Menu
FDA · Medical device recall

Flow sensors kept as stand-alone user replaceable spare parts for the… and 1 more product

Oct. 6, 2021 Class II

Date
Oct. 6, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare, LLC
Reason
The flow sensors built prior to June 2021 could have damaged tubes with small punctures or cuts that could cause leaks resulting in incorrect anesthesia machine tidal volumes potentially leading to over-delivery of volume to the patient.
Distribution pattern
Distribution was nationwide, including PR. Foreign distribution was made to Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote D'lvoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Fiji, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Macedonia, Madagascar, Malaysia, Maldives, Mali, Malta, Mauritius, Mexico, Moldova, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Niger, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom Uruguay, Venezuela, Vietnam, and Yemen.

2 products in this recall

Flow sensors kept as stand-alone user replaceable spare parts for the GEHC anethesia machines: 2089610-001…

Product
Flow sensors kept as stand-alone user replaceable spare parts for the GEHC anethesia machines: 2089610-001 FLOW SENSOR, LEGACY VAR ORF BCG (blue, cleanable) 2089610-001-S FLOW SENSOR, LEGACY VAR ORF BCG, SERVICE (blue, cleanable) 2087640-001 FLOW SENSOR, LEGACY VAR ORF AUTOCLAVABLE BCG (gray, autoclavable) 2087640-001-S FLOW SENSOR, LEGACY VAR ORF AUTOCLAVABLE BCG, SERVICE (gray, autoclavable) 2096513-001-S FLOW SENSOR ASSEMBLY 5697309 R-FMI34109-FLOW SENSOR, LEGACY VAR ORF BCG 5697310 R-FMI34109-FLOW SENSOR, LEGACY VAR ORF AUTOCLAVABLE BCG
Quantity
Estimated 310,072 flow sensors
Codes and lots
Flow sensors with date of manufacture prior to 2021-06 etched on body of the flow sensor.
Recall started
Aug. 2, 2021
Device type
Gas-Machine, Anesthesia
Root cause (FDA)
Under Investigation by firm
What the company told customers
The recalling firm issued Urgent Medical Device Correction letters on 8/2/2021 via FedEx explaining the safety issue. This letter informs the customer of the issue and provides safety instructions, including requesting the consignee inspect all inventory of flow sensors. GE Healthcare will replace the flow sensors indicated on returned customer response forms, when applicable. The consignee will be responsible for replacing the flow sensor. The letter was sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, Health Care Administrator / Risk Manager
Recall number
Z-2595-2021

Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI…

Product
Flow sensors that may be installed in the following GEHC anesthesia machines: (1) Aestiva MRI, GTIN 0080682102339; (2) Aestiva 7900; (3) Aestiva 7100; (4) Aespire System Carestation, S/5 Protiva Config., Aespire 100 Compact Wall Mount (Aespire 7100); (5) S/5 Aespire 7900; (6) Aespire View; (7) Avance; (8) Avance CS2, GTIN 00840682102292; (9) Aisys; (10) Aisys CS2, GTIN 00840682102322; (11) Amingo; (12) 9100 NXT Config, Aelite NXT; (13) Protiva 7100;
Quantity
Estimated 310,072 flow sensors
Codes and lots
The flow sensors built prior to June 2021 that are installed in the devices/anesthesia machines.
Recall started
Aug. 2, 2021
Device type
Gas-Machine, Anesthesia
Root cause (FDA)
Under Investigation by firm
What the company told customers
The recalling firm issued Urgent Medical Device Correction letters on 8/2/2021 via FedEx explaining the safety issue. This letter informs the customer of the issue and provides safety instructions, including requesting the consignee inspect all inventory of flow sensors. GE Healthcare will replace the flow sensors indicated on returned customer response forms, when applicable. The consignee will be responsible for replacing the flow sensor. The letter was sent to the following titles within the affected accounts: Chief of Anesthesia, Director of Biomedical / Clinical Engineering, Health Care Administrator / Risk Manager
Recall number
Z-2596-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-2595-2021, Z-2596-2021. The agency record governs; check it and the company before acting.

Report a mistake on this page