FDA · Medical device recall

BioPro Femoral Head (Metal 22… and 5 more products

Recall started · FDA enforcement report Oct. 22, 2025 Class II

Reason
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
Status
Ongoing
Distribution
US Domestic distribution to Texas and Michigan.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Oct. 22, 2025
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
BioPro, Inc.
Reason
Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
Distribution pattern
US Domestic distribution to Texas and Michigan.

What the record says about this device

A simplified drawing of a device, following the words of FDA's record. It shows the reason the record gives, not a photograph of the product.

Four steps: a device, the reason the record gives, the record's wordsA simplified drawing of a medical device with a mark for the reason the record gives. Left, for comparison: the same drawing without what the record names. Right: this record.1 The product: BioPro Femoral Head (Metal 22, 28, 32…2 The record describes a seal or package that may not hold3 What the record says4 What the record listsSEALSEALFOR COMPARISONthe same drawing, withoutwhat the record namesTHIS RECORDwhat the record names,step by stepcompromise the sterile…

“Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.”BioPro Femoral Head (Metal 22, 28, 32, 36, 40) cobalt chromium and ceramic heads Part ID/Description: 18593 HEAD FEMORAL 22MM -3 18594 HEAD FEMORAL 22MM +0 19003 HEAD FEMORAL 28MM-6 19004 HEAD FEMORAL 28MM-3 19005 HEAD FEMORAL 28MM+0 19006 HEAD FEMORAL 28MM+3 19007 HEAD FEMORAL 28MM+6 19008 HEAD FEMORAL 28MM+9 19053 HEAD FEMORAL 36MM -3 19054 HEAD FEMORAL 36MM +0 19055 HEAD FEMORAL 36MM +3 19056 HEAD FEMORAL 36MM +6 19057 HEAD FEMORAL 36MM +9 19130 HEAD FEMORAL 32MM-6 19131 HEAD FEMORAL 32MM-3 19132 HEAD FEMORAL 32MM+0 19133 HEAD FEMORAL 32MM +3 19134 HEAD FEMORAL 32MM+6 19135 HEAD FEMORAL 32MM +9, FDA enforcement report, 2025-10-22. The caption is the record's own words.

A generic device, not the product. The mark follows the record's own words.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Distributed in (FDA's record): Michigan Texas

6 products in this recall

BioPro Femoral Head (Metal 22…

Product
BioPro Femoral Head (Metal 22, 28, 32, 36, 40) cobalt chromium and ceramic heads Part ID/Description: 18593 HEAD FEMORAL 22MM -3 18594 HEAD FEMORAL 22MM +0 19003 HEAD FEMORAL 28MM-6 19004 HEAD FEMORAL 28MM-3 19005 HEAD FEMORAL 28MM+0 19006 HEAD FEMORAL 28MM+3 19007 HEAD FEMORAL 28MM+6 19008 HEAD FEMORAL 28MM+9 19053 HEAD FEMORAL 36MM -3 19054 HEAD FEMORAL 36MM +0 19055 HEAD FEMORAL 36MM +3 19056 HEAD FEMORAL 36MM +6 19057 HEAD FEMORAL 36MM +9 19130 HEAD FEMORAL 32MM-6 19131 HEAD FEMORAL 32MM-3 19132 HEAD FEMORAL 32MM+0 19133 HEAD FEMORAL 32MM +3 19134 HEAD FEMORAL 32MM+6 19135 HEAD FEMORAL 32MM +9
Quantity
147 total
Codes and lots
Part ID/ UDI-DI: 18593 M20918593 18594 M20918594 19003 M20919003 19004 M20919004 19005 M20919005 19006 M20919006 19007 M20919007 19008 M20919008 19053 M20919053 19054 M20919054 19055 M20919055 19056 M20919056 19057 M20919057 19130 M20919130 19131 M20919131 19132 M20919132 19133 M20919133 19134 M20919134 19135 M20919135 All lots and serial numbers affected
Recall started
Sept. 9, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Package design/selection
What the company told customers
Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
Recall number
Z-0148-2026

