FDA · Medical device recall

VITROS Immunodiagnostic Products Anti-HBc IgM Controls…

Nov. 24, 2021 Class III

Date
Nov. 24, 2021
Risk class
Class III (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences)
Status
Terminated
Filed by
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Immunodiagnostic Products Anti-HBc IgM Controls, Product Code 6801427 For use in monitoring the performance of the VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and the VITROS 5600/XT 7600 Integrated Systems
Reason
Ortho Clinical Diagnostics confirmed a complaint that some VITROS Immunodiagnostic Products Anti-HBc IgM Controls Lot 1690 Controls Values booklet contained information for the incorrect product. While the accuracy of patient test results is not affected by this issue, there is a potential for delayed results due to the troubleshooting of Quality Control results.
Distribution pattern
US (Domestic) distribution to states of: AZ, CA, CO, CT, FL, GA, LA, MA, MO, NJ, NY, OK, SC, SD, TX, UT, WA, WI, and WV. OUS (Foreign) distribution to Colombia.
Quantity
191 units
Codes and lots
Lot 1690 UDI 10758750001200
Recall started
Oct. 15, 2021
Device type
Single (Specified) Analyte Controls (Assayed And Unassayed)
Root cause (FDA)
Under Investigation by firm
What the company told customers
On October 15, 2021, the firm issued "Important Product Correction Notification" letters to affected customers. Customers were informed that up to 8 booklets in the affected lot contained the range of means information for VITROS Immunodiagnostic Products Anti-HBc Controls instead of the control values for VITROS Anti-HBc IgM. Customers were provided with a copy of the correct Controls Values booklet. Customers were asked to do the following: - Complete and return the included Confirmation of Receipt - Response Required customer response form by email to: ra-ocdus-confirmad@orthoclinicaldiagnostics.com or fax to: 1-888-557-3759 or 1-585-453-4110 - Retain the included Controls Values booklet to use in place of any recalled booklets - Forward the notification if the product was distributed outside your facility If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Recall number
Z-0247-2022