FDA · Medical device recall

HARMONIC HD 1000i Shears (20cm Shaft Length)-indicated for soft… and 1 more product

Sept. 1, 2021 Class II

Date
Sept. 1, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
Ethicon Endo-Surgery Inc
Reason
An internal component may be cracked and become lodged behind the energy button potentially resulting in continuous activation of the device, may result in burns
Distribution pattern
Worldwide distribution - US Nationwide distribution and the countries of Austria, Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Estonia, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Netherlands, New Zealand, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, United Arab Emirates.

2 products in this recall

HARMONIC HD 1000i Shears (20cm Shaft Length)-indicated for soft tissue incisions when bleeding control and…

Product
HARMONIC HD 1000i Shears (20cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. Product Code: HARHD20
Quantity
7870 units
Codes and lots
Lot Codes: U95126 U9571N U9526Z U95A80 U9543P U95T0X U9550G GTIN: 10705036015048
Recall started
July 20, 2021
Device type
Instrument, Ultrasonic Surgical
Root cause (FDA)
Process control
What the company told customers
Ethicon issued Urgent Medical Device Removal letter to Operating Room Supervisors, Materials Management Personnel, and Chief of Surgery, on July 20, 2021. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email it to Stericycle at 877-884-7828 or Ethicon7616@stericycle.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 5. Customers are required to return unused affected HARMONIC HD 1000i Shears subject to this recall that are in their inventory immediately. To receive replacement reimbursement, customers must return product subject to this recall no later than October 31, 2021. Any non-affected product and any product returned after the date specified will not receive replacement. 6. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your records. 7. To return product subject to this recall, photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for the shipping charges only if the authorized label is used. Extra shipping labels may be obtained by calling 888-3
Recall number
Z-2355-2021

HARMONICHD 1000i Shears (36cm Shaft Length)-indicated for soft tissue incisions when bleeding control and…

Product
HARMONICHD 1000i Shears (36cm Shaft Length)-indicated for soft tissue incisions when bleeding control and minimal thermal injury are desired. Product Code: HARHD36
Quantity
68960 units
Codes and lots
Lot Codes: U94Y9V U95C2A U94Z98 U95427 U9516E U95599 U9503G U95856 U94Y9W U95L0A U9518D U95526 U9507E U9548R U95127 U95525 U9530P U95T2Z U95366 U95R4F U9521Z U95F2T U9554U U95F8W U9564G U95U6A U95524 U95E6F U9512D U95K4D U94Z49 U95F07 U95A9X U95F6N U9569W U95E1U U9571P U95E72 U9587D U95E29 U95815 U95T90 U95814 U95Z67 U95754 U95Y7G U95D4T U9523L U95566 U9530X U9543R GTIN: 10705036015055
Recall started
July 20, 2021
Device type
Instrument, Ultrasonic Surgical
Root cause (FDA)
Process control
What the company told customers
Ethicon issued Urgent Medical Device Removal letter to Operating Room Supervisors, Materials Management Personnel, and Chief of Surgery, on July 20, 2021. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2. Remove the product subject to this recall and communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. 3. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 4. Complete the Business Reply Form (BRF) (Attachment 3) confirming receipt of this notice and fax or email it to Stericycle at 877-884-7828 or Ethicon7616@stericycle.com within three (3) business days. Please return the BRF even if you do not have product subject to this recall. 5. Customers are required to return unused affected HARMONIC HD 1000i Shears subject to this recall that are in their inventory immediately. To receive replacement reimbursement, customers must return product subject to this recall no later than October 31, 2021. Any non-affected product and any product returned after the date specified will not receive replacement. 6. Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Stericycle. While processing your returns, please maintain a copy of this notice with the product subject to this recall and keep a copy for your records. 7. To return product subject to this recall, photocopy the completed BRF, place it in the box with the product, and affix the pre-paid authorized shipping label included with the recall notification letter. Ethicon will pay for the shipping charges only if the authorized label is used. Extra shipping labels may be obtained by calling 888-3
Recall number
Z-2356-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement reports, recall numbers Z-2355-2021, Z-2356-2021. The agency record governs; check it and the company before acting.