U.S. safety records, as the agencies wrote them. Independent.
FDA · Medical device recall

OPMI LUMERA 300…

Recall started · FDA enforcement report Nov. 1, 2023 Class II

Reason
Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components.
Status
Ongoing
Distribution
US Nationwide distribution in the states of MD, TX, CA, GA, LA.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Nov. 1, 2023
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA has not recorded a termination for this recall; FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
Carl Zeiss Suzhou Co., Ltd.
Product
OPMI LUMERA 300, REF 6137
Reason
Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components.
Distribution pattern
US Nationwide distribution in the states of MD, TX, CA, GA, LA.
Quantity
8
Codes and lots
UDI-DI/Serial Numbers: 06909262061375/6137104455, 6137104627, 6137104623, 6137104678, 6137104233, 6137104620, 6137104480, 6137104613
Recall started
Aug. 18, 2023
Voluntary or mandated
Voluntary: Firm initiated
Device type
Microscope, Surgical, General & Plastic Surgery
Root cause (FDA)
Process control
What the company told customers
On 8/18/23, correction notices were mailed to customers who were asked to do the following: Before each use of the affected device, perform the following mandatory inspection: Refer to pictures in the notice showing a gap between the OPMI and the inspection arm. This gap should not be greater than 5mm. If the gap is greater than 5mm, please Immediately stop using the device! In this case the affected device can no longer be used. The firm is planning to inspect (repair if needed) the suspension arm immediately. Service staff will contact you to arrange an appointment for inspection (repair if needed) of the suspension arm. Complete and return the acknowledgement and receipt form and return to USFCA.meditec.us@zeiss.com If you have any questions, contact firm Customer Support at 877-486-7473.
Recall number
Z-0183-2024

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device or a product, the reason the record gives, the record's wordsA simplified drawing for a device record; the mark follows the reason the record gives, not a photograph of the product. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: OPMI LUMERA 300, REF 61372 The record describes a part that is missing3 What the record says4 What the record listsFOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepmay be missing

“Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components.”OPMI LUMERA 300, REF 6137, FDA enforcement report, 2023-11-01. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Microscope, Surgical, General & Plastic Surgery

Distributed in (FDA's record): California Georgia Louisiana Maryland Texas and nationwide

This recall among the others

  • FDA's enforcement report came 75 days after the firm started the recall. For half of the 835 FDA medical device recall events reported in 2023 with a start date, it took 51 days or fewer (by year).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-0183-2024. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-0183-2024, Nov. 1, 2023. "OPMI LUMERA 300." PlainRecords, https://plainrecords.com/recall/device/z-0183-2024/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance