FDA · Medical device recall

OPMI VARIO 700 Surgical Microscope…

Oct. 24, 2011

Date posted by the FDA
Oct. 24, 2011
Status
Terminated
Filed by
Carl Zeiss Meditec, Inc.
Product
OPMI VARIO 700 Surgical Microscope. Manufactured by: Carl Zeiss Meditec AG, Oberkochen, Germany. Provide a magnified view of the surgical field.
Reason
An incorrect nut and bolt set was used in the axis three (3) spring assembly of the OPMI VARIO 700 surgical microscope distributed to three US locations.
Distribution pattern
USA (nationwide) including the states of FL, GA, and LA .
Quantity
3
Codes and lots
Serial numbers: 6636101067; 6636101068; 6636101070.
Recall started
Sept. 26, 2011
Device type
Microscope, Surgical
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
Carl Zeiss Meditec had Field Service Engineers directly contact all affected customers. The Field Service Engineers will replace the defective component. Questions regarding this Field Corrective Action should be directed to the Director of Clinical/Regulatory Affairs at 925-557-4616.