BioPro Endo Head (Metal uni-polar…

Product
BioPro Endo Head (Metal uni-polar, 38-55) Product Number/Description 10179 ENDO MD 38 10180 ENDO MD 41 10181 ENDO MD 43 10182 ENDO MD 45 10183 ENDO MD 47 10184 ENDO MD 49 10185 ENDO MD 51 10186 ENDO MD 53 10187 ENDO MD 55 10188 ENDO SH 38 10189 ENDO SH 41 10190 ENDO SH 43 10191 ENDO SH 45 10192 ENDO SH 47 10193 ENDO SH 49 10194 ENDO SH 51 10195 ENDO SH 53 10196 ENDO SH 55
Quantity
147 total
Codes and lots
Product ID/UDI-DI 10179 M20910179 10180 M20910180 10181 M20910181 10182 M20910182 10183 M20910183 10184 M20910184 10185 M20910185 10186 M20910186 10187 M20910187 10188 M20910188 10189 M20910189 10190 M20910190 10191 M20910191 10192 M20910192 10193 M20910193 10194 M20910194 10195 M20910195 10196 M20910196 All lots, all serial numbers
Recall started
Sept. 9, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Root cause (FDA)
Package design/selection
What the company told customers
Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
Recall number
Z-0149-2026

BioPro Bipolar Head Product ID/Description 18130 BIPOLAR HEAD 38MM 18131 BIPOLAR HEAD 39MM 18132 BIPOLAR HEAD…

Product
BioPro Bipolar Head Product ID/Description 18130 BIPOLAR HEAD 38MM 18131 BIPOLAR HEAD 39MM 18132 BIPOLAR HEAD 40MM 18133 BIPOLAR HEAD 41MM 18134 BIPOLAR HEAD 42MM 18135 BIPOLAR HEAD 43MM 18136 BIPOLAR HEAD 44MM 18137 BIPOLAR HEAD 45MM 18138 BIPOLAR HEAD 46MM 18139 BIPOLAR HEAD 47MM 18140 BIPOLAR HEAD 48MM 18141 BIPOLAR HEAD 49MM 18142 BIPOLAR HEAD 50MM 18143 BIPOLAR HEAD 51MM 18144 BIPOLAR HEAD 52MM 18145 BIPOLAR HEAD 53MM 18146 BIPOLAR HEAD 54MM 18147 BIPOLAR HEAD 55MM 18148 BIPOLAR HEAD 56MM 18149 BIPOLAR HEAD 57MM 18150 BIPOLAR HEAD 58MM 18151 BIPOLAR HEAD 59MM 18152 BIPOLAR HEAD 60MM
Quantity
147 total
Codes and lots
Product ID/UDI-DI 18130 M20918130 18131 M20918131 18132 M20918132 18133 M20918133 18134 M20918134 18135 M20918135 18136 M20918136 18137 M20918137 18138 M20918138 18139 M20918139 18140 M20918140 18141 M20918141 18142 M20918142 18143 M20918143 18144 M20918144 18145 M20918145 18146 M20918146 18147 M20918147 18148 M20918148 18149 M20918149 18150 M20918150 18151 M20918151 18152 M20918152 All lots, all serial numbers
Recall started
Sept. 9, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Hemi-, Femoral, Metal/Polymer, Cemented Or Uncemented
Root cause (FDA)
Package design/selection
What the company told customers
Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
Recall number
Z-0150-2026

BioPro Femoral Head (Ceramic heads 28-30) Product ID/Description 19023 HEAD FEMORAL CERAMIC 28MM -3 19024…

Product
BioPro Femoral Head (Ceramic heads 28-30) Product ID/Description 19023 HEAD FEMORAL CERAMIC 28MM -3 19024 HEAD FEMORAL CERAMIC 28MM +0 19025 HEAD FEMORAL CERAMIC 32MM -3 19026 HEAD FEMORAL CERAMIC 32MM +0 19027 HEAD FEMORAL CERAMIC 32MM +3
Quantity
147 total
Codes and lots
Product ID/UDI-DI 19023 M20919023 19024 M20919024 19025 M20919025 19026 M20919026 19027 M20919027 All lots, all serial numbers
Recall started
Sept. 9, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
Root cause (FDA)
Package design/selection
What the company told customers
Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
Recall number
Z-0151-2026

