Vitros CRP (C-Reactive Protein) Slides…

Jan. 8, 2005

Date posted by the FDA
Jan. 8, 2005
Status
Terminated
Filed by
Ortho-Clinical Diagnostics
Product
Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Firm on label: Ortho-Clinical Diagnostics, a Johnson & Johnson Company, Rochester, NY 14626.
Reason
Results may be negatively biased due to a non-linear response at the upper end of the Reportable (Dynamic) Range.
Distribution pattern
Nationwide and worldwide distribution. 70% of product was shipped to foreign consignees.
Quantity
Cat. 1926740 - 20,903 boxes; Cat 8097990 - 34,877 boxes
Codes and lots
Coatings 401 thru 419.
Recall started
Nov. 22, 2004
Device type
C-Reactive Protein, Antigen, Antiserum, And Control
Root cause (FDA)
Other
What the company told customers
Letters dated 11/22/2004 issued to customers. End users instructed to manually modify the Reportable and/or Analyzer Ranges; dilute all samples with an initial CRP result greater than 9.0 mg/dL. Distributors instructed to destroy product in-house and to notify end-users to manually modify the Reportable and/or Analyzer Ranges.
Recall number
Z-0387-05