FDA · Medical device recall

GE Centricity Universal Viewer Zero Footprint Client

Dec. 8, 2021 Class II

Date
Dec. 8, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
GE Centricity Universal Viewer Zero Footprint Client
Reason
There is a potential to display incomplete patient imaging study.
Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, DE, FL, GA, IL, IN, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, WA, and WV. The countries of Argentina, Aruba, Australia, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Dominican Republic, Ecuador, France, Germany, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Panama, Qatar, Russia, Saudi Arabia, South Africa, Spain, Switzerland, Taiwan, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Quantity
638 devices
Codes and lots
Versions 6.0 SP9, SP9.0.1, SP9.0.1.1, SP9.0.1.2, SP9.0.1.3, SP9.0.1.4, SP9.0.1.5, SP10, SP10.1, SP10.2, SP10.2.1, SP10.2.1.1, SP10.2.2, SP10.2.2.1 configured with a Centricity PACS back end. GTIN 00840682102988; UDI 00840682102988.
Recall started
Oct. 15, 2021
Device type
System, Image Processing, Radiological
Root cause (FDA)
Software design
What the company told customers
The firm issued medical device correction letters dated 10/15/2021 on the same date via traceable means. The letter described the safety issue and actions to be taken. The letter informed the consignee they can continue to use the device by following the instructions provided in the letter until a GE Service Representative can restart their ZFP system.
Recall number
Z-0329-2022

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0329-2022. The agency record governs; check it and the company before acting.