing all ball studs on all affected devices. Should you continue to have users/patients use the NovoTHOR, then be aware it may potentially cause harm should the failure occur whilst they are lifting or lowering the canopy. " Ensure you warn all users of the potential risk " Ensure someone is present to lift and lower the canopy " Ensure the canopy is lifted and lowered gently and slowly " Ensure someone is present in the room or immediately outside the room during use. A service representative will contact customers within 6 weeks to schedule correction. Customers are requested to respond to the email notice with their acknowledgement. Contact phone number for Thor is (877) 355 3151.
Recall number
Z-0454-2024

NovoTHOR Gen 2.5 Regular whole body red light therapy bed…

Product
NovoTHOR Gen 2.5 Regular whole body red light therapy bed, model S2183
Quantity
26
Codes and lots
Serial numbers: 2002, 2050, 2051, 2052, 2053, 2054, 2055, 2056, 2057, 2059, 2060, 2061, 2062, 2064, 2065, 2066, 2075, 2076, 2077, 2078, 2081, 2084, 2085, 2086, 2088, 2089, 2090, 2094, 2097.
Recall started
Sept. 13, 2022
Device type
Lamp, Infrared, Therapeutic Heating
Root cause (FDA)
Device Design
What the company told customers
The "IMPORTANT MEDICAL DEVICE CORRECTION" notification issued October 05, 2023 by email with follow-up phone call. The notice advises users (1) how to recognize the issue. (2) THOR is replacing all ball studs on all affected devices. Should you continue to have users/patients use the NovoTHOR, then be aware it may potentially cause harm should the failure occur whilst they are lifting or lowering the canopy. " Ensure you warn all users of the potential risk " Ensure someone is present to lift and lower the canopy " Ensure the canopy is lifted and lowered gently and slowly " Ensure someone is present in the room or immediately outside the room during use. A service representative will contact customers within 6 weeks to schedule correction. Customers are requested to respond to the email notice with their acknowledgement. Contact phone number for Thor is (877) 355 3151.
Recall number
Z-0455-2024

NovoTHOR Gen 2.5 XL whole body red light therapy bed…

Product
NovoTHOR Gen 2.5 XL whole body red light therapy bed, model S2184.
Quantity
26
Codes and lots
Serial numbers: 3025, 3026, 3027, 3030, 3031, 3034, 3035, 3036, 3037, 3038, 3039, 3040, 3042, 3043, 3044, 3045, 3047, 3050, 3051, 3053.
Recall started
Sept. 13, 2022
Device type
Lamp, Infrared, Therapeutic Heating
Root cause (FDA)
Device Design
What the company told customers
The "IMPORTANT MEDICAL DEVICE CORRECTION" notification issued October 05, 2023 by email with follow-up phone call. The notice advises users (1) how to recognize the issue. (2) THOR is replacing all ball studs on all affected devices. Should you continue to have users/patients use the NovoTHOR, then be aware it may potentially cause harm should the failure occur whilst they are lifting or lowering the canopy. " Ensure you warn all users of the potential risk " Ensure someone is present to lift and lower the canopy " Ensure the canopy is lifted and lowered gently and slowly " Ensure someone is present in the room or immediately outside the room during use. A service representative will contact customers within 6 weeks to schedule correction. Customers are requested to respond to the email notice with their acknowledgement. Contact phone number for Thor is (877) 355 3151.
Recall number
Z-0456-2024

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0453-2024, Z-0454-2024, Z-0455-2024, Z-0456-2024. The agency record governs; check it and the company before acting.