Monitor suspensions used with Advantx Legacy…

Oct. 27, 2008

Date posted by the FDA
Oct. 27, 2008
Status
Terminated
Filed by
GE Healthcare
Product
Monitor suspensions used with Advantx Legacy, RFX Classical, SFX-90, Model # 46-240838G2, manufactured after March 8, 1985 for Advantx Legacy, RFX and SFX systems. General-purpose radiographic system component used to suspend the system's image monitor.
Reason
GE Healthcare has recently become aware of a potential risk of monitor dropping associated with the monitor suspension of the Advantx system that may impact patient safety. It has been reported that suspensions manufactured after March 8, 1995 did not have the required thread locking agent applied to the setscrew, allowing the setscrew to back out over time, which can lead to the monitor dropping
Distribution pattern
Worldwide Distribution --- USA including states of AZ, CA, FL, GA, IA, MD, MO, NC, NY, OH, OK, TN, TX,WI, and LA, and country of Slovak Republic.
Quantity
21
Codes and lots
00000486566WK3, 00000486586WK1, 00000486555WK6, 00000486573WK9, 00000486581WK2, 00000478555WK6, 00000486579WK6, 00000486588WK7, 00000486557WK2, 00000486560WK6, 00000486550WK7, 00000486556WK4, 00000486547WK3, 00000056366VP8, 00000486544WK0, 00000486541WK6, 00000486543WK2, 00000471291WK5, 00000486568WK9, and 00000486559WK8
Recall started
March 9, 2007
Device type
Generator, High-Voltage, X-Ray, Diagnostic
Root cause (FDA)
Component change control
What the company told customers
Consignees were sent a GE Healthcare "Product Safety Notification" letter dated September 13, 2007. The letter was addressed to Hospital Administrators/Managers of Radiology/Radiologists. The letter described the Safety Issue, Affected Product, Safety Instructions, Corrective Action and Contact Information. The letter recommends that customers move the monitor suspension to an open area in the room where viewing is still possible and make sure to minimize the movement of the monitor until a Qualified GE Healthcare Field Engineer has inspected the suspension. The letter also states that a Qualified GE Healthcare Field Engineer will schedule a field visit to inspect the monitor suspension. Customers should contact their sales or service rep is they have any questions.
Recall number
Z-0200-2009