FDA · Medical device recall
OPMI VARIO 700 Surgical Microscope…
- Date posted by the FDA
- Oct. 24, 2011
- Status
- Terminated
- Filed by
- Carl Zeiss Meditec, Inc.
- Product
- OPMI VARIO 700 Surgical Microscope. Manufactured by: Carl Zeiss Meditec AG, Oberkochen, Germany. Provide a magnified view of the surgical field.
- Reason
- An incorrect nut and bolt set was used in the axis three (3) spring assembly of the OPMI VARIO 700 surgical microscope distributed to three US locations.
- Distribution pattern
- USA (nationwide) including the states of FL, GA, and LA .
- Quantity
- 3
- Codes and lots
- Serial numbers: 6636101067; 6636101068; 6636101070.
- Recall started
- Sept. 26, 2011
- Device type
- Microscope, Surgical
- Root cause (FDA)
- Nonconforming Material/Component
- What the company told customers
- Carl Zeiss Meditec had Field Service Engineers directly contact all affected customers. The Field Service Engineers will replace the defective component. Questions regarding this Field Corrective Action should be directed to the Director of Clinical/Regulatory Affairs at 925-557-4616.
- Recall number
- Z-0105-2012