FDA · Medical device recall

GE Healthcare Optima XR 200amx with digital upgrade Mobile general… and 3 more products

Nov. 28, 2012 Class II

Date
Nov. 28, 2012
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Reason
It was discovered GE Healthcare has become aware of a firmware issue on the mobile radiographic product, Brivo XR285amx, Optima XR200amx, Optima XR22Oamx and Optima XR200amx with Digital Upgrade, related to Technique Accuracy as mandated by the US Code of Federal Regulations (21CFR) and IEC 60601-2-54.
Distribution pattern
Nationwide Distribution

4 products in this recall

GE Healthcare Optima XR 200amx with digital upgrade Mobile general purpose radiographic imaging of the human…

Product
GE Healthcare Optima XR 200amx with digital upgrade Mobile general purpose radiographic imaging of the human head and body.
Quantity
364 total units installed in US
Codes and lots
part number: 5421698
Recall started
Sept. 20, 2012
Device type
Image-Intensified Fluoroscopic X-Ray System, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue. Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard. Further questions please call (262) 513-4122.
Recall number
Z-0197-2013

GE Healthcare Optima XR 220amx Mobile general purpose radiographic imaging of the human head and body

Product
GE Healthcare Optima XR 220amx Mobile general purpose radiographic imaging of the human head and body.
Quantity
364 total units installed in US
Codes and lots
part numbers: 5555000-5, 5555000-6
Recall started
Sept. 20, 2012
Device type
Image-Intensified Fluoroscopic X-Ray System, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue. Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard. Further questions please call (262) 513-4122.
Recall number
Z-0198-2013

GE Healthcare Optima 200amx Mobile general purpose radiographic imaging of the human head and body

Product
GE Healthcare Optima 200amx Mobile general purpose radiographic imaging of the human head and body.
Quantity
364 total units installed in US
Codes and lots
part numbers: 5555000-3, 5555000-4
Recall started
Sept. 20, 2012
Device type
Image-Intensified Fluoroscopic X-Ray System, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue. Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard. Further questions please call (262) 513-4122.
Recall number
Z-0199-2013

GE Healthcare Brivo XR 285 Mobile general purpose radiographic imaging of the human head and body

Product
GE Healthcare Brivo XR 285 Mobile general purpose radiographic imaging of the human head and body.
Quantity
364 total units installed in US
Codes and lots
part numbers:5555000, 5555000-2
Recall started
Sept. 20, 2012
Device type
Image-Intensified Fluoroscopic X-Ray System, Mobile
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
GE Healthcare will remedy the defect or bring the product into compliance with each applicable Federal and IEC standard. A Service Representative will update the firmware on the system to address the issue. Field Modification Instructions (FMI) 10870, 10872, and 10873 are describing the rework plan the GEHC Engineers will make to the imaging systems compliant with the applicable performance standard. Further questions please call (262) 513-4122.
Recall number
Z-0200-2013

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0197-2013, Z-0198-2013, Z-0199-2013, Z-0200-2013. The agency record governs; check it and the company before acting.