Vitros Immunodiagnostic Products CEA REAGENT PACK…

Jan. 6, 2004

Date posted by the FDA
Jan. 6, 2004
Status
Terminated
Filed by
Ortho-Clinical Diagnostics
Product
Vitros Immunodiagnostic Products CEA REAGENT PACK, Lot 750, REF 192 0115, 100 coated wells per pack. Firm on the label: Ortho-Clinical Diagnostics, Amersham, UK.
Reason
Reagent packs may contain wells that produce a low light signal that could result in calibration failures or negatively biased results.
Distribution pattern
154 reagent packs shipped to consignees in the U.S. 370 reagent packs shipped to Ortho/J&J International Affiliates for further distribution to foreign users. Foreign consignees are located in Argentina, Australia, Belgium, Brazil, Chile, China, France, Germany, India, Italy, Japan, Mexico, Panama, Poland, Portugal, Puerto Rico, Singapore, Spain, United Kingdom, and Venezuela.
Quantity
524 reagent packs
Codes and lots
Lot #750, Exp. 14 January 2004
Recall started
Dec. 3, 2003
Device type
System, Test, Carcinoembryonic Antigen
Root cause (FDA)
Other
What the company told customers
U.S. customers notified via FED EX letter sent on 12/3/03. Letter includes modified QC procedures to use until replacement product is issued. Foreign affiliates notified via e-mail on 12/3/03 with instructions to notify customers.
Recall number
Z-0299-04

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0299-04. The agency record governs; check it and the company before acting.