GE Healthcare…

July 25, 2012 Class II

Date
July 25, 2012
Risk class
Class II (FDA's definition, 21 CFR 7.3(m): a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote)
Status
Terminated
Filed by
GE Healthcare, LLC
Product
GE Healthcare, Invasive Blood Pressure Care Cables. Product ID Number: 2016995-001, 2016995-002, 2016995-003, 2016995-004.
Reason
GE Healthcare has become aware through complaints of a potential safety issue with its Invasive Blood Pressure Cable made for use with a Spectramed / BD invasive blood pressure transducer. Due to use of an incorrect O-ring, the mating of the cable with the transducer is faulty. This situation inhibits blood pressure measurement from being monitored correctly.
Distribution pattern
Worldwide Distribution -- USA (nationwide) including the states of AK, CA, IL, MS, NJ, PR, SD, TN, TX, and VA. and the countries of ARGENTIAN, AUSTRALIA, BELGIUM, CHINA, GERMANY, ALGERIA, ECUADOR, EGYPT, FINLAND, UNITED KINGDOM, JORDAN, JAPAN, MOROCCO, MAURITUS, NETHERLANDS, OMAN, PERU, SEBIA, SUDI ARABIA, SINGAPORE, EL SALVADOR, TURKEY, URUGUAY and SOUTH AFRICA.
Quantity
279
Codes and lots
Product ID Number Lot Code 2016995-001 xxxx2010E 2016995-002 xxxx2010D 2016995-003 xxxx2010C 2016995-004 xxxx2010C
Recall started
June 12, 2012
Device type
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Root cause (FDA)
Nonconforming Material/Component
What the company told customers
GE Healthcare sent a Urgent Medical Device Correction letter dated June 12, 2012, to all affected customers. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Inforamtion. Advised customers to discontinue use, further distribution and to contact their customers. Please return any defective Invasive Blood Pressure Cables by isolating all affected product and completing the attached Confirmation form/fax back per its instructions. This step is required to confirm receipt of communications with all customers. If you have any questions or concerns regarding this notification, please contact Customer Service at +1-800-932-0760 (Domestic & International). Hours of Operation: 8:00 am EST to 6:00 pm EST. .
Recall number
Z-2015-2012

What should I do? How serious is a recall? Classes and warnings explained

Source: FDA · Medical device enforcement report, recall number Z-2015-2012. The agency record governs; check it and the company before acting.