FDA · Medical device recall
Neptune Plus…
- Date
- Nov. 25, 2015
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- TZ Medical Inc.
- Product
- Neptune Plus, HEMOSTASIS PAD WITH ANTIMICROBIAL BARRIER, For hemostasis following needle/sheath removal, Part number 8870-03
- Reason
- The IFU of the Neptune Plus Pads, does not match that submitted in the original 510k. Need to remove: for haemostasis following needle/sheath removal
- Distribution pattern
- US nationwide distribution.
- Quantity
- 32,170 units
- Codes and lots
- Lot Numbers: AGR554; AGR554\20; AGR554\24; AGR554/24; AGR570\28; AGR570/43; AGR744124; AGR744137.
- Recall started
- Oct. 14, 2015
- Device type
- Dressing, Wound, Drug
- Root cause (FDA)
- Labeling False and Misleading
- What the company told customers
- TZ Medical Inc. began sending the Urgent: Medical Device Correction 8870-03 Neptune Plus letter, dated 10-7-2015, to consignees in October 2015. The firm is notifying all users of Neptune Plus wound dressing to a correction to the Instructions for Use (IFU). Consignees should discard the IFU in your current inventory and replace with new IFU. Consignees with questions should call TZ Medical Inc. at (503) 639-0282.
- Recall number
- Z-0293-2016