FDA · Medical device recall
Vitros CRP (C-Reactive Protein) Slides…
- Date posted by the FDA
- Jan. 8, 2005
- Status
- Terminated
- Filed by
- Ortho-Clinical Diagnostics
- Product
- Vitros CRP (C-Reactive Protein) Slides. Catalog #192 6740 contains 250 slides; Catalog #809 7990 contains 90 slides. Firm on label: Ortho-Clinical Diagnostics, a Johnson & Johnson Company, Rochester, NY 14626.
- Reason
- Results may be negatively biased due to a non-linear response at the upper end of the Reportable (Dynamic) Range.
- Distribution pattern
- Nationwide and worldwide distribution. 70% of product was shipped to foreign consignees.
- Quantity
- Cat. 1926740 - 20,903 boxes; Cat 8097990 - 34,877 boxes
- Codes and lots
- Coatings 401 thru 419.
- Recall started
- Nov. 22, 2004
- Device type
- C-Reactive Protein, Antigen, Antiserum, And Control
- Root cause (FDA)
- Other
- What the company told customers
- Letters dated 11/22/2004 issued to customers. End users instructed to manually modify the Reportable and/or Analyzer Ranges; dilute all samples with an initial CRP result greater than 9.0 mg/dL. Distributors instructed to destroy product in-house and to notify end-users to manually modify the Reportable and/or Analyzer Ranges.
- Recall number
- Z-0387-05