FDA · Medical device recall

GE ApexPro CH Telemetry System…

Oct. 20, 2006

Date posted by the FDA
Oct. 20, 2006
Status
Terminated
Filed by
Ge Healthcare
Product
GE ApexPro CH Telemetry System. The ApexPro Telemetry System is composed of 6 major components. -The patient worn data acquisition transmitters; -The receiver antenna system infrastructure; -The receivers; -The receiver subsystem; -A computer platform running the Apex Pro Telemetry Application; -A computer platform running a central station application (which may be the same computer platform running the ApexPro Telemetry Application)
Reason
When a patient being monitored with ApexPro/Apex Pro CH Telemetry with ST monitoring disabled is in a pre-existing condition of continuous MESSAGE or ADVISORY level alarm preceding a SYSTEM WARNING level alarm, the SYSTEM WARNING audible alarm and flashing yellow border around the patient panel at the CIC (Clinical Information Center) does not occur.
Distribution pattern
Worldwide-USA and Puerto Rico and countries of ,Australia, Belgium, Canada, China, Croatia, Czech Republic, Estonia, Finland, France, Germany, Great Britain, Greece, Guam, Honduras, Hong Kong, Hunary, India, Indonesia, Ireland, Israel, Italy, Jordan, Korea, Kuwait, Malaysia, Malta, Mexico, Netherlands, New Zealand, Norway, Pakistan, Panama, Philippines, Poland, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerlant, Syria, Taiwan, Thailand, Tunisia, and Venezuela.
Quantity
5,203
Codes and lots
Software version 3.8 and earlier
Recall started
Aug. 4, 2006
Device type
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Root cause (FDA)
Other
What the company told customers
An Urgent Medical Device Correction letter dated August 04, 2006 was sent to customers with affected devices. The letter described the issue affecting devices containing software versions v3.8 and earlier, provides short term recommendations and a long term solution of updating the software.
Recall number
Z-0055-2007