FDA · Medical device recall

Neptune Plus…

Nov. 25, 2015 Class II

Date
Nov. 25, 2015
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
TZ Medical Inc.
Product
Neptune Plus, HEMOSTASIS PAD WITH ANTIMICROBIAL BARRIER, For hemostasis following needle/sheath removal, Part number 8870-03
Reason
The IFU of the Neptune Plus Pads, does not match that submitted in the original 510k. Need to remove: for haemostasis following needle/sheath removal
Distribution pattern
US nationwide distribution.
Quantity
32,170 units
Codes and lots
Lot Numbers: AGR554; AGR554\20; AGR554\24; AGR554/24; AGR570\28; AGR570/43; AGR744124; AGR744137.
Recall started
Oct. 14, 2015
Device type
Dressing, Wound, Drug
Root cause (FDA)
Labeling False and Misleading
What the company told customers
TZ Medical Inc. began sending the Urgent: Medical Device Correction 8870-03 Neptune Plus letter, dated 10-7-2015, to consignees in October 2015. The firm is notifying all users of Neptune Plus wound dressing to a correction to the Instructions for Use (IFU). Consignees should discard the IFU in your current inventory and replace with new IFU. Consignees with questions should call TZ Medical Inc. at (503) 639-0282.
Recall number
Z-0293-2016

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-0293-2016. The agency record governs; check it and the company before acting.