FDA · Medical device recall
Centricity PACS-IW with Universal Viewer… and 1 more product
- Date
- Sept. 29, 2021
- Risk class
- Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
- Status
- Ongoing
- Filed by
- GE Healthcare, LLC
- Reason
- Image acquisition failures and synchronization failure with the Centricity Enterprise Archive
- Distribution pattern
- AK, AZ, CA, CO, DE, FL, GA, IA, IN, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, SD, TX, UT, WA, WI, WV and WY. Austria, Brazil, Canada, France, Germany, Ghana, India, Ireland, Israel, Italy, Kuwait, Netherlands, Panama, Poland, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Switzerland, Turkey, United Kingdom and Uruguay.
2 products in this recall
Centricity PACS-IW with Universal Viewer…
- Product
- Centricity PACS-IW with Universal Viewer - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources.
- Quantity
- 392 (219 US; 173 OUS) all products
- Codes and lots
- 2068177-001 (Centricity PACS-IW with Universal Viewer software versions 5.0.x with PACS-IW foundation) System ID: 083055248481416, 1-121967267, 1-429329281, DGC19607785, HC3062UV01, HCIT1401012UW, HCIT4331279UV, PACS-IW-127, PACS-IW-219, PACS-IW-2770, PACS-IW-323, PACS-IW-421, RICHMONDU14767331, RICHMONDU15531177N1, X1347906 / X1347909, X1347906 / X1347909, YP315021, ZA2412PAC21 and ZA2533PAC1
- Recall started
- Aug. 6, 2021
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design
- What the company told customers
- This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected software versions on August 06, 2021. This letter informs customers of the issues and provides safety instructions. GE Healthcare will correct the issue through a product change. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-2502-2021
Centricity Universal Viewer and Viewer 6.0…
- Product
- Centricity Universal Viewer and Viewer 6.0 - Product Usage: Is a device that displays medical images (including mammograms) and data from various imaging sources. Includes the following Model Numbers: 1. Model Number: 2090255-001; 2. Model Number: 2088026-723; 3. Model Number: 2102676-001; 4. Model Number: 2104867-044.
- Quantity
- 462 (279 US; 183 OUS)
- Codes and lots
- 1. Model Number: 2090255-001 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1) System ID (UDI-DI if applicable): PMIO14194768, PMIO20084199, PACS-IW-231, CITRUSVAL20884363, CITRUSVAL20884364, CITRUSVAL20884365, PACS-IW-155, PRIMEHEAL02315681, PRIMEHEAL20501355, PARKVIEWC21200827 (840682103800), PARKVIEWC21200830 (840682103800), PACS-IWUVN2-441, QUEEN19630914, QUEEN19630915, QUEEN19630916T, QUEEN20553950, 203268UVSW, DELAWAREC01589220, PACS-IW-462UV, PACS-IW-462UVN1, PACS-IW-126, PACS-IW-1062, OFMC01819402, OFMC20791988, HGS01766912, HGS20584004 (840682103800), GA001PACTRL01, TRGI02180958, DRY01810490, DRY19670358, PACS-IW-101, RML20502096, PACS-IWUV-239, PACS-IW-00813, PHS01374810, 01722652SEAS, 02031921SEASN1, MARYLANDF17985489, MYRTLEHIL17713895, MYRTLEHIL01444483, MYRTLEHIL17713895, VT01766830, VT01977629, PACS-IW-2637, SRMC01982629, PACS-IWUV-568, 1-142022244, PACS-IWUV-1131, OGAA02373799NOD1, ALLG00558571 (840682103800), EXC02189547, EXC01584099, PACS-IW-574, PRIME HEAL19508102, PRIME HEAL19050922, PRIME HEAL19050921, PACS-IW-219, ANCL00558570, ANCL02157634, NSW01452088, NSW20080545, MARYLANDF17985489, PACS-IW-518UV (840682103800), MTGD02393053, MTGD00558454, PACS-IW-44, QUENTIN01970717, LAKESIDEM02256808, BMC12P00558644P1, BMC12T02267839, BMC12N01271417N1, PACS-IW-58, SMCO02039946PC (840682103800), BWH01301357, PACS-IW-349, PACS-IW-255, 1-357059471, CAROCOMMU02205222, CAROCOMMU01459013, CSPA01291589 (840682103800), CSPA02097167 (840682103800), PACS-IW-79, PACS-IW-93, 1-276423210, MFCI01448329, LSFH02032334, CORONAREG01300927, 951737IWUWWN1, CORONAREG02288447, 1-382753509, CITRUSVAL19618406, MGMC02288391, MGMC02398531, 1-473998727, BARRETTHO00558704, BARRETTHO17737529, FORTDUNCA01298556 (840682103800), FORTDUNCA02395799 (840682103800), 1-283597082, TTU01973509, AVHL02396030, PACS-IW-511, PACSIWTEST, PACS-IW-386UV (840682103800), PACS-IW-486-UV (840682103800), PACS-IW-486-UVN1 (840682103800), PACS-IW-390 (840682103800), PACS-IW-202TEST, PACS-IW-202, HZC14961014, AVHL02396026, AVHL02396027, AVHL02396028, AVHL02396029, PACS-IW-419, PMRM14768587UVProxy, HANNIBALR02179353 (840682103800), TESTHANNIBALR02255613 (840682103800), HANN16174584UVProxy (840682103800), PACS-IW-189 (840682103800), HARTSDALE14768195 (840682103800), HOA01451973 (840682103800), RML01448365 (840682103800), ICA01367198 (840682103800), ICA14776604 (840682103800), JKS01452133 (840682103800), JKS14776447UVSFNode (840682103800), JKS14776448UVProxy (840682103800), JKS14776449UVTest (840682103800), PACS-IW-181, GTEC01477448, GTEC01477447, GTEC15993728UVProxy3, 1-592001096 (840682103800), UHSD14767800 (840682103800), PACS-IW-551UVN1 (840682103800), PACS-IW-551UVN2 (840682103800), UV1-154550926, DHOL19603374, DHOL19603375, PACS-IW-551UV (840682103800), PACS-IW-424CTLR, PACS-IW-424UVN1, PACS-IW-424UVN2, SFCS19667347, SFCS15961473, SFCS15961473, MHOG01806096, PACS-IW-166, PACS-IW-166T, MCT02277843, MCT02278145, PACS-IW-565, PACS-IWT-565, WGH00558334CTRL, WGH02372377PROX, WGH02372375BACK, 1-464783615, PACS-IW-389, BVH00558643, PACS-IW-348UV (840682103800), NRVM19609993UVT (840682103800), NRVM19611189UVPROXY (840682103800), PACS-IW-441, PACS-IW-441TUVSW, PACS-IW-402, PACS-IW-318, MINNETOHE, PACS-IW-253, PACS-IW-581UV, PACS-IW-370, PACS-IW-UVT370, VISOU02345023, VISOU02288781, 1-218291079, 1-103108035, MFHO14767332, PALMDALEM00558700, PALMDALEM01810788, PACS-IW-454, PACS-IW-406, 20101378UBSW, ILCC01373579 (840682103800), ILCC14580319 (840682103800), 516473UVSW (840682103800), JOHNTMA19732280T (840682103800), 1-182669905, NCDC20279342, CENTERFOR01627896, CENTERFOR16023960UVProxy, PACS-IW-176, PACS-IW-176UVTEST, PACS-IW-405, PACS-IW-336, NASONHOSP14898835, RRA01814001, MYRTLEHIL01444483, PSPC01815255, 1-283596904, HOOO20091140, GZMG02182469CNTRLR (840682103800), GZMG02287985 (840682103800), GZMG02288300 (840682103800), PACS-IW-556UV, URSA20025013, PACS-IW-431, 1-598891598, PACS-IW-230, INTERCOAST17594527UVProxy, MDHC16204735UVProxy, 201392UVSW, 505368UVBSW (840682103800), NNAR14767799 (840682103800), PACS-IW-479, PACS-IW-557 (840682103800), URS20583582 (840682103800), PACS-IWUV-481, PACS-IW-476, PACSIW476UVTEST, GGMC02349137, 1-486091448, 1-486091475, 1-750710672, SUC00558667, 1-802428413, 954688UVBSW100, PACS-IW-1511, KDH217192544, ACP00558109, PACS-IW-31Test1, PACS-IW-31Test, ACP01613226, 1-219095008, OKOS14767330, 918277CVUWBSW, CCA20759776SF, PACS-IW-487, QTHBL14767810UVPROXY, QTHBL14767809UVSFBK, QTHBL14767808UVSF1KO, QTHBL14767807UVSFSIM, QTHBL14767806UVSFVAL, QTHBL02256595, VCM02255583T, VCM01819618, VCM02255586TSP, VMC02255587NSP, BOJC00558520, PACS-IW-56, PACS-IW-225, 1-375705893, DDNI14776603, PACS-IW-129, DISSFHW01, 