GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human… and 1 more product

Jan. 20, 2010

Date posted by the FDA
Jan. 20, 2010
Status
Terminated
Filed by
Ge Healthcare
Reason
A software defect - may result in a false indication and warning on the OEC 9900 Workstation and C-arm display that the X-Ray tube Anode is over-heated. This will also cause the displayed air kerma rate and cumulative air kerma to be inaccurate.
Distribution pattern
Nationwide and Canadian Distribution

2 products in this recall

GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human anatomy

Product
GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human anatomy.
Quantity
1411 units total
Codes and lots
All associated serial numbers
Recall started
Aug. 12, 2009
Device type
System, X-Ray, Fluoroscopic, Image-Intensified
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
An Important Electronic Product Radiation Warning letter will be sent to customers identifying the issues, interim solution, and permanent solution. Questions or concerns are to be directed to the GEHC OEC service team at 1-800-874-7378.
Recall number
Z-0004-2010

GE OEC 9900 Elite GSP Fluoroscopes To produce mobile fluoroscopic images of human anatomy

Product
GE OEC 9900 Elite GSP Fluoroscopes To produce mobile fluoroscopic images of human anatomy.
Quantity
1411 units total
Codes and lots
All associated serial numbers
Recall started
Aug. 12, 2009
Device type
System, X-Ray, Fluoroscopic, Image-Intensified
Root cause (FDA)
Radiation Control for Health and Safety Act
What the company told customers
An Important Electronic Product Radiation Warning letter will be sent to customers identifying the issues, interim solution, and permanent solution. Questions or concerns are to be directed to the GEHC OEC service team at 1-800-874-7378.
Recall number
Z-0005-2010

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-0004-2010, Z-0005-2010. The agency record governs; check it and the company before acting.