FDA · Medical device recall
Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product…
- Date
- Oct. 11, 2023
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- Blue Belt Technologies, Inc
- Product
- Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299
- Reason
- CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis
- Distribution pattern
- US Nationwide distribution.
- Quantity
- 56 units
- Codes and lots
- UDI: (01) 00885556773338 Lot Numbers: 51141585, 51108475, 51108689
- Recall started
- Sept. 7, 2023
- Device type
- Orthopedic Stereotaxic Instrument
- Root cause (FDA)
- Software Design Change
- What the company told customers
- Smith & Nephew, Inc. (Blue Belt Technologies) issued Urgent Medical Device Correction Notice (C-2023-08) on September 7, 2023 via overnight mail. Letter states reason for recall, health risk and action to take: Required Actions for user/hospital: 1. Complete Correction Notice Acknowledgement Form and submit to FieldActions@smith-nephew.com. 2. Prior to use of software, contact your sales representative to update settings. 3. Upon completion of settings update, you may use the CORI Surgical System until the software update is deployed. 4. A Smith & Nephew Robotics Representative will be deployed to your site to implement a new software version to correct the issue once it is available. Email: FieldActions@smith-nephew.com Actions for Smith & Nephew Representatives 1. Work with customer to update CORI settings to prevent incorrect configuration. 2. Complete the attached Correction Notice Acknowledgement Form. 3. Upon completion of software upgrade, submit CORI In-Field Service Record to Field Action Team.
- Recall number
- Z-0041-2024
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement report, recall number Z-0041-2024. The agency record governs; check it and the company before acting.