BioPro Endo Head (Ceramic) Product ID/Description 13006 ENDO MODULAR CERAMIC SH 38 13012 ENDO MODULAR CERAMIC…

Product
BioPro Endo Head (Ceramic) Product ID/Description 13006 ENDO MODULAR CERAMIC SH 38 13012 ENDO MODULAR CERAMIC SH 51 13013 ENDO MODULAR CERAMIC MD 38 13019 ENDO MODULAR CERAMIC MD 51 13007 ENDO MODULAR CERAMIC SH 41 13008 ENDO MODULAR CERAMIC SH 43 13009 ENDO MODULAR CERAMIC SH 45 13010 ENDO MODULAR CERAMIC SH 47 13011 ENDO MODULAR CERAMIC SH 49 13014 ENDO MODULAR CERAMIC MD 41 13015 ENDO MODULAR CERAMIC MD 43 13016 ENDO MODULAR CERAMIC MD 45 13017 ENDO MODULAR CERAMIC MD 47 13018 ENDO MODULAR CERAMIC MD 49 14823 ENDO MODULAR CERAMIC SH 53 14824 ENDO MODULAR CERAMIC MD 53 14825 ENDO MODULAR CERAMIC SH 55 14826 ENDO MODULAR CERAMIC MD 55 17295 ENDO MODULAR CERAMIC SH 36 17296 ENDO MODULAR CERAMIC MD 36
Quantity
147 total
Codes and lots
Product ID/UDI-DI 13006 M20913006 13012 M20913012 13013 M20913013 13019 M20913019 13007 M20913007 13008 M20913008 13009 M20913009 13010 M20913010 13011 M20913011 13014 M20913014 13015 M20913015 13016 M20913016 13017 M20913017 13018 M20913018 14823 M20914823 14824 M20914824 14825 M20914825 14826 M20914826 17295 M20917295 17296 M20917296 All lots, all serial numbers
Recall started
Sept. 9, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Hemi-, Femoral, Metal Ball
Root cause (FDA)
Package design/selection
What the company told customers
Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
Recall number
Z-0152-2026

BioPro Femoral Head Product ID/Description 14089 HEAD FEMORAL 28 SH LT 14090 HEAD FEMORAL 28 MD LT 14091 HEAD…

Product
BioPro Femoral Head Product ID/Description 14089 HEAD FEMORAL 28 SH LT 14090 HEAD FEMORAL 28 MD LT 14091 HEAD FEMORAL 28 LG LT 14094 HEAD FEMORAL 32 SH LT 14095 HEAD FEMORAL 32 MD LT 14096 HEAD FEMORAL 32 LG LT
Quantity
147 total
Codes and lots
Product ID/UDI-DI 14089 M20914089 14090 M20914090 14091 M20914091 14094 M20914094 14095 M20914095 14096 M20914096 All lots, all serial numbers
Recall started
Sept. 9, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Root cause (FDA)
Package design/selection
What the company told customers
Consignees were notified on September 9, 2025, via phone, followed by formal letters with instructions for returning the inventory to BioPro.
Recall number
Z-0153-2026

This recall among the others

  • FDA's enforcement report came 43 days after the firm started the recall. For half of the 972 FDA medical device recall events reported in 2025 with a start date, it took 44 days or fewer (by year).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement reports, recall numbers Z-0148-2026, Z-0149-2026, Z-0150-2026, Z-0151-2026, Z-0152-2026, Z-0153-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-0148-2026, Oct. 22, 2025. "BioPro Femoral Head (Metal 22… and 5 more products." PlainRecords, https://plainrecords.com/recall/device/z-0148-2026/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance