FDA · Medical device recall

Carl Zeiss INTRABEAM PRS 500…

Oct. 11, 2017 Class II

Date
Oct. 11, 2017
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
Carl Zeiss Meditec AG
Product
Carl Zeiss INTRABEAM PRS 500, product #304534-0000-500. The firm name on the label is Carl Zeiss Meditec AG, Jena, Germany. The INTRABEAM System is indicated for radiation therapy treatments
Reason
The device had an incorrect printed calibration value on the calibration certificate.
Distribution pattern
Nationwide Distribution including AZ, CA, CT, DC, FL, GA, OH, and PA.
Quantity
9 devices
Codes and lots
Intrabeam ser. #6402501246, Ion chamber ser. #SN001254; Intrabeam ser. #6402501241, Ion chamber ser. #SN001745; Intrabeam ser. #6402100932, Ion chamber ser. #SN001994; Intrabeam ser. #6402100958, Ion chamber ser. #SN002096; Intrabeam ser. #6402101042, Ion chamber ser. #SN002265; Intrabeam ser. #6402101029, Ion chamber ser. #SN002270; Intrabeam ser. #6402101034, Ion chamber ser. #SN002302; Intrabeam ser. #6402101105, Ion chamber ser. #SN002651; and Intrabeam ser. #6402101130, Ion chamber ser. #SN002684.
Recall started
Aug. 30, 2017
Device type
System, Therapeutic, X-Ray
Root cause (FDA)
Process control
What the company told customers
Zeiss sent a Field Safety Corrective Action letter dated August 30. 2017, to all affected customers via FedEx. Recall letters were issued by the U.S. Agent on behalf of the /recalling firm on 8/30/2017 via FedEx. Customers were given the following instructions: 1. A member of the ZEISS service staff will contact you to coordinate the steps to be taken and to schedule the update of your system. Alternatively, you may update the system yourself by following the attached instructions. In case you update the system yourself, a ZEISS service staff member will still be visiting your site to confirm the updated calibration. 2. Please complete and return the attached response form to acknowledge receipt of this notification. If you have question regarding this field safety corrective action, you can contact your ZEISS customer service at 800-442-4020 or by email at iort.meditec.US@zeiss.com with any additional questions. You may also contact me at 925-557-4151 or viet.nguyen@zeiss.com
Recall number
Z-0001-2018