FDA · Medical device recall

GE Healthcare Definium 8000 Digital Radiographic System

Sept. 25, 2008

Date posted by the FDA
Sept. 25, 2008
Status
Terminated
Filed by
GE Healthcare
Product
GE Healthcare Definium 8000 Digital Radiographic System.
Reason
Radiation dose reading greater than actual dose: Error in the generator software on the GE Healthcare Revolution XR/d configured with HP 8200 PC consoles and all Definium 8000 systems may impact patient safety. Occasional generator software errors may cause light X-ray Images with reported mAs readings higher than was actually exposed to the patient using the overhead X-ray tube. The mAs rea
Distribution pattern
Worldwide Distribution: USA, Australia, Austria, Belgium, Canada, China, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Guyana, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Martinique (French), Netherlands, Norway, Palestine, Portugal, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, Taiwan, and Yugoslavia.
Quantity
454 (277 US / 177 OUS)
Codes and lots
0000000002772E, 00000001191M36, 00000001561M30, 00000001895M32, 00000002041M32, 00000002051M31, 00000002052M39, 00000002065M31, 00000002364M38, 00000002439M38, 00000002441M34, 00000002650M30, 00000002767EN2, 00000002771M34, 00000002968M36, 00000003468M36, 00000003505M35, 00000003806M37, 00000004101M32, 00000004105M33, 00000004199M36, 00000004442M30, 00000096383DP5, 00000096401DP5, 00000097412DP1, 00000097656DP3, 00000097918DP7, 00000114134VE0, 00000114297VE5, 00000114452VE6, 00000TA41711-6, 00000TA41732-3, 00000TA41827-4, 00000TA41867-8, 00000TA41882-7, 00000TA41896¿4, 00000TA41972-3, 00000TA41984-4, 00000TA41984-6, 00000TA41995-7, 00000TA42004¿5, 00000TA42007-4, 00000TA42020-3, 00000TA42025¿4, 00000TA42274-8, 00000TA42869-6, 00001003925WK3, 00001003974WK1, 00001004022WK8, 00001004138WK2, 00001004163WK0, 00001004319WK8, 00001004331WK3, 00001004508WK6, 00001004509WK4, 00001004512WK8, 00001004763WK7, 00001004767WK8, 00001004769WK4, 00001004771WK0, 00001005016WK9, 00001005017WK7, 00001005018WK5, 00001005019WK3, 00001005021WK9, 00001005023WK5, 00001005025WK0, 00001005026WK8, 00001005027WK6, 00001005029WK2, 00001005032WK6, 00001005033WK4, 00001005034WK2, 00001005179WK5, 00001005180WK3, 00001005183WK7, 00001005184WK5, 00001005185WK2, 00001005186WK0, 00001005187WK8, 00001005319WK7, 00001005669WK5, 00001005698WK4, 00001005699WK2, 00001005701WK6, 00001005702WK4, 00001005705WK7, 00001005706WK5, 00001005707WK3, 00001005755WK2, 00001005757WK8, 00001005759WK4, 00001005761WK0, 00001005762WK8, 00001005763WK6, 00001005764WK4, 00001005841WK0, 00001005911WK1, 00001005912WK9, 00001005913WK7, 00001005915WK2, 00001006015WK0, 00001006040WK8, 00001006041WK6, 00001006043WK2, 00001006044WK0, 00001006177WK8, 00001006178WK6, 00001006179WK4, 00001006469WK9, 00001006470WK7, 00001006471WK5, 00001006475WK6, 00001006476WK4, 00001006477WK2, 00001006478WK0, 00001006631WK4, 00001006633WK0, 00001006634WK8, 00001006664WK5, 00001006665WK2, 00001006666WK0, 00001006667WK8, 00001006822WK9, 00001006833WK6, 00001006834WK4, 00001006872WK4, 00001006873WK2, 00001006874WK0, 00001006875WK7, 00001006876WK5, 00001006877WK3, 00001006878WK1, 00001006880WK7, 00001007160WK3, 00001007162WK9, 00001007163WK7, 00001007164WK5, 00001007165WK2, 00001007167WK8, 00001007168WK6, 00001007169WK4, 00001007188WK4, 00001007189WK2, 00001007190WK0, 00001007191WK8, 00001007192WK6, 00001007405WK2, 00001007406WK0, 00001007407WK8, 00001007409WK4, 00001007476WK3, 00001007477WK1, 00001007478WK9, 00001007497WK9, 00001007498WK7, 00001007499WK5, 00001007516WK6, 00001007517WK4, 00001007518WK2, 00001007525WK7, 00001007527WK3, 00001007528WK1, 00001007529WK9, 00001007530WK7, 00001007531WK5, 00001007532WK3, 00001007533WK1, 00001007534WK9, 00001007535WK6, 00001007536WK4, 00001007642WK0, 00001007643WK8, 00001007645WK3, 00001007646WK1, 00001007647WK9, 00001007649WK5, 00001007651WK1, 00001007713WK9, 00001007715WK4, 00001007716WK2, 00001007717WK0, 00001007718WK8, 00001007719WK6, 00001007720WK4, 00001007721WK2, 00001007776WK6…
Recall started
March 7, 2008
Device type
System, X-Ray, Stationary
Root cause (FDA)
Software design
What the company told customers
GE sent a Product Safety Notification dated March 7, 2008 to all customers. The letter described the safety issue, affected product, and stated that a GE Healthcare Field Engineer will schedule a field visit to install software updates.
Recall number
Z-1882-2008