Posterior Acetabular Retractor-Left with Light Cord Bracket…

Dec. 27, 2006

Date posted by the FDA
Dec. 27, 2006
Status
Terminated
Filed by
Whiteside Biomechanics
Product
Posterior Acetabular Retractor-Left with Light Cord Bracket, Model #BIO-1-1050-A, and Posterior Acetabular Retractor-Right with Light Cord Bracket, Model #BIO-1-1051-A, packaged 1 retractor (left or right retractor) per plastic bag. The firm name on the label is Whiteside Biomechanics, Inc., St. Louis, MO.
Reason
The soft tissue pin in the retractor can break off and become embedded in the tissue.
Distribution pattern
Distribution was made to two hospitals, one in MD and the other in MO.
Quantity
10 retractors (5 right, 5 left)
Codes and lots
Lot #P0835-1
Recall started
Sept. 15, 2006
Device type
Retractor
Root cause (FDA)
Other
What the company told customers
The recalling firm was notified of the problem via a telephone call from the surgeon on 9/15/06 and the remaining retractors were placed in quarantine by the hospital. The recalling firm requested return of the retractors. The recalling firm immediately visited the one remaining hospital on 9/15/06 and picked up their inventory of six retractors.
Recall number
Z-0316-2007