FDA · Medical device recall

GE Healthcare Optima XR 220amx 30kw Mobile X-ray System The GE… and 2 more products

Dec. 5, 2012 Class II

Date
Dec. 5, 2012
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Reason
GE Healthcare has become aware of a software issue on the interface of the mobile radiographic product, Optima XR22Oamx and Optima XR200amx with Digital Upgrade.
Distribution pattern
Nationwide distribution

3 products in this recall

GE Healthcare Optima XR 220amx 30kw Mobile X-ray System The GE Automatic Mobile X-Ray (AMX) Series: Brivo…

Product
GE Healthcare Optima XR 220amx 30kw Mobile X-ray System The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR2S5amx, Optima XR200amx, Optima XR22Oamx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR22Oamx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).
Quantity
217 total Installed in the US
Codes and lots
Model No - 5555000-5
Recall started
Aug. 1, 2012
Device type
System, X-Ray, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will without charge remedy the defect or bring the product into compliance with each applicable Federal and IEC standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication through a GE Healthcare field engineer site visit. Field Modification Instructions (FMI) 10868 describes the rework plan the GEHC Engineers will make to the imaging system compliant with the applicable performance standard. This CAP appears to adequately address the problem and is hereby approved. Call Center phone number: 800-437-1171 (United States). For other countries, please contact your local GE Healthcare Service Representative if you have any questions or concerns regarding this notification.
Recall number
Z-0448-2013

GE Healthcare Optima XR 220amx 15kw Mobile X-Ray System The GE Automatic Mobile X-Ray (AMX) Series: Brivo…

Product
GE Healthcare Optima XR 220amx 15kw Mobile X-Ray System The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR2S5amx, Optima XR200amx, Optima XR22Oamx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR22Oamx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).
Quantity
217 total Installed in the US
Codes and lots
Model No - 5555000-6
Recall started
Aug. 1, 2012
Device type
System, X-Ray, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will without charge remedy the defect or bring the product into compliance with each applicable Federal and IEC standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication through a GE Healthcare field engineer site visit. Field Modification Instructions (FMI) 10868 describes the rework plan the GEHC Engineers will make to the imaging system compliant with the applicable performance standard. This CAP appears to adequately address the problem and is hereby approved. Call Center phone number: 800-437-1171 (United States). For other countries, please contact your local GE Healthcare Service Representative if you have any questions or concerns regarding this notification.
Recall number
Z-0449-2013

GE Healthcare Optima XR 200 amx with Digital Upgrade…

Product
GE Healthcare Optima XR 200 amx with Digital Upgrade. The GE Automatic Mobile X-Ray (AMX) Series: Brivo XR2S5amx, Optima XR200amx, Optima XR22Oamx are intended to take exposures utilizing film or computed radiography (CR), however the Optima XR22Oamx utilizes the GE Wireless Detector, which is intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR).
Quantity
217 total installed in the US
Codes and lots
Model No - 5421698
Recall started
Aug. 1, 2012
Device type
System, X-Ray, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will without charge remedy the defect or bring the product into compliance with each applicable Federal and IEC standard in accordance with a plan to be approved by the Secretary of Health and Human Services, the details of which will be included in a subsequent communication through a GE Healthcare field engineer site visit. Field Modification Instructions (FMI) 10868 describes the rework plan the GEHC Engineers will make to the imaging system compliant with the applicable performance standard. This CAP appears to adequately address the problem and is hereby approved. Call Center phone number: 800-437-1171 (United States). For other countries, please contact your local GE Healthcare Service Representative if you have any questions or concerns regarding this notification.
Recall number
Z-0450-2013

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0448-2013, Z-0449-2013, Z-0450-2013. The agency record governs; check it and the company before acting.