FDA · Medical device recall
Disposable Smooth Interior Tubing Material: Polyethylene EVA copolymer…
- Date
- Oct. 24, 2018
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Terminated
- Filed by
- A M Systems Inc
- Product
- Disposable Smooth Interior Tubing Material: Polyethylene EVA copolymer. This tubing is supplied non-sterile P/N: 1024920 Rev. 01 Low temperature smooth interior tubing is a disposable pulmonary equipment interconnect used in ventilation and aesthesia machines.
- Reason
- Some of the Disposable Smooth Interior Tubing may pull apart when stretched.
- Distribution pattern
- US Distribution to states of: OH, PA, and NC.
- Quantity
- 610
- Codes and lots
- Product Code: 243301 Lot Numbers: 572015 574717 578973 579100 582567
- Recall started
- Aug. 10, 2018
- Device type
- Tubing, Pressure And Accessories
- Root cause (FDA)
- Process control
- What the company told customers
- On 08/10/2018, the firm, A-M Systems, sent an "URGENT: Notification of Product Recall/Correction" letter to customers via email informing them that it was determined that some of the Disposable Smooth Interior Tubing may pull apart when stretched. Customers were advised to provide the firm with the number of items distributed from the affected lots and the number of items still remaining in their inventory from these lots so that the firm could accurately track the status of this correction action. The customers are instructed to please use the enclosed form and self-addressed envelope to return the information to the firm immediately. Once the customers have provided the firm with an accurate count of items distributed and items remaining in the customer's control please destroy or discard any remaining product in your possession from these lot numbers. If you have any questions, please contact Quality Manager at Tel: 360.683.8300 x1302 or Email: rthompson@a-msystems.com.
- Recall number
- Z-0120-2019
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0120-2019. The agency record governs; check it and the company before acting.