FDA · Medical device recall
System… and 4 more products
- Date
- Oct. 28, 2020
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- GE Healthcare, LLC
- Reason
- Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.
- Distribution pattern
- Worldwide distribution - US Nationwide distribution including in the states of FL, KY, NC, NJ, NM, NY, TX, UT and WA. The countries of Canada, Hungary, India, Korea, China, Russia, Taiwan, Puerto Rico and Ubekistan.
5 products in this recall
System…
- Product
- System, Nuclear Magnetic Resonance Imaging - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
- Quantity
- 30 devices total
- Codes and lots
- Model: 5741000 and 5741000-20
- Recall started
- Sept. 8, 2020
- Device type
- System, Nuclear Magnetic Resonance Imaging
- Root cause (FDA)
- Process change control
- What the company told customers
- On 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction
- Recall number
- Z-0127-2021
System…
- Product
- System, Nuclear Magnetic Resonance Imaging - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
- Quantity
- 30 devices total
- Codes and lots
- Model: 5741000-23 and 5396469
- Recall started
- Sept. 8, 2020
- Device type
- System, Nuclear Magnetic Resonance Imaging
- Root cause (FDA)
- Process change control
- What the company told customers
- On 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction
- Recall number
- Z-0128-2021
System…
- Product
- System, Nuclear Magnetic Resonance Imaging - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
- Quantity
- 30 Devices total
- Codes and lots
- Model: 5352293, 5726207-5, 5741000-25 and 5741000-29 .
- Recall started
- Sept. 8, 2020
- Device type
- System, Nuclear Magnetic Resonance Imaging
- Root cause (FDA)
- Process change control
- What the company told customers
- On 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction
- Recall number
- Z-0129-2021
System…
- Product
- System, Nuclear Magnetic Resonance Imaging - Product Usage: is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times.
- Quantity
- 30 devices Total
- Codes and lots
- Model: 5148810-3, 5741000-100, 5741000-26 and 5741000-27
- Recall started
- Sept. 8, 2020
- Device type
- System, Nuclear Magnetic Resonance Imaging
- Root cause (FDA)
- Process change control
- What the company told customers
- On 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction
- Recall number
- Z-0130-2021
Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance…
- Product
- Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance - Product Usage: The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging.
- Quantity
- N/A
- Codes and lots
- Model: 5726207-8 and 5741000-28
- Recall started
- Sept. 8, 2020
- Device type
- Tomographic Imager Combining Emission Computed Tomography With Nuclear Magnetic Resonance
- Root cause (FDA)
- Process change control
- What the company told customers
- On 09/08/20 the firm sent a Customer Letter to its consignees with the following instructions: You can continue to use the system. In the event the PGR cabinet needs to be serviced, power down the PGR at the Main Disconnect Panel (MDP) then follow MDP Lockout/Tagout from the current released service methods and documentation. GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction
- Recall number
- Z-0131-2021
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-0127-2021, Z-0128-2021, Z-0129-2021, Z-0130-2021, Z-0131-2021. The agency record governs; check it and the company before acting.