Menu
FDA · Medical device recall

CARESCAPE PDM-Masimo SpO2…

Aug. 4, 2021 Class II

Date
Aug. 4, 2021
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Ongoing
Filed by
GE Healthcare, LLC
Product
CARESCAPE PDM-Masimo SpO2. Physiological data monitor.
Reason
Masimo SpO2 Saturation Values can become frozen after an extended length of use without a power down.
Distribution pattern
worldwide distribution US nationwide, Albania, Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Fiji, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Indonesia, Iraq, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, State of Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Vietnam.
Quantity
111,728 units
Codes and lots
CARESCAPE PDM: product codes SA3, SPX; model numbers 2042084-001, 2094504-001, 2016793-002, 2016793-003; Software kits, field replaceable units, and upgrade kits: part numbers 2034826-001, 2034826-002, 2034826-003, 2034826-004, 2034826-005, 2034826-006, 2034826-007, 2034826-008, 2034826-009, 2034826-010, 2034826-011, 2034826-012, 2034826-013, 2031069-010, 2045825-001, 2045825-002, 2045825-004, 2045825-005
Recall started
May 21, 2021
Device type
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Root cause (FDA)
Software Manufacturing/Software Deployment
What the company told customers
On May 21st the firm sent a letter to its customers with the following instructions: You can continue to use the CARESCAPE PDM with Masimo SET technology by completing the power down instructions below annually. 1. If the PDM is on a patient, provide alternate monitoring if needed, during the power down process. 2. Disconnect the PDM from the display device by removing the PDM from the mounting dock or by disconnecting the communication cable from the back of the PDM. 3. Open the battery door compartment and remove the battery (if present), for approximately 10 seconds. 4. Reinsert the battery and reconnect the PDM to the display device
Recall number
Z-2139-2021

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2139-2021. The agency record governs; check it and the company before acting.

Report a mistake on this page