FDA · Medical device recall

ClariTEE Miniaturized TEE probe…

Recall started · FDA enforcement report Oct. 15, 2025 Class II

Reason
Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
Status
Ongoing
Distribution
US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.

This box checks this one record only. Other recalls of the same product or company are on the brand page.

Enforcement report date
Oct. 15, 2025
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing (FDA's definition: a recall which is currently in progress; as in openFDA's data of Nov. 10, 2026)
Filed by
ImaCor Inc.
Product
ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.
Reason
Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
Distribution pattern
US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.
Quantity
45 units
Codes and lots
Model CLT-010; UDI-DI 00861589000108; Lot number: 20261209.
Recall started
Aug. 29, 2025
Voluntary or mandated
Voluntary: Firm initiated
Device type
Transducer, Ultrasonic, Diagnostic
Root cause (FDA)
Process control
What the company told customers
ImaCor, Inc. notified consignees on 08/29/2025 via email. Consignees were instructed to Immediately discontinue use of the affected ClariTEE probes, quarantine all affected units in inventory, confirm receipt of this notice, return all affected units to ImaCor, and complete and return the response form.
Recall number
Z-0038-2026

What this record is about

A generic icon of the kind of thing the record is about; the record's own words are below. It is not a picture of the product.

A generic icon of the kind of thing the record is aboutA simplified icon of the kind of thing the record is about and the phrases of the record's own text beside it.1 The record: ClariTEE Miniaturized TEE probe. Model…2 In the record's words:3 Only the kind of thing is drawn4 The record's own words are below“Production assembly error resulting in the potential for reversedarticulation, where moving the lever forward caused the probe tip…”CLARITEE MINIATURIZED TEE…

“Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.”ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH., FDA enforcement report, 2025-10-15. The caption is the record's own words.

The picture shows the kind of thing the record is about (a part, a product, a package), not the product itself; the ring marks it. Nothing in it is a claim beyond the record's own words below.

If the drawing and the record's words differ, the words govern. A drawing is chosen by rules and an AI reading of the record and can miss what it means; report one with the form below.

FDA device name: Transducer, Ultrasonic, Diagnostic

Distributed in (FDA's record): Florida Georgia New Jersey Pennsylvania South Carolina and nationwide

This recall among the others

  • FDA's enforcement report came 47 days after the firm started the recall. For half of the 972 FDA medical device recall events reported in 2025 with a start date, it took 44 days or fewer (by year).
  • ImaCor: 1 other recall or notice on this site in the three years before this one (from Oct. 16, 2022).

Counted from the agency records on this site, as on Statistics: a recall event once, by its first notice.

What should I do? How serious is a recall? Classes and warnings explained · How to read this record

Source: FDA · Medical device enforcement report, recall number Z-0038-2026. See this record in openFDA (FDA's own data service; it answers in JSON). The agency record governs; check it and the company before acting.

Cite this record
FDA recall record Z-0038-2026, Oct. 15, 2025. "ClariTEE Miniaturized TEE probe." PlainRecords, https://plainrecords.com/recall/device/z-0038-2026/ (as of the day it was built).

At each build a sample of records from each source is compared with the agency's published file and the page; every correction is public. Drawings are chosen by software that reads the record. Built daily. How we check · Corrections

Report a mistake on this page

Newest medical device recalls from other companies

  • Nov. 4, 2026Carevo…Class IIThe side support of the shower trolley may open unintentionally when the handles are not fully locked and additional force is applied during patient… Recall began Sept. 25, 2026
  • Nov. 4, 2026Precision Pointer Electromagnetic (8000-050-001)Class IIDevice may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the… Recall began Sept. 24, 2026
  • Nov. 4, 2026Nalu TD2 Therapy Disc with firmware…Class IITherapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a… Recall began Sept. 30, 2026
  • Nov. 4, 2026PENTAX Medical Ultrasound Endoscope… and 4 more productsClass IIWarnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and… Recall began Sept. 9, 2026

Check another brand or product

Recalled in the last 7 days: Game Changer Products · Eros Apothecary · Cummins (company) · Huffy · QIOMALA · SHEIN Distribution · Allstar Marketing Group · Arrow Reliance