FDA · Medical device recall
TC pCO2 Electrodes…
- Date posted by the FDA
- Dec. 1, 2011
- Status
- Terminated
- Filed by
- Radiometer America Inc
- Product
- TC pCO2 Electrodes, Model/Part #945-655-E5260, 510(k) #K043003. Product Usage: The device is intended for used with the firm's TCM series monitor. The E260mpCO2 electrode is a non-invasive device, used with the TCM series monitor to measure transcutaneous pCO2 (through the skin).
- Reason
- Transcutaneous pCO2-electrodes were wrongly coded as: pO2 electrode on the hardware plug for the firm's TCM series monitors. This causes the monitor to interpret the electrode wrongly, causing it to fail all calibrations and never get ready for use.
- Distribution pattern
- Worldwide Distribution - USA (nationwide) in the state of San Diego, CA.
- Quantity
- 4 units were distributed.
- Codes and lots
- Part Number: 945-655, Lot Code: DP-06
- Recall started
- Sept. 27, 2011
- Device type
- Monitor, Carbon-Dioxide, Cutaneous
- Root cause (FDA)
- Employee error
- What the company told customers
- Radiometer contacted the affected U.S. customer and informed them of the recall via telephone on September 27, 2011. Radiometer requested the return of the affected product received from the firm. For questions call (440) 871-0463.
- Recall number
- Z-0341-2012