1-644600677, PACS-IW-154UV, FLMC14777071, NEWYORK01291738, OANT02035061, PACS-IW-424CTLR, PACS-IW-424UVN1, PACS-IW-424UVN2, VBHCP20497227, PACS-IW-2635, PREMI02182462, PREMI02182462, NORTHWEST01804127, HCIT4271649UW, AH4325SV01, AH5291PC02, AP6264PC01, AP6469PC01, AH0116SV03, AH0106SV03, HCD4440494UV, HCIT4258559UV, HCIT7109403UV, HCIT4244609UV, HCIT4708563UV, HCIT4470563UV-CTRL, HCIT4261024UVCTRL, HCIT7113573UV, 519264PACSAW1 (840682103800), 416813UV (840682103800), 416813TESTRPACSUV (840682103800), HCIT4266470UW, HCIT4238540UV, HCIT4486723UV, E4167658, X314119802, X141385617, A5485262 / A5485266 / A5485267 / A5485269 / A5485270 / A5485271, A9190177, M41980107, M4487535, M44997011, HC0399UV01, CC4105PC02, PC1695SV01, 786787SV02, HBON01SV02, HC4105PC01, 509940UV01, HC0529PC01, HC0529PC01, 233638UV01, HC0245PC02, 600791PC02, 427990UV01, PC1961SV02, 531344PC01, 434487SV03, PX4900SV02, HC0565PC03, TP0575UV02, GH1077UV01, 083055820116113, 083055824999515, 083055861021012, 83055862088813, 083055800046112, 81111DBS01, IL1469IT05, IL1469IT16, A5992711, A5412010, A6197102, KW0000PP02, HCIT4175024UV, AMS3963, BON0164, ECT0265, HCIT4192962UV, HCIT4264206UW, PL3673RW01, PL1493RW02, PL2970RW01, PL1124RW01, PL1499RW03, PL3349PC01, PL1501RW03, PL2624RW01, PL3249RW02, PL3277RW01, PL1005RW01, 600062UV, PL3362RW01, PL0326RW01, PL2975RW02, PL1504RW02, PL2971RW01, RU9255ID01, RU6968ID01, RU8945ID01, RU7519ID01, RU1781RW01, RU2580ID01, RU4255ID01, RU0003ID01, RU0472ID03, RU9595ID02, RU9327ID02, RU1011ID05, RU8366ID01, RU7808ID02, RU1516ID03, RU1646ID01, RU1112ID02, 01977813, 01980259, RU3052ID01, RU1107RW01, RU1107RW01, RU7175ID01, RU1717ID04, RU9216ID02, RU3453RW03, 600101RW02, RU8002ID01, RU1544ID01, RU1516ID03, RU9940RW01, RU5388ID01, RU0994ID03, RU3794ID01, RU9226ID01, RU9697ID01, 1509505, RU0968ID04, RU0120ID01, RU6164ID01, RU9546ID01, RU7468ID01, RU3998ID01, 600105ID03, SA2402PP02, SA2110PP05, SA1339PP03, ZA2541UV01, ZA2414PAC1, ZA2415PAC1, ZA2474UV01, ZA2267PAC1, ZA1058UV01, ZA2551PAC1, ZA1058UV01, ZA2412PAC2, ZA2472PAC2, ZA000010, ZA2412PAC2, MPX35219, DP760316, FP502804, 802030PC01, 228493WS01, 124774CS01, 482896SV03, 06051PAC04, HCIT4133734PACS, HCIT4148106PEE, Others Tracked under Service Cloud product ID. 2. Model Number: 2088026-723 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: NYHOSPI16204605, NYHOSPI16204606, NYHOSPI16204607, NYHOSPI16204608, NYHOSPI16204609, NYHOSPI16204612, NYHOSPI16204613. 3. Model Number: 2102676-001 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: PACS-IW-119. 4. Model Number: 2104867-044 (Centricity Universal Viewer with PACS-IW foundation software versions 6.0 through 6.0 SP7.1); UDI-DI: 840682103800; System ID: DDOC02339896.
- Recall started
- Aug. 6, 2021
- Device type
- System, Image Processing, Radiological
- Root cause (FDA)
- Software design
- What the company told customers
- This issue will be resolved by sending an Urgent Medical Device Correction letter to customers with affected software versions on August 06, 2021. This letter informs customers of the issues and provides safety instructions. GE Healthcare will correct the issue through a product change. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Recall number
- Z-2503-2021
What should I do? How serious is a recall? Classes and warnings explained
Source: FDA · Medical device enforcement reports, recall numbers Z-2502-2021, Z-2503-2021. The agency record governs; check it and the company before acting.