INTRABEAM Balloon Applicator Set… and 2 more products

Nov. 28, 2012 Class II

Date
Nov. 28, 2012
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Carl Zeiss Meditec, Inc.
Reason
A manufacturing defect has been identified which could results in the small tube becoming dislodged between the luer fitting and the check valve.
Distribution pattern
Nationwide Distribution including IL and MD

3 products in this recall

INTRABEAM Balloon Applicator Set…

Product
INTRABEAM Balloon Applicator Set; 3.0 cm, Carl Zeiss Meditec AG, Oberkochen, Germany INTRABEAM System is intended for radiation therapy treatments
Quantity
36 tootal, all lots
Codes and lots
Part number 304534-7500-000; lot number 540958.
Recall started
Oct. 25, 2012
Device type
System, Therapeutic, X-Ray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Carl Zeiss Meditec sent an Important Field Corrective Action letter dated October 25, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately return the affected product to Carl Zeiss Meditec, Inc. A Carl Zeiss Meditec representative will contact the customer to schedule a visit to support the removal of the affected product. For questions customers were instructed to call 925-580-5371 or by email at v.venturo@meditec.zeiss.com. For questions regarding this recall call 925-557-4616.
Recall number
Z-0404-2013

INTRABEAM Balloon Applicator Set…

Product
INTRABEAM Balloon Applicator Set; 3.5 cm, Carl Zeiss Meditec AG, Oberkochen, Germany INTRABEAM System is intended for radiation therapy treatments.
Quantity
36 total, all lots
Codes and lots
Part number 304534-7500-001; lot number 540959.
Recall started
Oct. 25, 2012
Device type
System, Therapeutic, X-Ray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Carl Zeiss Meditec sent an Important Field Corrective Action letter dated October 25, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately return the affected product to Carl Zeiss Meditec, Inc. A Carl Zeiss Meditec representative will contact the customer to schedule a visit to support the removal of the affected product. For questions customers were instructed to call 925-580-5371 or by email at v.venturo@meditec.zeiss.com. For questions regarding this recall call 925-557-4616.
Recall number
Z-0405-2013

INTRABEAM Balloon Applicator Set…

Product
INTRABEAM Balloon Applicator Set; 4.0 cm, Carl Zeiss Meditec AG, Oberkochen, Germany INTRABEAM System is intended for radiation therapy treatments.
Quantity
36 total, all lots
Codes and lots
Part number 304534-7500-002; lot number 540960.
Recall started
Oct. 25, 2012
Device type
System, Therapeutic, X-Ray
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Carl Zeiss Meditec sent an Important Field Corrective Action letter dated October 25, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to immediately return the affected product to Carl Zeiss Meditec, Inc. A Carl Zeiss Meditec representative will contact the customer to schedule a visit to support the removal of the affected product. For questions customers were instructed to call 925-580-5371 or by email at v.venturo@meditec.zeiss.com. For questions regarding this recall call 925-557-4616.
Recall number
Z-0406-2013

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0404-2013, Z-0405-2013, Z-0406-2013. The agency record governs; check it and the company before acting.