FDA · Medical device recall

Aestiva/5 MRI Anesthesia System… and 6 more products

Nov. 11, 2020 Class II

Date
Nov. 11, 2020
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare, LLC
Reason
Flow sensors could have damaged tubes in the form of small punctures or cuts.
Distribution pattern
U.S. states where product was distributed - AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, MT, NC, NE, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI Foreign countries impacted - Australia, Canada, Czech Republic, Estonia, France, Germany, India, Indonesia, Israel, Italy, Kuwait, Latvia, Macedonia, Madagascar, Malaysia, Nicaragua, Panama, Qatar, Russia, Singapore, Slovenia, Spain, Thailand, Ukraine, United Kingdom

7 products in this recall

Aestiva/5 MRI Anesthesia System…

Product
Aestiva/5 MRI Anesthesia System - Product Usage: The Aestiva/5 MRI Anesthesia System provides the functional feature set offered by the conventional Aestiva/5 to the clinician with the added ability to be used in the MR environment. Among those standard Aestiva/5 features is the Datex-Ohmeda user interface, all the ventilation parameters of the SmartVent along with the Aestiva breathing circuit. The Aestiva/5 MRI Anesthesia System performed to specifications when tested directly next to 1.5 and 3.0 Tesla shielded MRI devices in a field strength that did not exceed 300 gauss.
Quantity
2 devices
Codes and lots
Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020)
Recall started
Sept. 17, 2020
Device type
Gas-Machine, Anesthesia
Root cause (FDA)
Process control
What the company told customers
GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.
Recall number
Z-0311-2021

GE Datex-Ohmeda Avance CS2 Anesthesia System…

Product
GE Datex-Ohmeda Avance CS2 Anesthesia System - Product Usage: intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric, adult). The device is intended for volume or pressure control ventilation.
Quantity
88 devices
Codes and lots
Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020)
Recall started
Sept. 17, 2020
Device type
Gas-Machine, Anesthesia
Root cause (FDA)
Process control
What the company told customers
GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.
Recall number
Z-0312-2021

GE Datex-Ohmeda Aisys CS2 Anesthesia System…

Product
GE Datex-Ohmeda Aisys CS2 Anesthesia System - Product Usage: is intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (neonatal, pediatric adult). The device is intended for volume or pressure control ventilation.
Quantity
101 devices
Codes and lots
Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020)
Recall started
Sept. 17, 2020
Device type
Gas-Machine, Anesthesia
Root cause (FDA)
Process control
What the company told customers
GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.
Recall number
Z-0313-2021

9100c NXT system…

Product
9100c NXT system - Product Usage: Device is licensed and distributed outside of the US only.
Quantity
37 devices
Codes and lots
Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020)
Recall started
Sept. 17, 2020
Device type
Gas-Machine, Anesthesia
Root cause (FDA)
Process control
What the company told customers
GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.
Recall number
Z-0314-2021

Flow Sensor Service Part…

Product
Flow Sensor Service Part, Catalog 2089610-001-S
Quantity
4794 units
Codes and lots
Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020). This flow sensor is compatible with the following device models: Aespire 7100/11, Aespire 7900, Aespire View, Aestiva 7100, Aestiva 7900, Aestiva MRI, Aisys, Aisys CS2, Avance, Avance CS2, Amingo (OUS), and 9100c NXT (OUS)
Recall started
Sept. 17, 2020
Device type
Gas-Machine, Anesthesia
Root cause (FDA)
Process control
What the company told customers
GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.
Recall number
Z-0315-2021

Flow Sensor Service Part…

Product
Flow Sensor Service Part, Catalog 2087640-001-S
Quantity
388
Codes and lots
Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020). This flow sensor is compatible with the following device models: Aespire 7100/11, Aespire 7900, Aespire View, Aestiva 7100, Aestiva 7900, Aestiva MRI, Aisys, Aisys CS2, Avance, Avance CS2, Amingo (OUS), and 9100c NXT (OUS)
Recall started
Sept. 17, 2020
Device type
Gas-Machine, Anesthesia
Root cause (FDA)
Process control
What the company told customers
GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.
Recall number
Z-0316-2021

Flow Sensor Service Part…

Product
Flow Sensor Service Part, Catalog 2096513-001-S
Quantity
2 units
Codes and lots
Devices with installed flow sensors with etched manufacturing date of 2020-08 (August 2020). This flow sensor is compatible with the following device models: Aespire 7100/11, Aespire 7900, Aespire View, Aestiva 7100, Aestiva 7900, Aestiva MRI, Aisys, Aisys CS2, Avance, Avance CS2, Amingo (OUS), and 9100c NXT (OUS)
Recall started
Sept. 17, 2020
Device type
Gas-Machine, Anesthesia
Root cause (FDA)
Process control
What the company told customers
GE will issue an Urgent Medical Device Correction letter to customers with affected flow sensors. This letter will inform the customer of the issue and provide safety instructions, including to inspect all inventory of Flow Sensors. GE Healthcare will replace the Flow Sensors indicated on returned customer response forms, if applicable. In the U.S., this letter will be sent by a traceable means. The letter will be sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, and Health Care Administrator / Risk Manager. GE Healthcare will document customer acknowledgment of the receipt and understanding of the letter, and the replacement of Flow Sensors, if needed.
Recall number
Z-0317-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0311-2021, Z-0312-2021, Z-0313-2021, Z-0314-2021, Z-0315-2021, Z-0316-2021, Z-0317-2021. The agency record governs; check it and the company before acting.